← Product Code [MAI](/productcode/MAI) · K990361

# HEADED BIO-ABSORBABLE CORKSCREW MODEL #AR-1921BU (K990361)

_Arthrex, Inc. · MAI · Apr 14, 1999 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K990361

## Device Facts

- **Applicant:** Arthrex, Inc.
- **Product Code:** [MAI](/productcode/MAI.md)
- **Decision Date:** Apr 14, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Headed Bio-Absorbable Corkscrew is intended for fixation of soft tissue to bone. This product is intended for the following indications: Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction. Midfoot Reconstruction, Metatarsal Ligament Repair Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligamant Reconstruction

## Device Story

Headed Bio-Absorbable Corkscrew is a threaded, bio-absorbable fastener used for soft tissue-to-bone fixation in orthopedic procedures. Device is manufactured from poly (L, DL-lactide). Surgeon prepares bone site using a tap before driving the anchor into place using an associated driver. Device provides mechanical fixation for various repairs including rotator cuff, Bankart, SLAP lesions, biceps tenodesis, and ligament reconstructions across multiple joints (shoulder, knee, ankle, wrist, elbow). By securing soft tissue to bone, the device facilitates healing and stabilization of the injured joint. Clinical benefit is derived from the bio-absorbable nature of the material, eliminating the need for secondary removal surgery.

## Clinical Evidence

Bench testing only. Pull-out strength testing was conducted using 20 lb. density foam blocks to simulate poor quality bone, comparing the subject device (79.3 lbs.) against predicates including FASTak (55.40 lbs.), Bio-Statak (69-71 lbs.), and Bio-Anchor (29-49 lbs.). No clinical data provided.

## Technological Characteristics

Material: Poly (L, DL-lactide). Design: Threaded fastener with slotted rounded head. Dimensions: 5.0mm. Insertion: Screw-in (requires pre-tapping of bone). Energy source: None (mechanical). Sterilization: Not specified.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- FASTak ([K971723](/device/K971723.md))
- Zimmer Bio-Statak (RSTA)
- Linvatec Bio-Anchor
- Ethicon Mitek 3.5mm Panalok Wedge Absorbable Suture Anchor

## Reference Devices

- Synthes Polypin

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# APR 1 4 1999

# 510(k) Summary

## 02/01/99

| Company: | Arthrex, Inc.                                            |
|----------|----------------------------------------------------------|
| Address: | 2885 S. Horseshoe Drive, Naples, FL 34104-               |
| Phone:   | (941) 643-5553                                           |
| Fax:     | (941) 643-6218                                           |
| Contact: | Vernon C. Brown<br>Manager Regulatory Affairs (ext. 117) |

| Trade Name:     | Arthrex Headed Bio-Absorbable Corkscrew        |
|-----------------|------------------------------------------------|
| Common Name:    | Bio-Absorbable Corkscrew Fastener              |
| Classification: | Fastener, Fixation, Biodegradable, Soft Tissue |

### Description:

The Headed Bio-Absorbable Corkscrew is manufactured using poly (L, DL-lactide). It is a threaded design with a rounded head which is slotted to accept the associated driver. Prior to driving in the anchor, it is necessary to prepare the bone using a tap.

#### Intended Use:

The Headed Bio-Absorbable Corkscrew is intended for fixation of soft tissue to bone. This product is intended for the following indications:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction. Midfoot Reconstruction, Metatarsal Ligament Repair

Knee: Medial Collateral Ligament Repair, Lateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligamant Reconstruction

## Substantial Equivalence:

Substantial equivalence is substantiated when a device has the same intended use and the same technological characteristics as the predicate device, or has the same intended use and different technological characteristics, but it can be demonstrated that the device is as safe and effective as the predicate device and does not raise different questions regarding safety and effectiveness

{1}------------------------------------------------

(121)

from the predicate device.

A substantial equivalence comparison is given in Table A.

In regards to the material. Poly (L.DL-Lactide), it is currently used in the Synthes Polypin which recently received FDA marketing clearance. This material has undergone extensive in-vitro and in-vivo testing by Synthes.

In addition to the testing performed by Synthes, an independent evaluation of the material was conducted by Claes et. al. ("New bioresorbable pin for the reduction of small bony fragments: design, mechanical properties and in vitro degradation" - Biomaterials, 1996, Vol. 17 No. 16).

None of the aforementioned differences make the Arthrex Headed Bio-Absorbable Corkscrew any less safe and effective than the predicate devices. Furthermore, they do not raise any different questions regarding safety and effectiveness from the predicate devices.

{2}------------------------------------------------

| Company  | Device                                                      | Intended Use                     | Material                       | Size(s)                  | Insertion             | Pull-Out                                                              |
|----------|-------------------------------------------------------------|----------------------------------|--------------------------------|--------------------------|-----------------------|-----------------------------------------------------------------------|
| Arthrex  | Headed Bio-<br>Absorbable<br>Corkscrew                      | Soft tissue to bone<br>fixation  | Poly (L, DL-lactide)           | 5.0mm                    | Screw in              | 79.3 lbs.*                                                            |
| Arthrex  | FASTak                                                      | Soft tissue to bone<br>fixation  | Titanium Alloy                 | 2.4mm<br>#2 suture       | Screw in              | 55.40 lbs.*<br>As reported: K971723                                   |
| Zimmer   | Bio-Statak<br>(RSTA)                                        | Soft tissue to bone<br>fixation  | Poly (L-lactic) acid           | 5.0mm<br>#2 suture       | Screw in              | Diaphyseal: 69 lbs. **<br>Metaphyseal: 71 lbs.<br>Cancellous: 70 lbs. |
| Linvatec | Bio-Anchor                                                  | Soft tissue to bone<br>fixation  | Poly (L-lactic) acid           | 3.5mm<br>#0 to #2 suture | Impact                | Diaphyseal: 33 lbs. **<br>Metaphyseal: 49 lbs.<br>Cancellous: 29 lbs. |
| Ethicon  | Mitek 3.5mm<br>Panalok Wedge<br>Absorbable<br>Suture Anchor | Soft tissue to bone<br>fixation. | Poly (L(-)-lactide)<br>polymer | 3.5mm                    | In tunnel<br>fixation | Unknown                                                               |

## Table A: Substantial Equivalence Comparison

* Pull-out testing performed using a 20 lb. density foam block to simulate poor quality bone

ર્દ્ર

** "Suture Anchors Product Information Guide" – Orthopedic Special Edition: Winter/Spring 1997

990361
38f3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with flowing lines representing hair or clothing.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 14 1999

Mr. Vernon C. Brown Requlatory Affairs Manager Arthrex, Inc. 2885 South Horseshoe Drive Naples, Florida 34104

Re: K990361 Headed Bio-Absorbable Corkscrew Trade Name: Regulatory Class: II Product Codes: MAI and HWC Dated: February 3, 1999 February 5, 1999 Received:

Dear Mr. Brown:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Mr. Vernon C. Brown

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Calia M. Whitten, Ph.D., M.

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

: :

The Headed Bio-Absorbable Corkscrew is intended for fixation of soft tissue to bone. This product is intended for the following indications:

Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction

Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair

Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis

Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction

(.

Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial Collateral Ligamant Reconstruction

(Divisi

Divis 510k; Number

Prescription Use
X
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K990361](https://fda.innolitics.com/device/K990361)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
