The Pacific Hemostasis ThromboScreen® 200 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma specimens in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), as well as PT and APTT-based factor assays.
Device Story
ThromboScreen® 200 (TS200) is a manual photo-optical coagulation instrument for clinical laboratories. Input: citrated plasma specimens. Operation: user manually pipettes samples and reagents into cuvettes; instrument uses 470 nm LASER light source to monitor optical changes during clotting. Incubator block maintains 36.5–37.5°C. Output: measured clotting time. Healthcare providers use results to assess coagulation status, including PT, APTT, fibrinogen levels, and factor activity. Benefits: provides standardized, automated optical detection for routine coagulation assays compared to manual visual methods.
Clinical Evidence
Bench testing only. Method comparison studies performed at two external sites and in-house using 137-141 samples for PT and 104-121 for APTT. Correlation coefficients (r) ranged from 0.97 to 0.99 across all assays (PT, APTT, Fibrinogen, Factor V/VIII). Precision studies (within-run and between-run) showed CVs generally comparable to predicate devices. No clinical patient outcome data provided.
Technological Characteristics
Photo-optical coagulation analyzer. Light source: 470 nm LASER. Temperature-regulated incubator block (36.5–37.5°C). Manual pipetting system. Two measuring positions, three reagent positions, 12 cuvette prewarming positions. Standalone benchtop unit.
Indications for Use
Indicated for in-vitro diagnostic coagulation testing of citrated plasma specimens in clinical laboratories. Includes PT, APTT, Fibrinogen (Clauss and Derived), and PT/APTT-based factor assays. No specific patient population age or gender restrictions provided.
Regulatory Classification
Identification
A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.
Special Controls
*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{1}------------------------------------------------
JUN 21 1999
# Premarket Notification 510(K) Summary ThromboScreen® 200
| 10.0 | PREMARKET NOTIFICATION 510(K) SUMMARY |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Laura A. Worfolk, Ph.D.<br>Pacific Hemostasis<br>11515 Vanstory Drive<br>Huntersville, NC 28078<br>704-875-0494<br>Fax # 704-875-2092 |
| Contact: | Larry Kopyta, Manager FDA Programs<br>704-875-0494<br>Fax # 704-875-2092 |
| Date: | April 16, 1999 |
| Trade Name: | Pacific Hemostasis ThromboScreen® 200 |
| Common Name: | Manual Coagulation Instrument |
| Classification Name: | Instrument, Coagulation (per 21 CFR section 864.5400) |
| Equivalent Devices: | MLA-900C &1000C, #K884863, #K894052 |
#### Description of the ThromboScreen® 200
The ThromboScreen® 200 (TS200) is a photo-optical instrument used for the performance of invitro diagnostic clotting procedures in the clinical laboratory. The instrument utilizes photo-optical principles to measure and record the time required for patient plasma specimens to clot. The ThromboScreen® 200 light source is provided by a 470 nm LASER. The incubator block is temperature regulated to 36.5 - 37.5℃ and contains two measuring positions, three reagent and 12 cuvette prewarming positions. A detailed description of the device, including an explanation of how it functions, is described in the ThromboScreen® 200 Operator's Manual, section 1, Introduction.
#### Intended Use of the ThromboScreen® 200
The Pacific Hemostasis ThromboScreen® 200 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing of citrated plasma specimens in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), as well as PT and APTT-based factor assays.
{2}------------------------------------------------
#### Summary of Substantial Equivalence Comparisons
The ThromboScreen® 200 (TS200) was compared to the MLA-900C and the MLA-1000C (K884863 & K894052, respectively). All three instruments have a similar intended use: for in-vitro diagnostic coagulation testing in the clinical laboratory. Further, the proposed device and the predicate devices have the same measurement system for clotting assays: photo-optical clot detection systems.
The TS200 is a "manual" coagulation instrument, in that the user must pipet both sample and test reagent. In contrast, the MLA-900C is semi-automated, and the MLA-1000C is a fully-automated instrument. The MLA-900C requires manual sample addition, but has an automatic pipet for reagent addition. The MLA-1000C has an automatic pipetting system, which adds both sample and test reagent. The light source for the MLA instruments is a Halogen lamp and the wavelength is set at 550 nm (for clotting assays). In contrast, the TS200 utilizes an LASER optic at 470 nm. Although differences in light source and wavelength exist, all instruments have been optimized for their light source/filter combinations. The performance data generated support this statement (Tables 1-3).
Comparison testing was performed in-house and at two external testing laboratories using Pacific Hemostasis (PH) brand reagents. Specimens were evaluated from apparently healthy individuals and from patients with different pathological conditions which are expected to affect the results for a particular assay. Table 1 summarizes the results of the comparison studies between the proposed and the predicate devices.
