CORDIA PALMAZ CORINTHIAN TRANSHEPATIC BILIARY STENT AND DELIVERY SYSTEM, MODEL PC125BPS, PC126BPS, PC127BPS,PC128BPS, PC
Applicant
Cordis Corp.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Apr 30, 1999
Decision
SESU
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is intended for use in the palliation of malignant neoplasms in the biliary tree.
Device Story
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is a balloon-expandable, stainless steel stent mounted on a balloon catheter delivery device. The system is advanced over a guidewire through a sheath lumen to an obstruction site within the biliary tree. Once positioned, the balloon is inflated to expand the stent, then deflated and removed, leaving the stent in place to maintain patency. The device is intended for single use and is provided sterile. It is operated by clinicians in a clinical setting to provide palliative treatment for malignant biliary neoplasms. The device benefits patients by relieving biliary obstructions caused by malignancies.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated via nonclinical design verification tests and analyses.
Technological Characteristics
Balloon-expandable stainless steel stent; mounted on balloon catheter delivery system. Stent lengths: 12-18 mm; expanded diameters: 5-8 mm. Sterile, single-use. Mechanical deployment via balloon inflation.
Indications for Use
Indicated for the palliation of malignant neoplasms in the biliary tree. Contraindicated for use in the vascular system.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
{1}------------------------------------------------
K990631
4/30/99
510(K) Summary of Safety and Effectiveness
Vage 1/3
| Submitter: | Cordis Corporation, a Johnson & Johnson Company<br>40 Technology Drive<br>Warren, New Jersey 07059 | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Telephone:<br>Fax: | (908) 755-8300<br>(908) 412-3915 |
| Contact Person: | Charles J. Ryan<br>Manager, Regulatory Affairs<br>Cordis Corporation, a Johnson & Johnson Company<br>40 Technology Drive<br>Warren, New Jersey 07059 | |
| | Telephone:<br>Fax: | (908) 412-7446<br>(908) 412-3915 |
| Date Prepared: | April 28, 1999 | |
| Trade Name: | Cordis PALMAZ CORINTHIAN Transhepatic Biliary<br>Stent and Delivery System | |
| Common Name: | Biliary Stent and Delivery Catheter | |
| Classification Name: | Biliary Catheter and Accessories (per 21 CFR 876.5010) | |
| Device Classification: | Class II | |
# Summary of Substantial Equivalence:
The design, matcrial, components, method of delivery, fundamental technology and I ne design, matchar, ochiponialer, ins PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System are substantially equivalent to those featured among Occh and Dollvory Bytiell are accosses 110(k) #K905720, #K911581, and #K964688).
In short, the Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery In short, the Cordis I Hillian Ocean the predicate Cordis PALMAZ Balloon-Expandable Stents (510(k) #K911581).
{2}------------------------------------------------
## Device Description:
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is a balloon-expandable, stainless steel stent that is provided mounted upon its balloon catheter delivery device. The Cordis PowerFlex Plus Balloon Catheters are used for the delivery of the Cordis PALMAZ CORINTHIAN Transhepatic Billiary Stant as to balloon catheters are substantially equivalent to the balloon catheter delivery devices used with the aforementioned predicate Cordis metal biliary stents.
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is advanced over a guidewire through a sheath lumen to an obstruction site in the biliary tree where the balloon is then inflated to expand the stent. After full expansion of the stent, the balloon is then deflated and removed
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is provided sterile and is intended for single use only.
#### Intended Use:
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is intended for use in the palliation of malignant neoplasms in the biliary tree.
## Technological Characteristics:
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System incorporates the same method of deployment, materials, fundamental technology and intended use as those found among predicate Cordis metal billiary stent products (see 510(k) #K905720, #K911581 and #K964688). The Cordis PALMAZ CORINTHIAN Transhepatic Billiary Stent and Delivery System incorporates a modified stent design that is substantially equivalent to that found with predicate Cordis metal biliary stents.
The Cordis PALMAZ CORINTHIAN Stent is provided in a range of nominal, unexpanded lengths from 12 to 18 mm and in a range of expanded diameters from 5 to 8 mm.
#### Performance Data:
The safety and effectiveness of the Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System have been demonstrated via data collected from nonclinical design verification tests and analyses.
