K990809 · Caditec AG · DQD · Jun 9, 1999 · Cardiovascular
Device Facts
Record ID
K990809
Device Name
CADISCOPE ELECTRONIC STETHOSCOPE
Applicant
Caditec AG
Product Code
DQD · Cardiovascular
Decision Date
Jun 9, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Indications for Use
The CADIscope is a patented electronic stethoscope with an integrated graphics display to show the amplified heart sounds as phonocardiograms and show parallel to the heart sounds the ECG in an Einthoven Triangle. It is intended for use as a diagnostic aid as part of a physical assessment of a patient by health care professionals or other individuals trained to administer emergency first aid or otherwise care for a patient. It can be used for the amplification of heart, lung, blood vessel, enteral and other body sounds. When the device's integral electrodes are placed on the chest of the patient, it is capable of verifying the existence of cardiac biopotential activity. The quality of the biopotential activity display depends upon user technique and environmental conditions and it is in no way meant to be diagnostic.
Device Story
CADIscope is a battery-operated electronic stethoscope with integrated ECG monitoring. It captures body sounds via a subminiature back-electret microphone and cardiac biopotentials via three AgCl-plated electrodes forming an Einthoven triangle. A microprocessor processes these inputs, providing sound amplification (0-35 dB) and filtering (heart, lung, entero modes). Outputs include audio via earphones and visual representations (phonocardiograms and 1-lead ECG) on an integrated 64x96 pixel LCD. Used by clinicians or trained first-aid personnel in any environment for physical assessment. The device provides real-time status (heart rate, signal quality, battery) and includes a freeze function for the display. It aids clinical decision-making by providing visual confirmation of cardiac activity alongside auscultation. It does not store data or transmit to external systems.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Battery-operated electronic stethoscope; surface mount devices; mixed signal processor. Features AgCl-plated electrodes for ECG, subminiature back-electret microphone for sound. 64x96 pixel LCD display. Three selectable frequency filters (heart: 15-600 Hz; lungs: 300-1500 Hz; entero: 15-1500 Hz). 16-level volume control (0-35 dB). Powered by 3x1.5V alkaline batteries. Portable form factor (170g).
Indications for Use
Indicated for use as a diagnostic aid during physical assessment of patients by healthcare professionals or individuals trained in emergency first aid. Used for amplification of heart, lung, blood vessel, enteral, and other body sounds, and for verifying cardiac biopotential activity via chest-placed electrodes.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
E-Scope Electronic Stethoscope Model 718-7120 (K961301)
{0}------------------------------------------------
K990809
CH-6343 Rotkreuz-Zug, Switzerland Phone xx41-41-79 80 120 Fax xx41-41-79 80 133
#### SUMMARY OF SAFETY AND EFFECTIVENESS 12.
1 February, 1999
- SUBMITTER/APPLICANT COMPANY INFORMATION 1. Registration #: None CADItec AG. EURO 1 Rotkreuz-Zug Switzerland, CH6343 Phone : xx41-41-79 80 120 Fax : xx41-41-79 80 133 Email : sales@caditec.com Contact Person: Fridolin Voegeli Contact Title: President
- SUBMISSION CORRESPONDENT 2. Medline Industries, Inc. One Medline Place Mundelein. IL 60060 Registration #: 1417592 (847) 949-5500 x1131 Phone: (847) 949-2643 Fax: Contact Name: Christine M. Galea, Submission Correspondent
| REGULATORY INFORMATION: | |
|-------------------------|------------------------------------------------------|
| DEVICE NAME: | Stethoscope, Electronic |
| PROPRIETARY NAME: | CADIscope Electronic Stethoscope with Integrated ECG |
| COMMON NAME: | Electronic Stethoscope |
| CLASS: | II |
| PRO CODE: | DQD (870.1875) |
| PERFORMANCE STANDARDS: | None |
- 4. DEVICE DESCRIPTION: The CADIscope is a patented electronic stethoscope with an integrated graphics display to show the amplified heart sounds as phonocardiograms and show parallel to the heart sounds the ECG in a standard Einthoven Triangle.
- 5. INTENDED USE: It is intended for use as a diagnostic aid as part of a physical assessment of a patient by health care professionals or other individuals trained to administer emergency first aid or otherwise care for a patient.
It can be used for the amplification of heart, lung, blood vessel, enteral and other body sounds. When the device's integral electrodes are placed on the chest of the patient, it is capable of verifying the existence of cardiac biopotential activity. The quality of the biopotential activity display depends upon user technique and environmental conditions and it is in no way meant to be diagnostic.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "CADltec" in a bold, sans-serif font. The letters "C", "A", "D", and "I" are capitalized, while "tec" is in lowercase. A horizontal line made of small squares runs beneath the word, adding a visual element to the design. The overall impression is a logo or branding element, possibly for a technology-related company.
