K-DEFIB/PACE RADIOLUCENT ADULT ELECTRODE, MODEL KRT-60

K991007 · Katecho, Inc. · MKJ · Aug 23, 1999 · Cardiovascular

Device Facts

Record IDK991007
Device NameK-DEFIB/PACE RADIOLUCENT ADULT ELECTRODE, MODEL KRT-60
ApplicantKatecho, Inc.
Product CodeMKJ · Cardiovascular
Decision DateAug 23, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The Katecho External Radiolucent electrode is indicated for use in external Pacing, Defibrillation and Monitoring applications as a non-sterile, disposable device, for single patient use, only. The KRT-60 Radiolucent electrode provides the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patients skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 Joule maximum).

Device Story

KRT-60 Radiolucent Defib/Pace Adult Electrode; non-sterile, disposable, single-patient use. Provides conductive interface between patient skin and external defibrillator/transcutaneous pacemaker. Used in clinical settings for defibrillation, cardioversion, or external pacing. Operated by healthcare professionals. Device facilitates energy delivery from defibrillator to patient; enables cardiac monitoring. Radiolucent design allows for use during imaging procedures. Benefits patient by providing necessary electrical therapy for cardiac arrhythmias.

Clinical Evidence

Bench testing only. Device tested to ANSI/AAMI DF2-1996 standard for compatibility with specified defibrillator/pacemaker equipment.

Technological Characteristics

Radiolucent electrode; conductive interface; disposable. Tested to ANSI/AAMI DF2-1996 standard. Designed for low power (360 Joule max) defibrillator output.

Indications for Use

Indicated for adult patients requiring external pacing, defibrillation, or monitoring. Used as a conductive interface between the patient's skin and defibrillator/transcutaneous pacemaker equipment. Intended for low power (max 360 Joule) defibrillators.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 AUG 23 1999 Mr. Warren R. Walters -Katecho, Inc. 2500 Bell Avenue · Des Moines, IA 50321 Re: K991007 K-Defib/Pace Radio Lucent Adult Electrode Model KRT-60 Regulatory Class: III (three) Product Code: MLN Dated: July 9, 1999 Received: July 12, 1999 Dear Mr. Walters: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {1}------------------------------------------------ Page 2 - Mr. Warren R. Walters This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97), Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Figure/0 description: This image shows a document with the title "Exhibit 14, Page 1 of 1". The document includes the date 8-3-99 and the time 11:59 AM. It also includes the text "Indications for use Statement", the 510k Number K991007, and the Device Name "KRT-60 Radiolucent Defib/Pace Adult Electrode". Indications For Use: The Katecho External Radiolucent electrode is indicated for use in external Pacing, Defibrillation and Monitoring applications as a non-sterile, disposable device, for single patient use, only. The KRT-60 Radiolucent electrode provides the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patients skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 Joule maximum). This device has been tested to the ANSI/AAMI DF2-1996 standard for use with equipment: Laerdal, Heartstream, Marquette, Physio Control, Survivalink, First Medic, and Hewlett-Packard. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) leA. Weathersbee (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number Prescription Use OR Over-The Counter Use (Por 21 CFR 801.109) 2500 Bell Avenue · Des Moines, lowa 50321 · Phone (515) 244-1212 · FAX (515)-244-4912
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...