← Product Code [DTB](/productcode/DTB) · K991169

# SX 53/15-BP, MODEL 124 853, SX 60/15-BP, MODEL 124 854 (K991169)

_Biotronik, Inc. · DTB · Jul 1, 1999 · Cardiovascular · ST_

**Canonical URL:** https://fda.innolitics.com/device/K991169

## Device Facts

- **Applicant:** Biotronik, Inc.
- **Product Code:** [DTB](/productcode/DTB.md)
- **Decision Date:** Jul 1, 1999
- **Decision:** ST
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3680
- **Device Class:** Class 3
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Synox leads are designed for use with implantable pulse generators that require pacing leads with a bipolar 3.2 mm IS-1 connector configuration; they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pulse generator therapy is medically indicated. This indication follows that recommended in the Class I definition of the ACC/AHA Task Force Report, entitled "Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmic Devices" (JACC, Vol. 18, No. 1, July 1991:1 - 13).

## Device Story

Synox leads are bipolar, passive fixation endocardial pacing leads; designed for implantation in heart trabeculae. Device features titanium tip electrode with fractal iridium surface; platinum/iridium ring electrode with fractal iridium surface; quadrifilar MP35N coaxial conductor; silicone rubber insulation; 3.2 mm IS-1 connector. Modification from predicate involves reducing ring-to-tip electrode spacing from 31 mm to 15 mm and standardizing insulation thickness. Leads connect to implantable pulse generators to deliver pacing therapy. Used by physicians in clinical settings for patients requiring cardiac pacing. Benefits include reliable electrical sensing and stimulation for cardiac rhythm management.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Bipolar, passive fixation endocardial lead. Tip electrode: titanium base, fractal iridium surface (1.3 mm²). Ring electrode: platinum/iridium base, fractal iridium surface (34 mm²). Conductor: quadrifilar MP35N wire, coaxial configuration. Insulation: silicone rubber. Connector: 3.2 mm IS-1. Sterilization: not specified.

## Regulatory Identification

Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.

## Predicate Devices

- Synox passive fixation endocardial leads ([K980869](/device/K980869.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

BIOTRONIK, Inc., Synox Passive Fixation Endocardial Lead, 510(k)

# Synox Passive Fixation Endocardial Lead 510(k) Notification

#### 510(K) SUMMARY 1.

| Name and Address of Sponsor:       | BIOTRONIK, Inc.<br>6024 Jean Road<br>Lake Oswego, OR 97035                                                                                                                    |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 1028232                                                                                                                                                                       |
| Device Name:                       | Proprietary Names: Synox Leads<br>Classification: Class III (21 CFR 870.3680(b))<br>Classification Name: Cardiovascular Permanent<br>Pacemaker Electrode<br>Product Code: DTB |
| Date Prepared:                     | April 2, 1999                                                                                                                                                                 |

## General Description and Predicate Devices:

The Synox 53/15-BP and 60/15-BP leads introduced in this 510(k) notification are modified versions of BIOTRONIK's currently marketed straight Synox leads (SX 53-BP and SX 60-BP) approved September 10, 1998 under 510(k) #K980869. The modification is a reduction of the distance between the ring and tip electrodes from 31 mm to 15 mm. Because of this change, the silicone insulation between the tip and ring electrodes was shortened and the prior tapered design was changed to one with a consistent thickness. The new 15 mm spacing distance between the ring and tip electrodes is identical to the electrode spacing of the "J" shaped Synox leads (SX 45-JBP and SX 53-JBP) also approved September 10, 1998 under 510(k) #K980869. The SX xx/15-BP straight leads with 15 mm spacing will not replace the existing straight leads, but will be distributed in conjunction with the currently marketed Synox leads.

Synox xx/15-BP leads are bipolar, passive fixation endocardial leads that feature three for fixation in the heart's trabeculae. The tip electrode is made of a titanium base material with a surface treatment of fractal iridium and a surface area of 1.3 mm². The ring electrode is made of a platinum/iridium base material with a surface treatment of fractal iridium and a surface area of 34 mm². The lead conductor is quadrifilar MP35N wire in a coaxial configuration, insulated with sillcone rubber tubing. All Synox leads utilize a 3.2 mm IS-1 connector.

The predicate devices are BIOTRONIK'S Synox passive fixation endocardial leads (straight and "U" configurations) cleared for commercial distribution on September 10, 1998 under 510(k) #K980869.

### Indications for Use:

Synox leads are designed for use with implantable pulse generators that require pacing leads with a bipolar 3.2 mm IS-1 connector configuration; they may be used with single or dual chamber pacing systems. The leads are designed for use in patients for whom single or dual chamber pulse generator therapy is medically indicated. This indication follows that recommended in the Class I definition of the ACC/AHA Task Force Report, entitled "Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmic Devices" (JACC, Vol. 18, No. 1, July 1991:1 - 13).

#### Name and Address of Manufacturing Site:

BIOTRONIK GmbH & Co. (reg. no. 7010992) Woermannkehre 1, 12359 Berlin, Germany 011-49-30-689-05-304

Contact Person(s) and Phone Number: Jon Brumbaugh Regulatory Affairs Manager Phone (888) 345-0374 Fax (503) 635-9936

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an eagle or bird with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the graphic.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 1 1999

Mr. Jon Brumbaugh Manager, Regulatory Affairs BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035

Re : K991169 Trade Name: Synox SX 53/15-BP and SX 60/15-BP Endocardial Pacing Leads Regulatory Class: III Product Code: DTB Dated: April 2, 1999 Received: April 7, 1999

Dear Mr. Brumbaugh:

We have reviewed your Section 510 (k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obliqation you might have under section 531 through 542 of the Act for devices

{2}------------------------------------------------

Page 2 - Mr. Jon Brumbaugh

under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

On August 16, 1993 the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirements of the regulation are found in 21 CFR 821 as described in the August 16, 1993 Federal Reqister beginning on page 43447.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) Additionally, for questions on the promotion and advertising 594-4586. of your device, please contact the Office of Compliance at (301) 594-4639. Also, please not the regulation entitled, " Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Quitymthater

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

{3}------------------------------------------------

#### INDICATIONS FOR USE 2.

Synox leads are designed for use with implantable pulse generators that require pacing hads with a Synox leads are designed for use with implanable pares with single or dual chamber pacing bigolar 3.2 min 10-1 connector ochingation, they may whom single or dual chamber pulse generator Systems. The leads are dolly indication follows that recommended in the Class I definition of the theraly is including indicated. This includines for Implantation of Cardiac Pacemakers and Antiarrhythmic Devices" (JACC, Vol. 18, No. 1, July 1991:1 - 13).

C. Witzman for Callahan

(Division Sign-Off) Division of Cardiovascular rological Devic 510(k) Numb

Image /page/3/Picture/6 description: The image shows a handwritten note with the name "Mike" at the top left. Below "Mike", the words "prescription-device" are written inside a roughly drawn circle. At the bottom of the image, there is a signature that appears to read "Chi".

---

**Source:** [https://fda.innolitics.com/device/K991169](https://fda.innolitics.com/device/K991169)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