{3}------------------------------------------------
A statement of substantial equivalence to another product is required by 21 CFR 807.87 and relates only to whether the product can be marketed without Prior reclassification or clinical approval. The proceet submission is therefore not related to the coverage of any patent and is not to be interpreted as an admission or used as evidence in a patent infringement lawsuit. As the commissioner of the stated, "A determination of substantial equivalence under the Federal Food, Drug and Cosmetic Act related to the fact that the product can be lawfully marketed without pre-market approval or reclassification. This determination is not intended to have any bearing whistsoever on the resolution of patent infringement suits." 42 Federal Register 42, 5
tage 3/3
1990631
{4}------------------------------------------------
Public Health Service
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and features an abstract image of an eagle. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged around the top half of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 3 0 1999
Mr. Chuck Ryan Manager, Regulatory Affairs Cordis, a Johnson & Johnson Company 40 Technology Drive Warren, New Jersey 07059
K990631 Re:
K990631
Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System Regulatory Class: II 21 CFR 876.5010 Product Code: 78 FGE Dated: March 31, 1999 Received: April 1, 1999
Dear Mr. Ryan:
We have reviewed your Section 510(k) notification of intent to market the device referenced We have reviewed your Secuon > IV(K) notincation of the indications for use indications for use above and we have determined the uevice is subsited in interstate commence
stated in the enclosure) to legally marketed prodical Parise Amendments, or to devices th stated in the enclosure) to legally marketed predical Device Amendments, or to devices that
prior to May 28, 1976, the enactment and wiss a stre Federal Hood Drug, and prior to May 28, 1976, the enactment use or the Provisions of the Federal Food, Drug, and have been reclassified in accordance with the provisions, subject to the general controls provisions Cosmetic Act (Act). You may, deletote, marked below. The general colorols provisions of the provisions of the Act and the inmultions described of devices, good manufacturing practice,
Act include requirements for annual registration, listing of devices, Act merude requirements against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this The Office of Device Evanuation has decemined the proposed labeling and that such use device will be used for an intended use hot tacititive in as p(i)(1)(E) of the Act, the following
could cause harm. Therefore, in accordance with Section in the Act, the fol Could Cause marm: " earlings section of the device's labeling:
The safety and effectiveness of this device for use in the vascular system have not been established.
Furthermore, the indication for biliary use must be prominently displayed in all labeling, Furthermore, the indication for othary use must be provincies and other promotional materials, in
including pouch, box, and carton labels, instructions for use, and other pro meluding pouch, 00%, and carton itsels, interest, and in bold print.
{5}------------------------------------------------
Page 2 - Mr. Chuck Ryan
If your device is classified (see above) into cither class II (Special Controls) or class III If your device is classified (Scc above) into onthols. Existing major regulations (Premarket Apploval), it may of subject to aden frontal Regulations, Title 21, Parts 800 to 895. affecting your device can or round intinues compliance with the Current Good A substantially Equivalent mintenents, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspections, the rood and Drug Trainmesult in regulatory action. In addition, FDA may publish comply with the GMT regulation may result in the Federal Register. Please note: this response to your premarks octification submission does not affect any obligation you might response to your promations and of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device The FDA Inding Of Substantial equivalence or your device to proceed to the market. This results in a classified.com for your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. I teast note that the above racomig manuse these limitations are modified in any way or removed from the device's labeling.
If you desire specific information about the application of other labeling requirements to your If you desire specific information as our of on in vitro diagnostic devices), please contact device (21 OF R Part 001 and 2011) 594-4616. Additionally, for questions on the promotion and the Office of Comphance at (201) 57 + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + + please note the regulation entitled, "Misbranding by reference to premarket notification" prease note the regalation on your responsibilities under the Act may be (21 Cr K 60). The Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Susan Abert, Ph.D., M.D.
. Ph.D., M.D. Susan A Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): K990631
Device Name: _Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System
FDA's Statement of the Indications For Use for device:
The Cordis PALMAZ CORINTHIAN Transhepatic Biliary Stent and Delivery System is indicated for the palliation of malignant neoplasms in the biliary tree.
Prescription Use V_ OR (Per 21 CFR 801.109)
Over-The-Counter Use_
David M. Heggem
Abdominal. 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.