← Product Code [JPD](/productcode/JPD) · K991127

# VARIANT II B-THALASSEMIA (K991127)

_Bio-Rad · JPD · Jun 10, 1999 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K991127

## Device Facts

- **Applicant:** Bio-Rad
- **Product Code:** [JPD](/productcode/JPD.md)
- **Decision Date:** Jun 10, 1999
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 864.7400
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The VARIANT™ II β-thalassemia Program is intended for the separation and area percent determinations of hemoglobins A2 and F and as an aid in the identification of abnormal hemoglobins in whole blood using ion-exchange high performance liquid chromatography (HPLC). The VARIANT™ II β-thalassemia Program is intended for use only with the Bio-Rad VARIANT™ II Hemoglobin Testing System. For in vitro diagnostic use only.

## Device Story

The VARIANT™ II β-thalassemia system is an automated ion-exchange HPLC platform for hemoglobin analysis. It processes whole blood samples, performing automated mixing and dilution via the VARIANT™ II Sampling Station (VSS). Samples are injected into an analytical cartridge where hemoglobin components are separated based on ionic interactions using a programmed buffer gradient. A filter photometer measures absorbance at 415 nm, with 690 nm used for background correction. The Clinical Data Management (CDM) software reduces raw data, calculates HbA2/F values, and generates patient reports and chromatograms. The system uses barcode-based sample identification and run sequencing. Compared to the manual VARIANT predecessor, this system automates sample preparation and run setup, stores chromatograms, and allows for reanalysis. Healthcare providers use the generated reports to aid in identifying abnormal hemoglobins, supporting clinical diagnosis of hemoglobinopathies.

## Clinical Evidence

Bench testing only. Performance characteristics were evaluated by comparing the VARIANT™ II β-thalassemia program to the predicate VARIANT™ system. Testing confirmed excellent concordance between the two methods and met all pre-defined acceptance criteria.

## Technological Characteristics

Ion-exchange HPLC system; automated sample preparation and dilution; dual-pump buffer gradient delivery; filter photometer (415 nm/690 nm); barcode-based sample tracking; CDM software for data reduction and reporting; one-level calibration for value adjustment.

## Regulatory Identification

A hemoglobin A2 assay is a device used to determine the hemoglobin A2 content of human blood. The measurement of hemoglobin A2 is used in the diagnosis of the thalassemias (hereditary hemolytic anemias characterized by decreased synthesis of one or more types of hemoglobin polypeptide chains).

## Predicate Devices

- VARIANT™ β-thalassemia ([K924122](/device/K924122.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# K99川27

JUN 1 0 1999

----

## Special 510(k): Device Modification Summary

| Submitter:                   | Bio-Rad Laboratories, Inc.<br>4000 Alfred Nobel Drive,<br>Hercules, California 94547<br>Phone: (510) 741-6188<br>FAX: (510) 741-6471                                                                                                                                                     |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:              | Juliet Carrara<br>Regulatory Affairs/Quality Assurance Manager                                                                                                                                                                                                                           |
| Date of Summary Preparation: | March 29, 1999                                                                                                                                                                                                                                                                           |
| Device Name:                 | VARIANT™ II β-thalassemia                                                                                                                                                                                                                                                                |
| Classification Name:         | Class II, 81JPD Hemoglobin A₂ Quantitation                                                                                                                                                                                                                                               |
| Unmodified Device:           | VARIANT™ β-thalassemia<br>K924122<br>Bio-Rad Laboratories<br>Hercules, CA 94547                                                                                                                                                                                                          |
| Statement of Intended Use:   | The VARIANT™ II β-thalassemia Program is<br>intended for the separation and area percent<br>determinations of hemoglobins A2 and F and as an<br>aid in the identification of abnormal hemoglobins in<br>whole blood using ion-exchange high performance<br>liquid chromatography (HPLC). |
|                              | The VARIANT™ II β-thalassemia Program is<br>intended for use only with the Bio-Rad<br>VARIANT™ II Hemoglobin Testing System                                                                                                                                                              |
|                              | For in vitro diagnostic use only.                                                                                                                                                                                                                                                        |

{1}------------------------------------------------

#### Description of Device

The VARIANT™ II is a fully automated HPLC system which can be used to separate and determine area percentages for hemoglobins A, and F and to provide qualitative determinations of abnormal hemoglobins.