| Test<br>(Reagent, Unit) | Site & Sample # | Correlation<br>Coefficient, r | Regression Equation |
|----------------------------------------------------------------------------|-----------------------------------------|-------------------------------|------------------------------------------------------------------|
| Prothrombin Time (PT)<br>(Thromboplastin DS,<br>seconds) | Site #1 - 137<br>Site #2 - 141 | 0.98<br>0.99 | y = 1.078x + 0.309<br>y = 1.076x - 0.526 |
| Prothrombin Time<br>(Thromboplastin DS,<br>INR) | Site #1 - 137<br>Site #2 - 141 | 0.98<br>0.99 | y = 0.899x + 0.186<br>y = 1.0754x - 0.1063 |
| Activated Partial<br>Thromboplastin Time*<br>(APTT-LS reagent,<br>seconds) | Site #1 - 104<br>Site #2 - 121 | 0.98<br>0.97 | y = 1.328x - 0.305<br>y = 1.072x + 4.351 |
| Clauss Fibrinogen<br>(PH Thrombin reagent,<br>mg/dL) | Site #1 - 20<br>Site #2 - 20<br>PH - 49 | 0.99<br>0.98<br>0.98 | y = 1.059x - 11.526<br>y = 1.038x - 14.293<br>y = 1.001x - 9.257 |
| Derived Fibrinogen<br>(Thromboplastin DS,<br>mg/dL) | Site #1 - 19<br>Site #2 - 47 | 0.99<br>0.99 | y = 0.775x + 96.12<br>y = 0.896x + 23.02 |
| Factor VIII<br>(APTT-LS, % activity) | PH - 49 | 0.97 | y = 0.844x + 14.07 |
| Factor V,<br>(Thromboplastin DS, %<br>activity) | PH - 45 | 0.97 | y = 1.071x + 2.01 |
#### Table 1. Summary of Method Comparison Studies Between the TS200 & the MLA-900C/1000C
The predicate device used at site #1 & PH was the MLA-1000C, at site # 2 , the MLA-900C. *= APTT
{3}------------------------------------------------
The following coefficients of variation were obtained for within-run and between-run precision studies:
| Test | Low | TS200 | | MLA-900C/1000C | | |
|---------------|-----|--------|------|----------------|--------|------|
| | | Normal | High | Low | Normal | High |
| PT | | 1.9 | 2.3 | | 1.1 | 2.8 |
| Site 1 | | | | | | |
| Site 2 | | 2.0 | 5.7 | | 1.5 | 2.0 |
| PH | | 2.8 | 3.2 | | 1.0 | 2.8 |
| APTT | | | | | | |
| Site 1 | | 5.1 | 1.6 | | 1.4 | 2.9 |
| Site 2 | | 4.7 | 2.3 | | 3.3 | 2.2 |
| PH | | 3.3 | 2.5 | | 0.9 | 1.2 |
| Clauss Fib.* | 6.7 | 6.8 | 5.0 | 2.0 | 2.1 | 2.8 |
| Derived Fib.* | 5.6 | 2.5 | 4.1 | 2.2 | 3.4 | 2.1 |
| Factor V* | 2.3 | 2.5 | | 4.0 | 2.0 | |
| Factor VIII* | 9.6 | 10.2 | | 5.3 | 4.7 | |
Table 2. Summary of Within-run Precision Studies, %CV
*Testing at PH only. (Shaded areas, no testing performed. Only clinically significant ranges tested.)
Table 3. Summary of PT & APTT Between-run Precision Testing
| Prothrombin Time Testing | | | | Activated Partial Thromboplastin<br>Time Testing | | | | |
|--------------------------|---------------|------|-----------------|--------------------------------------------------|---------------|------|-----------------|------|
| | Normal Plasma | | Abnormal Plasma | | Normal Plasma | | Abnormal Plasma | |
| | TS200 | MLA | TS200 | MLA | TS200 | MLA | TS200 | MLA |
| mean | 12.9 | 12.8 | 45.1 | 40.0 | 35.4 | 29.7 | 69.5 | 59.0 |
| SD | 0.22 | 0.18 | 2.68 | 3.24 | 1.28 | 1.27 | 1.96 | 4.41 |
| CV | 1.7% | 1.4% | 5.9% | 8.1% | 3.6% | 4.3% | 2.8% | 7.5% |
| n | 39 | 40 | 38 | 40 | 39 | 40 | 40 | 40 |
Testing performed at PH only.
In conclusion, the similar intended use, technological characteristics and performance data support the claim that the ThromboScreen® 200 is substantially equivalent to the MLA-900C and the MLA-1000C.
4
{4}------------------------------------------------
#### PREMARKET NOTIFICATION
TRUTHFUL AND ACCURATE STATEMENT [As required by 21 CFR 807.87(j)]
I certify that, in my capacity as a Research Scientist at Pacific Hemostasis, a Fisher Scientific Company, I believe to the best of my knowledge, that all data and information submitted in the premarket notification are truthful and accurate and that no material fact has been omitted,
이 대한민국 대학교 동문학교 대학교 대학교 대학교 대학교 대학
មាន ១០
德国际网址 利来国际官网 利来
Laura A. Worfolk
**Laura A. Worfolk, Ph.D.**
K991321
*(Premarket Notification [510(k)] Number)
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract caduceus or a series of interconnected, flowing lines, possibly representing people.
JUN 21 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Laura A. Worfolk, Ph.D. Research Scientist Pacific Hemostasis 11515 Vanstory Drive Suite 125 Huntersville, North Carolina 28078-8144
Re: K991321
> Trade Name: Pacific Hemostasis ThromboScreen 200 (TS200) Regulatory Class: II Product Code: KQG Dated: June 4, 1999 Received: June 7, 1999
Dear Dr. Worfolk:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Page of of
510(k) Number (if known): 4991321
Device Name:__________________________________________________________________________________________________________________________________________________________________
## Indications For Use:
The Pacific Hemostasis ThromboScreen® 200 is a photo-optical instrument used for the performance of in-vitro diagnostic coagulation testing in the clinical laboratory. Coagulation testing capabilities of the device include routine clotting tests such as Prothrombin Time (PT), Activated Partial Thromboplastin Time (APTT), Fibrinogen (Clauss and Derived methods), as well as PT and APT-based factor assays.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Peter E. Maderi
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K99321
Prescription Use V (Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.