CH-6343 Rotkreuz-Zug, Switzerland Phone xx41-41-79 80 120 Fax xx41-41-79 80 133
- SUBSTANTIAL EQUIVALENCE: The CADIscope, with it's sound channel, is substantially 6. equivalent to other electronic stethoscopes on the market. CADIscope claims for it's sound channel, substantial equivalence to the E-Scope Electronic Stethoscope Model 718-7120, K961301 manufactured by Cardionics Inc., 1100 Hercules Ave, Suite 210, Houston, 77058 TX.
The CADIscope, with it's ECG display, is substantially equivalent to other small portable electrocardiographs on the market. CADIscope claims for it's ECG display, substantial equivalence to the Miniscope MS-3, Electrocardiograph, K954066, manufactured by Schiller AG, Altgasse 68, CH-6340 Baar, Switzerland.
- COMPARISON TO PREDICATE DEVICE: The CADIscope is similar to the E-Scope in 7. that they both are electronic stethoscopes that operate on batteries. Both stethoscopes work by amplifying and filtering the heart sounds and allowing the user of the device to adjust the volume in order to better hear the heart sounds.
The CADIscope is different than the E-Scope in that it has a visual representation of the heart sounds (or the ECG) and the E-Scope does not. (And therefor the CADIscope contains a microprocessor and the E-Scope does not.)
The CADIscope is similar to the Miniscope MS-3 in that they both are battery operated portable monitors with 3 built-in electrodes to pick up the ECG in an Einthoven Triangle (or Tripode), in both devices the microprocessor optimizes the signal's amplitude and displays a 1-lead ECG on a LCD using similar resolution and base-line and noise filtering technology.
The CADIscope is different than the Miniscope MS-3 in that it does not have memory to store the (two) ECG signal's data over minutes to be played back on the internal LCD or to be transmitted to and analyzed by special ECG analysis programs on special cardiographs or PC's. The Miniscope MS-3 does also offer the possibility to connect an external patient cable with 3 or 5 leads, the CADIscope does not.
| | Predicate Devices | | New Device | |
|----------------|------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | Cardionics | Schiller | CADItec | |
| Characteristic | E-Scope<br>Electronic Stethoscope | Miniscope MS-3<br>Electrocardiograph | CADIscope<br>Electronic Stethoscope | |
| 1 Intended use | diagnostic aid, part of physical assessment of patient :<br>amplification of body sounds | diagnostic aid, part of physical<br>assessment of patient :<br>Display of ECG | diagnostic aid, part of physica<br>assessment of patient :<br>amplification of body sounds | |
| 2 Indications | any time an assessment of a<br>patient's physical condition is<br>required | any time an assessment of a<br>patient's physical condition is<br>required | any time an assessment of a<br>patient's physical condition is<br>required | |
| | | Predicate Devices | | New Device |
| | Characteristic | Cardionics<br>E-Scope<br>Electronic Stethoscope | Schiller<br>Miniscope MS-3<br>Electrocardiograph | CADItec<br>CADIscope<br>Electronic Stethoscope |
| 3 | Environment | anywhere | anywhere | anywhere |
| 4 | Personnel<br>requirements | individuals that have had<br>medical or first aid training<br>with a stethoscope, plus<br>teachers, students | physicians, nurses, or allied<br>health professionals and their<br>students | individuals that have had<br>medical or first aid training<br>with a stethoscope |
| 5 | Sound pickup | screw-on selectable :<br>- diaphragm flat shape<br>- bell shape | none | bell shape |
| 6 | Microphone | subminiature back-electret<br>microphone | none | subminiature back-electret<br>microphone |
| 7 | Sound<br>amplification | electronic, analog amplifier, 0<br>- 27 dB at 200 Hz, continuos | none | electronic: analog amplifier,<br>0-35 dB, 16 levels 5 dB each |
| 8 | Sound<br>Operating<br>modes | bell and diaphragm mode,<br>heart and breast sound filter | none | Organ specific :<br>heart, lungs, and entero filter |
| 9 | Sound<br>Frequency<br>response | 2 filters :<br>heart : 100-240 Hz<br>(-900Hz@ -20dB)<br>breast : 125-350 Hz.<br>(-2000Hz@ -20dB) | none | 3 filters :<br>heart: 15-600 Hz.<br>lungs: 300-1500 Hz.<br>Entero: 15-1500 Hz.<br>(@ -20 dB) |
| 10 | Sound speaker | 1 speaker, in control box | none | 2 earphones in the ear,<br>reduced noise induction into<br>binaurals |
| 11 | Binaurals | short metal tubes | none | none |
| 12 | On/Off controls | On switch<br>Off time controlled by<br>electronics | On switch<br>Off time controlled by<br>electronics | manual On<br>Off time controlled by<br>microprocessor |
| 13 | Mode / filter /<br>sound controls | mode switch > filters<br>analog potentiometer for<br>volume | none | mode switch > filters<br>+ / - switch for volume |
| 14 | Technology | discrete electronics,<br>integrated op amps | surface mount devices,<br>digital signal processor | surface mount devices,<br>mixed signal processor |
| 15 | Power<br>consumption | est 100 mW peak<br>est 10 mW normal | est 250 mW | 30 mW peak<br>5 uW sleep mode |
| 16 | Batteries | 1 x 9V, Alkaline, 45 gr<br>in control-box, 150 hours<br>operation = ½ year normal<br>operation | 2 x 1.5V, Alkaline, 20 grams<br>12 hours operation | 3 x 1.5V, Alkaline, 30 grams<br>in chestpiece, 160 hours<br>operation = ½ year normal<br>use |
