← Product Code [LPH](/productcode/LPH) · K991325

# MODIFICATION OF ENCORE LINEAR POROUS COATED HIP (K991325)

_Encore Orthopedics, Inc. · LPH · Jun 25, 1999 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K991325

## Device Facts

- **Applicant:** Encore Orthopedics, Inc.
- **Product Code:** [LPH](/productcode/LPH.md)
- **Decision Date:** Jun 25, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3358
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The indications for use of the total hip replacement prosthesis include: noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; rheumatoid arthritis; correction of functional deformity; revision procedures where other treatments or devices have failed; and treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement, which is unmanageable using other techniques. This stem is to be press-fit.

## Device Story

Cementless hip stem for total hip replacement; fabricated from Ti-6Al-4V substrate with commercially pure titanium porous coating for press-fit fixation. Features rectangular cross-section for rotational stability, 6° proximal flare, and 135° stem/neck angle. Available in standard and lateralized versions to accommodate patient anatomy without increasing leg length. Utilizes Morse type taper for modular head attachment. Intended for surgical implantation by orthopedic surgeons in clinical settings. Device provides structural support and joint articulation to restore hip function and mobility in patients with degenerative or traumatic hip conditions.

## Clinical Evidence

Bench testing only. Evidence includes finite element analysis of fatigue stem strength and laboratory testing of the Morse type taper.

## Technological Characteristics

Materials: Wrought/forged Ti-6Al-4V (ASTM F136/F620) with commercially pure titanium porous coating (ASTM F67 grade 2). Design: Collarless, rectangular cross-section, 6° proximal flare, 135° stem/neck angle. Fixation: Press-fit. Interface: Morse type taper for modular heads.

## Regulatory Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Encore Orthopedics®. Inc. 9800 Metric Blvd Austin, TX 78758 512-832-9500

K991325

Trade Name: Linear™ Porous Coated Hip Stem

Common Name: Cementless hip stem

Classification Name: Hip joint metal/polymer/metal semi-constrained porous coated uncemented prosthesis

## Description

When viewed in the mediolateral plane the Linear™ Porous Coated Hip Stem tapers slightly proximal to distal beginning at the distal border of the porous coating. Proximal to this point the stem flares approximately 6°. The stem has a rectangular cross-sectional geometry to provide rotational stability

The Linear™ Porous Coated Hip Stem is fabricated from wrought/forged Ti-6Al-4V that conforms to ASTM F136 or F620, respectively. The outside surface of the stern is porous coated with commercially pure titanium (ASTM F67 grade 2) to provide a porous surface for enhanced fixation. The stem is intended to be press-fit.

The Linear™ Porous Coated Hip Stem is collarless and has a Morse type taper to receive modular heads. This stem is available with a standard head/stem offset and a lateralized version that provides additional lateralization of the patient's femur without increasing leg length. The stem/neck angle is 135°.

Intended Use: The indications for use of the total hip replacement prosthesis include: noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; rheumatoid arthritis; correction of functional deformity; revision procedures where other treatments or devices have failed; and treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement, which is unmanageable using other techniques. This stem is to be press-fit.

Comparable Features to Predicate Device(s): Features comparable to predicate devices include Ti-6Al-4V substrate, straight stem, symmetric, no collar, narrow slightly taper stem in lateral view, and modular heads.

Test Results: Finite element analysis of the fatigue stem strength and laboratory testing of the Morse type taper was conducted.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract human profiles facing to the right, with flowing lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 5 1999

Ms. Debbie De Los Santos Requlatory/Clinical Specialist Encore Orthopedics, Inc. 9800 Metric Boulevard Austin, Texas 78758

Re: K991325 Linear™ Porous Coated Hip Stem - Size 5 Trade Name: Regulatory Class: II Product Codes: LPH and LZO Dated: April 16, 1999 Received: April 19, 1999

Dear Ms. De Los Santos:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ಕ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{2}------------------------------------------------

## Page 2 - Ms. Debbie De Los Santos

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) Number (if known): __

K991325

Device Name: __ Linear™ Porous Coated Hip Stem_

Indications For Use:

## Linear™ Porous Coated Hip Stem Indications For Use

The indications for use of the total hip replacement prosthesis include: noninflammatory I he midications for alle or allo to and the interest and avascular necrosis; theumatoid arthritis; correction of functional deformity, revision procedures where other treatments or devices have colronon of rancelonal coronial) fenoral neck and trochanteric fractures of the proximal femur with head involvement, which is umanageable using other techniques. This stem is to be pressfit.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use X
(Per 21 CFR 801.109)

OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)_

Division Sign Off

(Division of General Restor 5 IUlia) Number

.

---

**Source:** [https://fda.innolitics.com/device/K991325](https://fda.innolitics.com/device/K991325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