The VARIANTM II B-thalassemia Short Program utilizes principles of ion exchange high performance liquid chromatography(HPLC). The samples are automatically mixed and diluted on the VARIANT™ II Sampling Station(VSS) and injected into the analytical cartridge. This is a change from the unmodified program(VARIANT) where samples had to be mixed and diluted manually before being placed on the instrument. The VARIANT™ II chromatographic station(VCS) dual pumps deliver a programmed buffer gradient of increasing ionic strength to the cartridge, where the HbA, F are separated based on their ionic interactions with the cartridge material. The separated HbA. F then pass through the flow cell of the filter photometer, where changes in the absorbance at 415 nm are measured. An additional filter at 690 nm corrects the background absorbance. The VARIANT™ II Clinical Data Management(CDM) software performs reduction of raw data collected from each analysis. One level calibration is used for the adjustment of the calculated HbAyF values. A patient sample report and a chromatogram are generated by CDM for each sample. Minor differences in the separation efficiency of individual analytical cartridges are corrected by the use of the Hemoglobin A /F Calibrator.

## Technical Characteristics Compared to Unmodified Device

The main difference between the VARIANT™ and VARIANT™ II involves the patient sample preparation. In VARIANT™ II the preparation is automated, in VARIANT™ the preparation is manual.

| Technical Characteristic                       | VARIANT I                               | VARIANT II                           |
|------------------------------------------------|-----------------------------------------|--------------------------------------|
| Sample Preparation:<br>Calibrator and controls | Manual Preparation.                     | Manual Preparation                   |
| Sample Preparation:<br>Patient Samples         | Manual Preparation.                     | Automatic Preparation                |
| Run SET UP                                     | Run sequence is programmed<br>manually. | Run sequence read from bar<br>codes. |

{2}------------------------------------------------

| Technical<br>Characteristic | VARIANT                                                          | VARIANT II                                                       |
|-----------------------------|------------------------------------------------------------------|------------------------------------------------------------------|
| STAT Samples                | Available.                                                       | Not Available.                                                   |
| Analyte Identification      | Analyte peaks are labelled.                                      | Analyte peaks are labelled.                                      |
| Calibration                 | Calibration response factors are used to adjust observed values. | Calibration response factors are used to adjust observed values. |
| Data Retrieval              | Chromatograms are not stored in the database.                    | Chromatograms are stored in the database.                        |
| Reanalysis                  | Not available                                                    | Available                                                        |
| Sample Identification       | Manual identification                                            | Barcode identification                                           |
| Summary Report              | Not available                                                    | Available                                                        |

·

.

រុ

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

、

6

{3}------------------------------------------------

## Performance Characteristics

Testing described in Section F of this submission focuses on performance characteristics of VARIANT™ II ß-Thal Short. Testing met all acceptance criteria.

When considering the similarities of the intended use, characteristics of the two VARIANT ß-thalassemia programs, the use of the same technology, and the excellent concordance between the two methods, it can be concluded that the VARIANT and the VARIANT II B-thalassemia programs are substantially equivalent. Based on the establishment of substantial equivalence, the safety and effectiveness of the VARIANT II ß-thalassemia program is confirmed.

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with its wings spread, depicted with three curved lines.

JUN 1 0 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Juliet Carrara Regulatory Affairs and Quality Assurance Manager Bio-Rad Laboratories, Inc. 4000 Alfred Nobel Drive Hercules, California 94547

Re: K991127

Trade Name: VARIANT™ II ß-thalassemia Short Program Regulatory Class: II Product Code: JPD Dated: May 28, 1999 Received: June 3, 1999

Dear Ms. Carrara:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

## Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

## Statement of Indications for Use

510(k) Number:

K991127

Device Name:

Indications for Use:

Bio-Rad VARIANT™ II ß-thalassemia

The VARIANT™ II ß-thalassemia Program is intended for the separation and area percent determinations of hemoglobins A2 and F and as an aid in the identification of abnormal hemoglobins in whole blood using ion-exchange high performance liquid chromatography (HPLC).

The VARIANT™ II ß-thalassemia Program is intended for use only with the Bio-Rad VARIANT™ II Hemoglobin Testing System

For in vitro diagnostic use only.

|                  | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE                           |                         |
|------------------|-----------------------------------------------------------------------------------------|-------------------------|
|                  | Concurrence of CDRH, Office of Device Evaluation (ODE)                                  |                         |
|                  | <div> <img alt="signature" src="signature.png"/> </div>                                 |                         |
|                  | (Division Sign-Off)<br>Division of Clinical Laboratory Devices<br>510(k) Number K991127 |                         |
| Prescriptive Use | <img alt="check" src="check.png"/>                                                      | OR over-the-counter Use |
|                  | (Per 21 CFR 801.109)                                                                    |                         |

---

**Source:** [https://fda.innolitics.com/device/K991127](https://fda.innolitics.com/device/K991127)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
