Relieves minor muscle aches and pains . Relieve muscle spasm . Temporary improvement in local circulation . Temporarily reduces the appearance of cellulite . Relieves minor muscle aches and pains, relieves muscle spasm and temporarily improves local . circulation during Burn rehabilitation
Device Story
LPG Therapeutic Massager/Vibrator uses vacuum-assisted tissue mobilization to treat muscle aches, spasms, and cellulite; also used in burn rehabilitation. Device comprises main console with vacuum pump and computerized regulation system; motorized or auxiliary treatment heads with rollers create skin fold via suction (50-500 mBar). Motorized heads feature two rollers moving in rotation and sliding to mobilize tissue. Cyclic pulsation mode provides vibration by interrupting suction; frequency (0.41-19.23 Hz) and cycle rate (10-90% suction) are adjustable. Operated by practitioners (e.g., physical therapists) in clinical settings. Output is mechanical tissue stimulation; provider observes tissue reaction via endoscopic visualization or physical assessment. Benefits include improved local circulation, muscle relaxation, and scar/tissue mobilization for burn patients.
Clinical Evidence
Bench testing only. Clinical evidence includes literature citations regarding vibration effects on microcirculation, pain, and muscle spasm. One study by Dr. Gavroy compared burn victim treatment using LPG with rhythmicity versus LPG without rhythmicity.
Technological Characteristics
System includes main console with vacuum pump and computerized regulation. Treatment heads (5-12) feature motorized or non-motorized rollers. Suction range: 50-500 mBar. Cyclic pulsation: 0.41-19.23 Hz frequency; 10-90% cycle rate. Electrical safety: ground leakage <50 microamps. Patented motorized head (US 4,729,368) contains >200 parts. Connectivity: standalone console.
Indications for Use
Indicated for relief of minor muscle aches/pains, muscle spasm, temporary improvement of local circulation, and temporary reduction in appearance of cellulite. Also indicated for use during burn rehabilitation. Contraindicated for patients with known device sensitivity, open wounds, skin cancer in treated area, or HIV positive status.
Regulatory Classification
Identification
A therapeutic massager is an electrically powered device intended for medical purposes, such as to relieve minor muscle aches and pains.
{0}------------------------------------------------
K99 0445
JAN 18 2000
# 510 (k) Summary
Applicant's Name: LPG USA
Applicant's Address: 3101 North Federal Highway (Suite (301) Fort Lauderdale, FL 33306
Contact Person: Walter L. Wasserman
Phone Number/contact person: (954)568-5005 Fax Number/contact person: (954)568-6611
Address of manufacturing site: LPG Systems Technoparc de la Plaine 30 Rue du Docteur Abel BP 35-26902 Valence Cedex 9, France
Date Summary Prepared: January 18, 2000
This summary includes sections 2 through 8.
{1}------------------------------------------------
#### 2. Device Name
- Therapeutic Massager (21 CFR 890.5660) .
- Proprietary names LPG Therapeutic Massager Models ES1 .
- Therapeutic Massager/Vibrator (21 CFR 890.5660 and 890.5975) .
- Proprietary names LPG Therapeutic Massager/ Vibrator Models Cellu M6; ES/M60; S6; and LPG . Equine
#### 3. Establishment registration number
### 1062948
- Classification of device 4.
- . Class I
- Product codes 89 ISA and 89 IRO .
- . Panel - Physical medicine
- ડ . Performance standards
There are no performance standards established under section 514
#### 6. Labeling
# 6.1
# Indications for use
- Relieves minor muscle aches and pains ●
- . Relieve muscle spasm
- . Temporary improvement in local circulation
- . Temporarily reduces the appearance of cellulite
- Relieves minor muscle aches and pains, relieves muscle spasm and temporarily improves local . circulation during Burn rehabilitation
# Contraindications
- . Known sensitivity to the device
- Do not treat over open wounds .
- Skin cancer in the treated area .
- . Do not treat HIV positive patients
# Caution
Federal law restricts thus device for sale to or use under of a licensed physician
### Precautions;
When treating patients for burn rehabilitation care should be taken to carefully follow the instruction manual and not to exceed suction levels that would be obviously uncomfortable for the patient.
{2}------------------------------------------------
510(k) Number K99045 Device Name: LPG Therapeutic Massager/Vibrator
## Intended uses
- Relieves minor muscle aches and pains .
- Relieve muscle spasm ●
- . Temporarily Improve local circulation
- Temporarily reduces the appearance of cellulite .
