← Product Code [HOZ](/productcode/HOZ) · K990749

# SOF-SUCTION FLUID EVACUATION SYSTEM (K990749)

_X-Tech Medical, LLC · HOZ · Apr 2, 1999 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K990749

## Device Facts

- **Applicant:** X-Tech Medical, LLC
- **Product Code:** [HOZ](/productcode/HOZ.md)
- **Decision Date:** Apr 2, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.4790
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The device is used to continuously wick away irrigation fluid during eye surgery in an atraumatic manner since the suction does not reach the delicate tissues of the eye. The device can also be used in neurosurgery and other surgeries that cannot tolerate direct suction on tissue.

## Device Story

Device consists of polyvinyl acetal sponge tip heat-bonded to 18Ga 301 stainless steel cannula, attached to 5 Fr polyvinyl acetate suction tubing with standard adapter. Used in ophthalmic and neurosurgery to wick away irrigation fluid. Operates by continuous wicking: sponge absorbs fluid, which is then evacuated via suction through cannula and tubing. Prevents direct suction contact with delicate tissues, reducing trauma. Used by surgeons in clinical settings. Benefits include atraumatic fluid management and prevention of stagnant fluid accumulation.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Materials: polyvinyl acetal sponge, 301 stainless steel cannula, polyvinyl acetate suction tubing. Dimensions: 11mm x 7mm x 7mm sponge tip, 7mm 18Ga cannula, 36-inch tubing. Principle: passive wicking via sponge with active aspiration via suction tubing. Non-powered, manual assembly.

## Regulatory Identification

An ophthalmic sponge is a device that is an absorbant sponge, pad, or spear made of folded gauze, cotton, cellulose, or other material intended to absorb fluids from the operative field in ophthalmic surgery.

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Merocel Sponge Points
- Datex-Ohmeda Suction Tubing

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4/2/99

# X-TECH® MEDICAL

K9907499

#### 510(k) Summary

Date of Application: March 4, 1999

| Applicant:      | X-TECH® Medical                |
|-----------------|--------------------------------|
|                 | 111 Wild Oak Court             |
|                 | Danville, CA 94506             |
| Contact Person: | Edgar A. Rainin, MD, President |
| Phone:          | 925-736-5001                   |
| Fax:            | 925-736-8315                   |

- 1. TRADE NAME: SOF-SUCTION™ FLUID EVACUATION SYSTEM
- 2. COMMON NAMES: eye spears, sponge points & suction tubing
- 3. CLASSIFICATION NAME: sponge, ophthalmic & tube, aspirating, flexible, connecting
- 4. EQUIVALENT TO: Merocel Sponge Points and Datex-Ohmeda Suction Tubing.
- 5. DESCRIPTION OF DEVICE: Consists of a proximal portion consisting of a 11 mm x 7 mm x 7 mm polyvinyl acetal sponge tip fastened by heat bonding to a 7 mm 18Ga thin wall 301 stainless steel cannula friction attached to a 36'' 5 Fr polyvinyl acetate suction tubing, attached with medical grade adhesive to a ¼" standard suction tubing adapter. Included with the device is adhesive tape for fixating the suction tubing.
- 6. INTENDED USE: The device is used to continuously wick away irrigation fluid during eye surgery in an atraumatic manner since the suction does not reach the delicate tissues of the eye. The device can also be used in neurosurgery and other surgeries that cannot tolerate direct suction on tissue.
- 7. TECHNOLOGICAL CHARACTERISTICS: Our device consists of a combination of the Merocel Sponge Point and Suction Tubing. This results in continuous wicking by the sponge since the sponge is continuously emptied at its distal reservoir by aspirating the fluid from the reservoir. We are introducing no new materials. We are simply combining two proven technologies in a new, safe and effective manner.

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three lines forming the snake and staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 2 1099

Edgar A. Rainin, MD President X-TECH®MEDICAL 111 Wild Oak Court Danville, CA 94506

Re: K990749 Trade Name: Sof-Suction Fluid Evacuation System Regulatory Class: II Product Code: 86 HOZ Dated: March 4, 1999 Received: March 8, 1999

Dear Dr. Rainin:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{2}------------------------------------------------

Page 2 - Edgar A. Rainin, MD

ﺪ

· This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

A Roerl. Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health .......

{3}------------------------------------------------

Page 2a of 3

## 510(k) Number: K 990749

Device Name: Sof-Suction™ Fluid Evacuation System

### Indications For Use:

The Sof-Suction™ Fluid Evacuation System is used to remove excess irrigation fluids The Bot-Baction - Pland 24 assumed aring microscopic eye surgery to mitigate the effects n of the medial cannies of the eye aid as a see and to prevent stagnant irrigating fluid from entering the eye

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Laurel W.C. Brown, Ph.D
(Division Sign Off)

js

sion Sion-C ivision of Ophthalmic D 510(k) Number.

Prescription Use_v (Per 21 CFR 801.109).

---

**Source:** [https://fda.innolitics.com/device/K990749](https://fda.innolitics.com/device/K990749)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