| 17 | Dimensions | overall length 84 cm<br>total weight 250 grams | 77 x 126 x 20 mm<br>total weight 250 grams | cord length 64 cm<br>total weight 170 grams |
| 18 | Display | none | LCD 64 x 192 pixels,<br>graphics and status display | LCD 64 x 96 pixels,<br>graphics and status display |
| | | Predicate Devices | | New Device |
| | Characteristic | Cardionics<br>E-Scope<br>Electronic Stethoscope | Schiller<br>Miniscope MS-3<br>Electrocardiograph | CADItec<br>CADIscope<br>Electronic Stethoscope |
| 1<br>9 | Phonocardio-<br>gram<br>visualization of<br>sounds | no | no | yes, analog sound envelope<br>on built-in display, 4 time-<br>bases,<br>with freeze function |
| 2<br>0 | Display of<br>cardiac<br>biopotentials | no | Built-in three-point electrode<br>or 3 or 5 lead patient cable,<br>display 1 lead, with freeze and<br>playback | swiveling electrodes, forming<br>Einthoven triangle,<br>display time parallel to<br>hearing, with freeze function |
| 2<br>1 | Role of micro-<br>processor | none | switches modes, amplifi-<br>cation, displays status<br>controls | switches modes, amplifi-<br>cation, displays status<br>controls, displays and selects<br>optimum scales |
| 2<br>2 | Signal storage | no | 30 min 1 lead ECG | no |
| 2<br>3 | Signal trans-<br>mission to other<br>persons or to<br>PC | - option second headset<br>- option cable for<br>telemedicine over phone | Optional adapter with RS232<br>and analog output | Optional analog output to<br>hearing aides, second<br>earphone |
| 2<br>4 | Infection control | | | cleaning and disinfection<br>procedure in Operators'<br>Manual |
| 2<br>5 | Materials in:<br>chest contact | aluminum rings<br>polycarbonate diaphragm | AgCl plated electrodes | AgCl plated electrodes |
| 2<br>6 | Storage | | Polycarbonate cover | Optional cushion bag |
| 2<br>7 | Alarms | none | none | none |
| 2<br>8 | Status display | none | Status-line shows heart rate,<br>scale settings, battery low | status-line in display shows<br>mode, "good" signal, "direct"<br>heart rate, volume setting,<br>battery low |
| 2<br>9 | Use with<br>defibrillator | safety hazard | safety hazard | safety hazard |
| 3<br>0 | Source of user<br>hazards | defibrillator; bio-hazard | defibrillator; bio-hazard | defibrillator; bio-hazard |
| 3<br>1 | Patient hazards | bio-hazard | bio-hazard | bio-hazard |
| 3<br>2 | Environmental<br>hazards | none known | none known | none known |
| | | | | |
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for CADItec. The logo is in black and white and features the word "CADItec" in a stylized font. The letters "CADI" are in a larger, bolder font than the letters "tec". The "A" in "CADI" is stylized to look like a mountain peak. There is a dotted line underneath the word "CADItec".
CH-6343 Rotkrouz-Zug, Switzerland
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CADItec. The logo is in black and white, with the letters "CADItec" in a bold, sans-serif font. A jagged line runs underneath the letters. The logo is simple and modern.
CH-6343 Rotkreuz-Zug, Switzenland
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized caduceus or a human figure with outstretched arms.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
ਰ JUN
CADItec c/o Ms. Betty Lock Submission Correspondent Medline Industries, Inc. One Medline Place Mundelein, IL 60060
Re: K990809 CADIscope Electronic Stethoscope Regulatory Class: II (two) Product Code: DQD Dated: February 8, 1999 Received: March 11, 1999
Dear Ms. Lock:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food Drug and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) , it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions.
{5}------------------------------------------------
regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, 2Misbranding by reference to premarket notification? (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J.Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### INDICATIONS FOR USE 5.
Page 1 1 1
#### INDICATIONS FOR USE STATEMENT
510(k) Number : K980105
Device Name: CADIscope Electronic Stethoscope
Indications for Use:
The CADIscope is a patented electronic stethoscope with an integrated graphics display to show the amplified heart sounds as phonocardiograms and show parallel to the heart sounds the ECG in an Einthoven Triangle.
It is intended for use as a diagnostic aid as part of a physical assessment of a patient by health care professionals or other individuals trained to administer emergency first aid or otherwise care for a patient.
It can be used for the amplification of heart, lung, blood vessel, enteral and other body sounds. When the device's integral electrodes are placed on the chest of the patient, it is capable of verifying the existence of cardiac biopotential activity. The quality of the biopotential activity display depends upon user technique and environmental conditions and it is in no way meant to be diagnostic.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
V. Ruth Hillin
11 Carringscular No irologica Douges 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.