The nasal CPAP unit is intended for treatment of adult Obstructive Sleep Apnea (OSA). Obstructive sleep apnea is defined as the absence of air movement for ten seconds during sleep. Obstructive sleep apnea is usually diagnosed by a sleep study. The study obtains the optimum level of pressure required to maintain an unobstructed airway, the positive pressure allows the is ve or prosuropen. The nasal CPAP unit is intended only for spontaneously breathing adult patients.
Device Story
SoftAire Nasal CPAP System provides continuous positive airway pressure to treat adult obstructive sleep apnea. Device uses 12 VDC brushless motor driving impeller to generate airflow; air intake passes through filter and muffler. Clinician adjusts therapeutic pressure (4-18 cmH2O) and ramp timer (0-20 min). System connects to patient via 22mm smooth bore tubing and user-selected nasal mask; requires external exhaust port. Operated by clinician or patient; user controls limited to on/off and ramp mode. CPU (4 MHz) manages motor speed and monitors pressure via independent sensor. External 12V power source required. Device maintains airway patency during sleep, potentially reducing apnea events.
Clinical Evidence
Bench testing only. Laboratory performance testing verified compliance with device specifications: therapeutic pressure range (4-18 cmH2O), pressure accuracy (±10%), timer functionality, noise level (<38dB), and electrical/environmental safety (EN 60601-1, EN 60601-1-2, IEC 68-2).
Indicated for spontaneously breathing adult patients diagnosed with Obstructive Sleep Apnea (OSA) via sleep study.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
# 510(K) SUMMARY
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# SoftAire Nasal CPAP System
Bird Products Corporation
Tom Gutierrez P.E. Regulatory Compliance Engineer Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262-6267
760.778.7341 (voice) 760.778.7274 (fax)
June 18, 1999
{1}------------------------------------------------
510(k) Summary for the tem, page 2 of 4
## General Information
. . .
Device Trade Name: SoftAire Nasal CPAP System
#### Device Common/Classification Name:
The SoftAire Nasal CPAP System is Bird Products Corporation ventilator classified under 73 BZD "Non-Continuous Ventilator"
- · 868.5905 Non-Continuous Ventilator, 73 BZD
| Predicate Device: | ResMed Sullivan III Nasal CPAP System<br>FDA 510(k) No: 930656 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------|
| Year 2000 Compatibility: | The SoftAire Nasal CPAP System does not perform any date<br>calculation or processing and is therefore year-2000 compliant. |
### Device Description
The SoftAire Nasal CPAP System is intended to provide continuous positive airway pressure for the care and treatment of individuals suffering from obstructive sleep apnea. The positive pressure is clinician adjustable within the designated operating range and a clinician adjustable timer will allow timed rise to pressure ("Ramp" mode). The user controls will be limited to an on/off switch and option of using the timed pressure rise. This product is
designed to be used with a 22 millimeter (smooth bore) tube air delivery system and user selected nasal mask. This mask is not marketed by Bird.
Power will be provided by AC to DC adapter s which accommodate a variety of input voltages and frequencies. This adapter will plug into a 12 VDC connector on the side of the unit.
Image /page/1/Picture/12 description: The image shows a close-up of a black electrical box with a label on it. The label contains the model number 15682 and serial number KJTC00313. It also includes electrical requirements of 12VDC 2 amps and several warnings and cautions. The label also has several certification logos.
Bottom View of SoftAire Nasal CPAP System
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a close-up of a gas mask. The mask is black and has a boxy shape. Two corrugated tubes extend from the mask, one on each side. The tubes are also black and appear to be made of a flexible material.
SoftAire Nasal CPAP System with tube and nasal mask.
The SoftAire Nasal CPAP System does not utilize an actively controlled exhalation valve. The patient circuit consists of two essential parts: a large-bore tube from the ventilator to the patient and a required exhaust port. The exhaust is an orifice which may be part of the adapter between the tubing an the patient mask, or may be an orifice which is part of the patient mask.
