← Product Code [LPL](/productcode/LPL) · K991134

# ACUVUE (ETAFILCON A)SOFT (HYDROPHILIC) CONTACT LENSES, CLEAR AND VISIBILITY TINT, WITH UV BLOCKER, FOR DAILY WEAR (K991134)

_Vistakon, Johnson & Johnson Vision Products, Inc. · LPL · May 6, 1999 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K991134

## Device Facts

- **Applicant:** Vistakon, Johnson & Johnson Vision Products, Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** May 6, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism. The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism. The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism. The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less. ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye. Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

## Device Story

ACUVUE soft (hydrophilic) contact lenses (etafilcon A) are indicated for daily wear to correct refractive errors including myopia, hyperopia, presbyopia, and astigmatism. Lenses incorporate UV-blocking properties to reduce transmission of UV radiation to the eye. Prescribed by eye care practitioners for single-use disposable or frequent/planned replacement wear. For frequent replacement, lenses require cleaning and chemical disinfection. Device serves as a vision correction aid; UV blocking provides supplemental ocular protection. No changes to material, manufacturing, or physical properties from predicate; submission solely updates indications for use labeling.

## Clinical Evidence

No new clinical studies were conducted. Safety and efficacy are supported by clinical data previously submitted in N18-033/S31 and K962804. The submission relies on the established clinical history of the lens material and design.

## Technological Characteristics

Soft (hydrophilic) contact lens composed of etafilcon A. Features UV-blocking capability (82% UVA, 97% UVB). No changes to materials, dimensions, or manufacturing processes from predicate.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- ACUVUE (etafilcon A) soft (hydrophilic) contact lenses ([K962804](/device/K962804.md))

## Reference Devices

- N18-033/S31

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Summary of 510(k) Submission

| Name and<br>address of<br>submitter | Vistakon, Johnson & Johnson Vision Products, Inc.<br>4500 Salisbury Road, Suite 300<br>Jacksonville, Florida 32216<br>Contact: Martine D. Martino<br>Phone: (904) 443-1808<br>Date Prepared: April 1, 1999                                                                                                          |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Identification of<br>device         | The trade name is ACUVUE (etafilcon A) soft (hydrophilic) contact<br>lenses, clear and visibility tint, with UV blocker, for daily wear The common or usual name is Soft (hydrophilic) Contact Lens (daily<br>wear). The FDA Classification is Class II.                                                            |
| Predicate<br>devices                | The predicate device is ACUVUE (etafilcon A) soft (hydrophilic) contact<br>lenses, clear and visibility tint, with UV blocker, for daily wear covered under<br>K962804.                                                                                                                                             |
| Description of<br>device            | The device description does not change from that cleared under K962804.                                                                                                                                                                                                                                             |
|                                     | Continued on next page                                                                                                                                                                                                                                                                                              |
| Intended use                        | The reason for the 510(k) is to add the following Indications Statement (see <b>bold text</b> ) for ACUVUE (etafilcon A) soft (hydrophilic) contact lenses, clear and visibility tint, with UV blocker, for daily wear. The revised Indications Statement is included with this 510(k):                             |
|                                     | The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.                                                                              |
|                                     | The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.                                                                                         |
|                                     | The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.                                                                                    |
|                                     | The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.                                                                                     |
|                                     | ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.                                                                                                                                                                                         |
|                                     | Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only. |

{1}------------------------------------------------

## Summary of 510(k) Submission, Continued

Continued on next page

{2}------------------------------------------------

## Summary of 510(k) Submission, Continued

| Characteristics                      | The characteristics do not change. They are the same as previously submitted<br>in N18-033/S31 and K962804.                                                                                                                                                                                                                                                                                               |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Non-clinical<br>studies              | Non-clinical studies (microbiology, toxicology, chemistry, shelf-life, and<br>leachability) on the lens material were not conducted because the lens<br>material, etafilcon A, does not change. Non-clinical studies were conducted to support the following modified<br>Actions in the labeling. The average UV blocking for ACUVUE Contact Lenses with UV<br>Blocker for UVA is 82% and for UVB is 97%. |
| Clinical studies                     | This 510(k) describes a labeling modification: an additional Indications<br>Statement. There is no change in lens material, the manufacturing process,<br>nor the parameters and properties, therefore, the clinical data previously<br>submitted in N18-033/S31 and K962804 supports the clinical safety of the<br>subject device.                                                                       |
| Conclusions<br>drawn from<br>studies | Additional studies were not conducted, therefore, the conclusions drawn from<br>studies previously submitted in N18-033/S31 and K962804 support the non-<br>clinical and clinical safety of the subject device.                                                                                                                                                                                           |

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized representation of a human figure, consisting of three overlapping profiles facing to the right, with flowing lines suggesting movement or connection.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 6 1999

Vistakon™ Johnson & Johnson Vision Care Products, Inc. C/O Ms. Martine D. Martino P.O. Box 10157 Jacksonville, FL 32247

Re: K991134

Trade Name: ACUVUE (etafilcon A) soft (hydrophilic) contact lens, clear and with visibility tint, with UV blocker, for daily wear Regulatory Class: II Product Code: LPL Dated: April 1, 1999 Received: April 2, 1999

Dear Ms. Martino:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Martine D. Martino

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marked predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Ralph L. Rosenthal, M.D.

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications Statement

K991134 510(k) Number:

ACUVUE (etafilcon A) Soft (hydrophilic) Contact Lens Clear and Visibility Device Name: Tint with UV Blocker

The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or nonaphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.

ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)

Karin Waibmtor
(Division Sign-Off)
Division of Ophthalmic Devices
Prescription Use ✓ 510(k) Number K991134 OR Over the Coumer

Continued on next page

---

**Source:** [https://fda.innolitics.com/device/K991134](https://fda.innolitics.com/device/K991134)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