- Relieves minor muscle aches and pains, relieves muscle spasm and improves local circulation during . Burn rehabilitation
# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use (Per 21 CFR 801.109) OR
1
Over-The-Counter-Use
(Optional Format 1-2-96)
6.2
{3}------------------------------------------------
#### 7. Substantial equivalence comparison
This product is substantially equivalent to a product already on the market LPG Therapeutic Massager (21 CFR 890.5660). The LPG technique involves a mobilization of the tissue accomplished with the LPG devices. All models with the exception of the ES1 and LPG Equine have both continuous and cyclic or pulsating modes of action. This pulsating function is equivalent to Therapeutic Vibrators (21 CFR 890.5975). And as such Substantial Equivalence is also claimed to the following legally marketed devices, which are also Class I.
- G5-GK1 K850851 .
- VIBRATOUCH™ II K912383 .
## Table of Comparison
| Device | Indications | Design | Specifications |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| LPG Massager/Vibrator | Minor muscle pain<br>Relieve muscle spasm; Temp.<br>improvement in appearance of<br>cellulite; Temp. increase of<br>local circulation; Burn Rehab. | 5 models<br>5 applicators<br>12 treatment heads<br>3 motorized heads | Ground current<br>Leakage <50<br>micamp. |
| G5-GK1 Vibrator | Minor muscle pain | 17 applicators<br>3 models<br>hand held | 12'x3'x3' - 3.8 lbs.; 0.25HP<br>110V, 60 cycle<br><75 micamp. ground<br>current leakage |
| VIBRATOUCH II | Minor muscle pain | 2 speeds; Battery Power | not specified |
#### Our objective is not to substantiate new claims but to prove that the exempt claims are appropriate for the burn victim and that the product is safe and effective for its general uses when used in burn rehabilitation.
{4}------------------------------------------------
8. 9. 9. 8. 9
### The LPG Technique:
The LPG technique involves mobilization of tissue accomplished with the LPG devices.
LPG stands for Louis Paul Guitay inventor of these therapeutic massagers.
LPG after a car accident that had left scars on his body was receiving treatments from a physical therapist using a tissue mobilization technique called Skin Fold Rolling. After a number of treatments, LPG noticed a difference in the Skin Fold Technique due to the practitioner performing the maneuvers or the time of the treatment. LPG had the idea to make the technique more consistent and replicable using vacuum power to lift the skin and make a fold that is mobilized by two motorized rollers.
#### The LPG devices:
Based on similar concepts, the LPG devices are intended for practitioners who work in specific fields but all are substantially equivalent devices:
- ES1 for Endermologie ●
- Cellu M6 for Endermologie and therapeutic applications ●
- S6 sport for sport applications, before, after exercise and rehabilitation after injuries .
- ES/M60 for therapeutic applications and cosmetic applications on the face or on small area .
- LPG Equine for animal treatment. �
These devices are substantially equivalent to the ES1 therapeutic massager because the mode of operation and operating parameters and specifications are basically the same.
Each LPG device comprises a main console housing a vacuum pump and a computerized regulation system.
All of them contain treatment heads (from 5 to 12), that are designed to accommodate all parts of the body. ES1, Cellu M6, S6, LPG Equine devices have a main motorized head, the other device contain auxiliary heads (non-motorized) only. The equine device head is the largest in order to be able to be applied on the horse's body surface.
The suction power of all these machines range from 50mBar to 500mBar which is represented by a scale from 1 to 10; for a specific machine and application, the suction power must be cyclically interrupted to provide vibration to the skin which is another stimulation for the tissues.
## The motorized head:
This patented head (number 4.729.368 US) contains more than 200 individual parts and contains two motorized rollers that move together and apart, at each end of the suction chamber. The two rollers are mobilized in rotation and in sliding back and forth. They provide, to the
{5}------------------------------------------------
and the comments
head, the capability of creating a fold of skin and subcutaneous fat with a very efficient contact, which allows better mobilization. This is also enhanced by specific maneuvers.
To visualize the reaction of tissue in-vivo, some endoscopic views were recorded within : the main head.
#### The auxiliary heads:
In addition to the main head, a set of auxiliary heads can be used with two non-motorized rollers on smaller parts of the body, particularly the face, the neck, and all the specific area, which need treatment. These heads are called regarding the size of the rollers: 15, 30, 44 millimeters; the ES/M60 device contains a larger head with specific ergonomic features (i.e. handle on the top of the head) but the same design. Recently, LPG had added small springs at each end of the rollers to provide lateral mobilization of them during the treatment.