#### Electrical/Pneumatic System
Image /page/2/Figure/5 description: This image is a block diagram of a system. The system includes a 100 PPI filter, a muffler, and a 12 VDC blower with a 22 MM outlet. The system also includes a power supply that can accept either 110 VAC or 220 VAC, a connector that is 2.5MM x 5.5MM, an electronic module, a user input, and a temp fuse.
Flow to the patient is provided via 12 VDC Brushless Motor driving an impeller. Air intake passes through a filter and muffler into the motor, which operates at a preset speed established by the (clinical) operator. Patient tubing pressure is verified by an independent pressure measurement.
An external 12-Volt power source supplies a 5-Volt regulator, which provides power to all interval components. The CPU runs at 4 MHz. There are 5 switches underneath the unit for Mode, Plus, Minus, "Ramp Mode On" and "Ramp Mode Off",
{3}------------------------------------------------
#### Comparison to Predicate Device
The SoftAire Nasal CPAP System is not significantly different from the predicate device, the ResMed Sullivan III Nasal CPAP, first cleared for market under 510(k) K930656. Both devices utilize Blower type flow generators with the Bird SoftAire Nasal CPAP having a maximum working pressure of 22cmH,O.
----
### Summary of Performance Testing
Performance testing was conducted in the laboratory to confirm operation compliance to the device specifications. Testing to Environmental, EMI/RFI and Electrical Safety Standards were performed by certified test facilities. The following table specifies all SoftAire Nasal CPAP functions. All functions were verified to operate as designed and intended, as denoted by the checkmark.
| Specification | Requirement | |
|------------------------------------------|--------------------------------------------------------------------------|---|
| Adjustable Therapeutic<br>Pressure Range | 4-18 cmH2O. | ☑ |
| Adjustable Beginning<br>Pressure Range | 4 and 18 cmH2O and below therapeutic range. | ☑ |
| Pressure Accuracy | $\pm$ 10% to the nearest displayed digit (claibrated at 500ft elevation) | ☑ |
| Timer | 0, 5, 10, 15, 20 minutes, $\pm$ 2 minutes | ☑ |
| Input Voltage | 12 $\pm$ 10% VDC, 2A | ☑ |
| Noise Level | < 38dB | ☑ |
| Dimensions | 11X8X4 inches | ☑ |
| Weight | 5.2 lb | ☑ |
| Tubing | 6 ft, 22mm female connectors | ☑ |
| Electrical Safety | EN 60601-1 | ☑ |
| EMI/RFI | EN 60601-1-2 & Draft Reviewers Guidance for Ventilators 1995 | ☑ |
| Environmental | IEC 68-2 | ☑ |
| Operating Temperature | +10 to +40°C, 30 to 75% RH | ☑ |
| Storage Temperature | -20 to +60°C, up to 95% RH | ☑ |
Performance testing verified that the SoftAire Nasal CPAP meets all of it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare, featuring a staff with a serpent entwined around it.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 1 1 2000
Mr. Tom Gutierrez Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262
Re: K990856 SoftAire Nasal CPAP System Regulatory Class: II (two) Product Code: 73 BZD Dated: October 13, 1999 Received: October 14, 1999
Dear Mr. Gutierrez:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Mr. Tom Gutierrez
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Jeanroth W. Westerhausen for,
Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
| Page | | of | |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----|--|
| | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | |
| Acres of Children | l | 1 | |
| | 1 | | |
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The nasal CPAP unit is intended for treatment of adult Obstructive Sleep Apnea (OSA). Obstructive sleep apnea is defined as the absence of air movement for ten seconds during sleep. Obstructive sleep apnea is usually diagnosed by a sleep study. The study obtains the optimum level of pressure required to maintain an unobstructed airway, the positive pressure allows the is ve or prosuropen. The nasal CPAP unit is intended only for spontaneously breathing adult patients.
#### (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF REQUIRED) * ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ -----------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Wutcherson
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K990856 510(k) Number ________________________________________________________________________________________________________________________________________________________________
Prescription Use
✓
OR
Over-the-Counter Use
(Per 21 CFR 801.109)
-----------------------
(Optional format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.