← Product Code [MQL](/productcode/MQL) · K991331

# MEDI-CULT M3 MEDIUM (K991331)

_Medicult A/S · MQL · Jun 20, 2000 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K991331

## Device Facts

- **Applicant:** Medicult A/S
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Jun 20, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Medi-Cult M3 Medium is intended for the culture of embryos from day 3 and up to day 7.

## Device Story

Medi-Cult M3 Medium is a specialized cell culture medium used in assisted reproduction technology (ART) laboratories. It is formulated based on MCDB 302, a modification of HAM's F10 and F12, containing amino acids, vitamins, inorganic salts, and glucose. The medium is supplemented with sodium bicarbonate, human serum albumin (HSA), and optionally penicillin, streptomycin, and phenol red. It is designed to support the extended culture of embryos from day 3 through day 7, facilitating morula and blastocyst development. The medium is used by embryologists in clinical IVF settings. By providing a controlled environment for extended embryo development, it aims to improve blastocyst formation rates, which can assist in clinical decision-making regarding embryo selection for transfer. The product is subject to quality control testing for bioburden, sterility, pH, osmolarity, and endotoxin levels.

## Clinical Evidence

Clinical evidence includes two prospective randomized studies. One study at Sahlgrenska hospital compared Medi-Cult M3 to Scandinavian Science S2 medium, finding similar blastocyst formation rates. A second study at The Jones Institute for Reproductive Medicine compared a sequential culture system (Medi-Cult universal IVF-medium followed by M3) against a competitor system (S1 followed by S2), reporting similar blastocyst formation rates (26% vs 27%).

## Technological Characteristics

Formulation based on MCDB 302 (modified HAM's F10/F12). Components: amino acids, vitamins, inorganic salts, glucose, sodium bicarbonate, human serum albumin. Optional additives: penicillin, streptomycin, phenol red. Quality control includes bioburden, filter integrity, sterility, pH, osmolarity, and endotoxin testing.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Reference Devices

- Scandinavian Science S2 medium
- Scandinavian IVF Science S1 medium

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 2

## 510(k) Summary Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination (K991331)

The products:

- "Medi-Cult M3 Medium" Cat.No. 1033 -
- "Medi-Cult M3 Medium w/o Pen/Strep." Cat.No. 1032 ﮯ
- -"Medi-Cult M3 Medium w/o Phenol Red" Cat.No. 1037

Indication of use:

"Medi-Cult M3 Medium" is intended for the culture of embryos from day 3 and up to day 7.

Product formulation:

"Medi-Cult M3 Medium" is based on MCDB 302; a modification of HAM's F10 and F12 composed of amino acids, vitamins, inorganic salts and glucose supplemented with:

Assisted Reproduction Technique Supplement (ARTS) Sodium Bicarbonate Human Serum Albumin (HSA) (obtained from a licensed source, U.S. license No. 140) Penicillin (Except product 1032) Streptomycin (Except product 1032) Phenol Red (Except product 1037)

Products testing control contents:

- Bioburden, production-test -
- Integrity filter testing, production-test -
- -Sterility, QC-test
- pH, QC-test -
- -Osmolarity, QC-test
- -Endotoxin, OC-test

The culture media from Medi-Cult have been used by many European IVF-units since the end of the 1980's. (Produced and distributed by GEA-Biotech 1987 to 1989 and by Medi-Cult a/s from 1989)

The Medi-Cult M3 medium was introduced in 1992 to allow extended culture of embryos beyond day 2/3 after fertilisation. The Medi-Cult M3 medium has been used by several clinics obtaining good morula/blastocyst formation and good pregnancy rates.

A few prospective randomised studies have been performed showing that the Medi-Cult M3 medium performs equally well as other products available for extended culture.

{1}------------------------------------------------

K991331
Page 2 of 2

One center performed a prospective randomised study comparing the Medi-Cult M3 medium with the Scandinavian Science S2 medium. They found a similar rate of blastocyst formation (Source Dr. Anita Sjogren the Dept of Ob. & Gyn., Sahlgrenska hospital in Gothenburg, Sweden, data presented as poster at The world IVF meeting in Boston 1995).

The Jones Institute for Reproductive Medicine, Eastern Virginia Medical school, Norfolk, USA, has performed a prospective randomised study comparing the rate of blastocyst formation when human embryos are cultured in two different commercially available culture systems. The embryos were cultured either in the Medi-Cult universal IVF-medium for 2 days followed by The Medi-Cult M3 or in S1 for 2 days followed by S2 from Scandinavian IVF Science. The rate of blastocyst formation was similar in both groups (26% vs 27% respectively).

There have been no registered complaints on the product and there is no evidence in the last 1.5 year that the product has been the cause of any serious adverse events in connection with its intended use.

Thus based on the Federal Register Notice (Final Rule, Vol. 63, No. 175, page 48429, September 10, 1998), "Obstetric and Gynecologic Devices; Reclassification and Classification of Medical Devices Used for In-Vitro Fertilization and Related Assisted Reproduction Procedures" effective on October 13, 1998 and the supportive clinical data we feel that the safety and effectiveness of the product for its intended use is shown in the present submission.

Prepared and Submitted by:

Rute Wicheep. Lenardi 1/22/02

**Ronald G. Leonardi, Ph. D.** **Date**

Ronald G. Leonardi, Ph. D. President R & R Registrations P.O. Box 262069 San Diego CA 92196 1-858-586-0751

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 0 2000

Medi-Cult A/S c/o Ronald G. Leonardi, Ph.D. President R & R Registrations P.O. Box 262069 San Diego, CA 92196-2069

Re: K991331 Medi-Cult M3 Medium Dated: May 22, 2000 Received: May 23, 2000 Regulatory Class: II 21 CFR §884.6180/Procode: 85 MQL

Dear Dr. Leonardi:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 20, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act indude requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalion assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulator, In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notfication submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html",

Sincerely yours,

Daniel G. Schultz, M.D.

Captain, USPHS Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

510(k) Number (if known) K991331

Device Name: Medi-Cult M3 Medium

Indications for Use:

M3 Medium is used for the culture of embryos from day 3 and up to day 7.

(PLEASE DO NO WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

David A. Ingram

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K991331](https://fda.innolitics.com/device/K991331)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