### Rhythmicity or cyclic pulsation and mobilization:
For a specific machine and application (Cellu M6, S6, M60, and LPG Equine), the suction power is cyclically interrupted to provide vibration to the skin fold between the two rollers, which is another form of stimulation for tissues. Added to the device for the comfort of the patient, the cyclic suction power, according to scientific literature, provides additional effects to the tissue. Because of that, the therapeutic massager that includes vibration could be described as a form of therapeutic vibrator as well.
This cyclic pulsation feature is modulated by two variables:
#### . Cycle rate
The cycle rate regulates the proportion of time devoted to suction and rest. It is a ratio, the cycle rate ranges from 1 to 9, 1 for 10% suction versus 90% rest, 9 for 90% suction versus 10% rest, 5 for 50% suction versus 50% rest.
#### Frequency .
The frequency is the number of repetitions of one cycle per second. The frequency range from 0.41 to 19.23 Hz; these numbers appears on the screen as 0.0 for 0.41 Hz to 199 for 19.23 Hz.
In order to apply to the skin fold effective vibration, it is necessary to choose an adjustment with high frequency and low cycle rate or low frequency and high cycle rate. If the chosen adjustment is high frequency and high cycle rate, the skin fold is constantly lifted and the action is equivalent to the constant suction. If the chosen adjustment is low cycle rate and low frequency, the skin fold is mostly relaxed and therefore there will be less or no effectiveness.
This cyclic pulsation should not be considered as a single event but rather as repetitive stimulation that could produce harmonics (additional modes of stimulation). Thus the specific adjustments of the machine are less significant than it appears because of the harmonic summation.
{6}------------------------------------------------
The literature contains many citations showing the effectiveness of vibration of various frequencies; (i.e. has been shown on microcirculation, on pain and on muscle spasm).
LPG has investigated this area in one studies:
Doctor Gavroy compared treatment on burn victim with LPG and rhythmicity and LPG without rhythmicity.
# History of the LPG Technique:
A number of French medical doctors advised Louis Paul Guitay about using the device to treat low back pain, a condition very frequently associated with some types of change in the skin and subcutaneous fat. This is called secondary cellulitis on the back or around the pelvic bone (iliac crest). They also recommended treating neck ache with huge fatty localization around C7 (which is called buffalo hump).
Therapist successfully using the CelluM6 device on female patients suffering low back pain, also noticed that their patients were showing an improvement in the appearance of their cellulite or commonly called orange peel skin.
Results on the appearance of cellulite were so significant that LPG decided to market the product for this effect. The device was then sold for two applications (cosmetic and therapeutic).
When LPG decided to sale the LPG therapeutic Massager for its effect on cellulite, a specific methodology based on physiological and pathological data was designed:
- Global methodology involving torso massage designed to improve circulation at the । pelvic and abdominal level
- Tissue mobilization with specific maneuvers acting upon the venous system, lymphatic system, blood system which are sensitive to surrounding tissue movements
- Adjusted suction power to the sensitivity and the thickness of the tissues -
- -An average of fourteen sessions (two sessions per week) without interruption, during 35 to 45 minutes each.
Because this methodology represents a specific and innovative way to tackle this skin condition, the company decided to invent a name for it: the word "Endermologie" was chosen and became a trademark.
Thus, it was evident that Endermologie could be only done with the LPG Therapeutic Massagers because of the specificity of the patented heads.
To date the LPG devices such as therapeutic massagers have got the exempted claims: Relieves minor muscles aches and pain
Temporarily increases local blood circulation
Relaxes muscles spasm
Since, April 1998, the LPG therapeutic massager has been permitted the claim:
Temporary reduction in the appearance of cellulite.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three heads facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 18 2000
Mr. Frederic Barthe Vice President LPG USA, Inc. 3101 North Federal Highway Suite 301 Fort Lauderdale, Florida 33306
Re: K990445
Trade Name: LPG Therapeutic Massager Vibrator Regulatory Class: I Product Code: ISA Dated: October 18, 1999 Received: October 20, 1999
Dear Mr. Barthe:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{8}------------------------------------------------
Page 2 - Mr. Frederic Barthe
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Sincerely yours,
James E. Dillard III Acting Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
{9}------------------------------------------------
## 6.2
510(k) Number K990445 Device Name: LPG Therapeutic Massager/Vibrator
Intended uses
- · Relieves minor muscles aches and pains
- · Relieve muscle spasm
- · Temporarily improves local blood circulation
- · Relieves minor muscle aches and pains, relieves pains, relieves muscle spasm and improves local circulation during burn rehabilitation
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription use (Per 21 CFR 901.109
Over-the-Counter-Use
(Optional Format 1-2-96)
Acsilla
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K990445
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.