NUVOLASE 532 LASER SYSTEM

K990725 · American Laser Corp. · GEX · Mar 22, 1999 · General, Plastic Surgery

Device Facts

Record IDK990725
Device NameNUVOLASE 532 LASER SYSTEM
ApplicantAmerican Laser Corp.
Product CodeGEX · General, Plastic Surgery
Decision DateMar 22, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Nuvolase 532 Laser System for Ophthalmology is intended for use in retinal and macular photocoagulation and trabeculoplasty.

Device Story

Nuvolase 532 is a continuous-wave frequency-doubled diode-pumped Nd:YAG laser system. It uses 808 nm diodes to pump an Nd:YAG crystal, producing 1064 nm light, which is frequency-doubled to 532 nm green laser light. An electro-mechanical shutter controls exposure durations (0.05 to 1.0 seconds or continuous). The system includes a 670 nm red diode aiming beam. The device is used by ophthalmologists in clinical settings for retinal/macular photocoagulation and trabeculoplasty. The laser beam is delivered via fiber optic and laser slit lamp. The output provides therapeutic thermal energy to ocular tissues to treat specific retinal or macular conditions or to perform trabeculoplasty. The physician uses the slit lamp delivery system to visualize the target tissue and apply the laser energy, affecting clinical decision-making by enabling precise photocoagulation or trabecular meshwork modification.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Continuous-wave frequency-doubled diode-pumped Nd:YAG laser; 532 nm therapeutic wavelength; 670 nm red diode aiming beam; 50 mW to 1.5 W (up to 2.0 W) power output; electro-mechanical shuttering; fiber optic and slit lamp delivery; standalone system.

Indications for Use

Indicated for retinal and macular photocoagulation and trabeculoplasty in patients requiring ophthalmic laser treatment.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 3/22/99 Nu Nuvolase 532 Laser System for Opthalmology American Laser Medical, Inc. March 4, 1999 # K990725 # Summary of Safety and Effectiveness ### Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92 # Company Name/Contact: Daniel Hoefer American Laser Medical, Inc. 1832 South 3850 West Salt Lake City, UT 84104 (801) 972 1311, FAX (801) 972 5251 #### Name of Device: Trade Name: Nuvolase 532 Laser System for Opthalmology Common Name: Ophthalmic Laser Photocoagulator Classification Name: Ophthalmic Laser (per 21 CFR 886.4930) #### Predicate Devices: The Nuvolase 532 Laser System for Opthalmology has been modified from the American Laser Medical, Inc. Nuvolase 660 Laser System for Ophthalmology, K972561. ## Description of Device: The Nuvolase 532 Laser System for Ophthalmology is a continuous-wave frequency-doubled diode-pumped Nd:YAG laser system. Treatement beam power output for the system is 50 milliwatts to 1.5 watts at a wavelength of 532 nm. Depending on the delivery device efficiency, the maximum power level may be as much as 2.0 Watts CW. The aiming beam is provided by a red diode laser operating at 670 nm. Exposure durations for the Nuvolase 532 Laser System for Ophthalmology (in seconds) are 0.05, 0.1, 0.25, 0.5, 1.0, and continuous. Delivery of the beam occurs via fiber optic and laser slit lamp. {1}------------------------------------------------ Nuvolase 532 Laser System for Opthalmology American Laser Medical, Inc. March 4, 1999 #### Intended Use: The Nuvolase 532 Laser System for Ophthalmology is intended for use in retinal and macular photocoagulation and trabeculoplasty. ## Technological Characteristics/Device Comparison: The Nuvolase 532 Laser Sysem is a modification of the Nuvolase 660 Laser System for Ophthalmology, already in legal commercial distribution. Each of the systems optically pumps an Nd: Y AG crystal using 808 nm diodes to produce laser light at 1064 nm. This light passes through a second crystal which exhibits a non-linear optical response, re-emitting the laser energy at the first harmonic of the 1064 line, 532 nm greeen. The continuous wave beam is then shuttered electro-mechanically to produce the desired exposure durations. Each device is intended for retinal and macular photocoagulation and trabeculoplasty. The delivery system is the same in each case. ### Conclusion: The device modification does not affect the indications for use, materials, method of manufacture, or technology of the legally marketed device. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird with three wing-like shapes. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 2 1000 Mr. Daniel Hoefer Regulatory Affairs American Laser Medical, Inc. 1832 South 3850 West Salt Lake City, Utah 84104 Re: K990725 > Trade Name: Nuvolase 532 Laser System For Ophthalmology Regulatory Class: II Product Code: GEX Dated: March 4, 1999 Received: March 5, 1999 Dear Mr. Hoefer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ ### Page 2 - Mr. Daniel Hoefer This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. ~Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ - さ 990725 510(k) Number (if known): Device Name: NUVOLASE 532 IASER SYSTEM FOR OPHTHALMOLOGY Indications For Use: RETINAL AND MACULAR PHOTOCOAGULATION 1. TRABECULOPLASTY 2. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED! Concurrence of CDRH, Office of Device Evaluation (QDE) | (Division Sign-Off) | | |-----------------------------------------|---------| | Division of General Restorative Devices | | | 510(k) Number | L990725 | | Prescription Use | <span style="text-decoration: overline;">X</span> | OR | Over-The-Counter Use | __ | |----------------------|---------------------------------------------------|----|--------------------------|----| | (Per 21 CFR 801.109) | | | (Optional Format 1-2-96) | |
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%