← Product Code [FAS](/productcode/FAS) · K991314

# COAGULOOP (K991314)

_American Medical Systems, Inc. · FAS · Jun 29, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K991314

## Device Facts

- **Applicant:** American Medical Systems, Inc.
- **Product Code:** [FAS](/productcode/FAS.md)
- **Decision Date:** Jun 29, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The AMS Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder.

## Device Story

Electrosurgical device; used with resectoscopes and monopolar electrosurgical generators. Input: radio frequency electrical current. Function: removes and coagulates soft tissue via RF energy. Used in clinical settings (e.g., OR) by physicians. Output: tissue resection/coagulation. Benefit: enables surgical removal and hemostasis of prostate/bladder tissue.

## Clinical Evidence

Bench testing only. Biocompatibility testing followed ISO-10993 recommendations. Sterilization validated per ANSI/AAMI and ISO 11137 standards.

## Technological Characteristics

Monopolar electrosurgical resection electrode; compatible with standard resectoscopes and generators. Sterilization: gamma radiation (SAL 10^-6). Biocompatibility: ISO-10993 compliant.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- AMS Coaguloop™ Resection Electrode ([K971512](/device/K971512.md))
- Circon/ACMI Vaportome™ Resection Loop Electrode ([K973820](/device/K973820.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 2 9 1999

Image /page/0/Picture/2 description: The image is a black and white logo. The logo is an abstract design that resembles the letter 'A'. The 'A' is formed by a series of vertical and horizontal lines. The lines are arranged in a way that creates a sense of depth and dimension. The logo is simple, yet visually striking.

# AMERICAN MEDICAL SYSTEMS

# 510(k) SUMMARY

#### SUBMITTER I.

Name and Address:

American Medical Systems, Inc. 10700 Bren Road West Minnetonka, MN 55343 USA

2183959

David Worrell

April 15, 1999

2183959

-

Establishment Registration Number:

Contact Person:

Date of Summary Preparation:

#### II. MANUFACTURER

Name and Address:

Device Trade Name:

Classification Name:

American Medical Systems 10700 Bren Road West Minnetonka, MN 55343-9679

Establishment Registration Number

#### III. DEVICE NAME

Device Common or Usual Name:

Resection Loop or Resection Electrode

AMS Coaguloop™ Resection Electrode

Endoscopic electrosurgical unit with accessories (21 CFR 876.4300)

{1}------------------------------------------------

#### DEVICE DESCRIPTION IV.

The AMS Coaguloop™ Resection Electrode is an electrosurgical device designed for use with commercially available resectoscopes and general purpose monopolar electrosurgical generators. The Coaguloop™ Resection Electrode is designed to remove and coagulate soft tissue by means of radio frequency electrical current.

#### V. INDICATION FOR USE

The Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder.

#### SUBSTANTIAL EQUIVALENCE SUPPORT VI.

American Medical Systems' Coaguloop™ Resection Electrode is similar in indications, design, and features to various resectoscope electrodes that are in commercial distribution. The predicate devices that AMS is claiming substantial equivalence to are the AMS Coaguloop™ Resection Electrode K971512 and the Circon/ACMI Vaportome™ Resection Loop Electrode K973820.

#### STERILIZATION METHODOLOGY VII.

Sterilization is based on the recommendations in ANSI/AAMI and ISO 11137 Guideline for the Sterilization of health care products - Requirements for validation and routine control - Radiation sterilization. The devices are terminally sterilized by exposure to gamma radiation. This gamma radiation sterilization cycle will achieve a minimum Sterility Assurance Level (SAL) of 10°.

### VIII. -BIOCOMPATIBILITY

Biocompatibility testing generally followed the recommendations of FDA Blue Book Memorandum #G95-1 entitled "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing."

## CONCLUSION

In summary, the data and information provided within this 510(k) Premarket Notification adequately support that the AMS Coaguloop™ Resection Electrode is substantially equivalent to the AMS Coaguloop™ Resection Electrode K971512 and the Circon/ACMI Vaportome™ Resection Loop Electrode K973820.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN 2 9 1999

Mr. David Worrell Regulatory Affairs Specialist American Medical Systems, Inc. 10700 Bren Road West Minnetonka, MN 55343

Re: K991314 AMS Coaguloop™ Resection Electrode Dated: April 15, 1999 Received: April 16, 1999 Requlatory Class: II 21 CFR §876.4300/Procode: 78 FAS

Dear Mr. Worrell:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compilance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page _1_ of _1_

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

510(k) Number (if Known): ____________________________________________________________________________________________________________________________________________________

AMS Coaguloop™ Resection Electrode Device Name:

Indications For Use:

The AMS Coaguloop™ Resection Electrode is intended to be used for resection, ablation, and fulguration in the prostate gland and the bladder.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|

| Prescription Use     | <div style="text-align:left;">/</div> |
|----------------------|---------------------------------------|
| (Per 21 CFR 801.109) |                                       |

OR

| Over-The-Counter Use | _________________ |
|----------------------|-------------------|
|----------------------|-------------------|

| (Division Sign-Off)                                                |         |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |         |
| 510(k) Number                                                      | K991314 |

---

**Source:** [https://fda.innolitics.com/device/K991314](https://fda.innolitics.com/device/K991314)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
