The genX Single Lumen Sideport Needle is a stainless steel needle (32 cm length, 16 or 17 gauge) with a polycarbonate hub. It features a dual-tubing configuration: one tube connects to the needle rear for oocyte aspiration into a collection test tube; the second connects to a side flush port for syringe-based media injection into the follicle. Used by clinicians during assisted reproductive procedures to retrieve oocytes. The device facilitates follicle flushing to improve oocyte recovery. It is provided sterile via gamma irradiation.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Stainless steel needle tubing (32 cm length, 16/17 gauge); polycarbonate hub; dual-tubing for aspiration and flushing. Sterilization: Gamma irradiation (2.5 Mrad) per ANSI/AAMI/ISO 11137-1994. SAL 10^-6.
Indications for Use
Indicated for obtaining oocytes from the body for assisted reproductive procedures.
Regulatory Classification
Identification
Assisted reproduction needles are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to obtain gametes from the body or introduce gametes, zygote(s), preembryo(s) and/or embryo(s) into the body. This generic type of device may include a single or double lumen needle and component parts, including needle guides, such as those used with ultrasound.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
K191291 — Wallace Dual Lumen Oocyte Recovery System · CooperSurgical, Inc. · Jun 12, 2019
Submission Summary (Full Text)
{0}------------------------------------------------
SEP g 1999
<genX> international. Inc. 510(K) Premarket Notification
#### 510K Summary
K991264
- a) DEVICE NAME Proprietary Name: <genX> Single Lumen Sideport Needle
Classification Name: Assisted Reproduction Needle
#### ESTABLISHMENT REGISTRATION No.: 9003605 b)
<genX> international. Inc. 170 Fort Path Road Madison, CT 06443 Tel: 203-245-4901 Fax: 203-245-4994
Contact Individual: Michael D. Cecchi President
- CLASSIFICATION: Class II C) Assisted reproduction needles Product: <genX> Single Lumen Needle with Sideport Procode: 85 MQE CFR#: 884.6100
- PERFORMANCE STANDARDS d) Performance Standards under Section 514 of the ACT have not been developed for this device. However, Special controls have been identified by the FDA to provide reasonable assurance of safety and effectiveness of the device in assisted reproductive procedures.
- PROPOSED LABELS, LABELING, AND ADVERTISING e) The proposed labeling and instruction material is included in this package in the appropriate section.
- f) DESCRIPTION
The <genX> Single Lumen Sideport Needle is intended to obtain gametes from the body. The catalog number is GNSP-732.
The basic design is a single stainless steel needle tubing of 32 cm in exposed length, with a grounded point. It is attached to a hub of polycarbonate designed especially for this purpose. The outer diameter of the outer needle tube maybe 16 or 17gauge, 17 being more desirable.
Attached to the hub are two lengths of tubing. One is attached to the rear of the needles, which allows the occytes to be aspirated into a collection test tube, and the second length is attached to the flush port on the side. This second port is
{1}------------------------------------------------
attached to a syringe, which allows the operator to inject media into the follicle to allow "flushing"
- SUBSTANTIALLY EQUIVALENT g) The Single Lumen Sideport Needle is substantially equivalent to several products currently marketed in the US. See the appropriate section for the comparisons. (Section 2)
## h) STERILIZATION PROCEDURES AND FACILITIES
All needles manufactured, will be Gamma Irradiated by and according to the guidelines in place.
Dosage: 2.5 Mrad Sterility Assurance Level (SAL): of (10 to the -6 ) After sterilization the contractor will issue a "Certificate of Sterilization".
Validation assurance is in accordance with ANSI / AAMI / ISO 11137-1994 Standards.
See Section 11.0
・・
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 9 1999
Mr. Michael D. Cecchi President <genX> International, Inc. 170 Fort Path Road, Unit 14 Madison, CT 06443
Dear Mr. Cecchi:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act indude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
Re: K991264
<genX> Single Lumen Sideport Needle
21 CFR §884.6100/Procode: 85 MQE
Dated: June 10, 1999
Regulatory Class: II
Received: June 22, 1999
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
<genX> international, Inc. 510(K) Premarket Notification
Response #2
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Re: K991264 Single Lumen Sideport Needle
# INDICATIONS FOR USE
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
510 (k) Number ( if known )___________________________________________________________________________________________________________________________________________________
Device Names: <genX> Single Lumen Sideport Needle Indication for Use:
.
"Used to obtain oocytes from the body"
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED )
and the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the comments of the commen
Concurrence of CDRH, Office of Device Evaluation ( ODE )
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K991264 |
|---------------|---------|
|---------------|---------|
| Prescription Use | <div></div> |
|----------------------|-------------|
| or | <div></div> |
| Over-the Counter Use | <div></div> |
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.