VEGA RETIC AUTOMATED HEMATOLOGY ANALYZER

K990311 · Abx Diagnostics · GKZ · Mar 31, 1999 · Hematology

Device Facts

Record IDK990311
Device NameVEGA RETIC AUTOMATED HEMATOLOGY ANALYZER
ApplicantAbx Diagnostics
Product CodeGKZ · Hematology
Decision DateMar 31, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.5220
Device ClassClass 2

Indications for Use

The VEGA RETIC Automated Hematology Analyzer is an automated hematology analyzer providing complete blood count (CBC), differential leucocyte count (DIFF) as well as reticulocyte count for in vitro diagnostic use in clinical laboratories. The VEGA RETIC Automated Hematology Analyzer provides appropriate flags and alarms to assist users in final checking and identification of abnormal cell populations.

Device Story

Benchtop clinical laboratory instrument; analyzes whole blood samples. Uses cytochemistry, focused flow impedance, light scattering, and fluorescence to perform complete blood counts, differential leucocyte counts, and reticulocyte counts. Microprocessor-driven system; provides flags and alarms for abnormal cell populations. Operated by clinical laboratory personnel. Output used by clinicians to assist in hematological assessment and identification of abnormal cell populations; aids in diagnostic decision-making regarding patient blood health.

Clinical Evidence

Clinical performance study comparing 200 human whole blood samples between VEGA RETIC and SYSMEX R-3000. Results showed high correlation for absolute/proportional reticulocyte and erythrocyte counts. Precision evaluated via within-run, between-run, and between-day testing using low, normal, and high control specimens, showing results similar to SYSMEX R-3000. Additional assessments included assay linearity, sample stability (up to 48-72 hours), and carry-over effect, confirming reliable performance.

Technological Characteristics

Benchtop automated hematology analyzer. Measurement principles: cytochemistry, focused flow impedance, light scattering, and fluorescence. Uses fluorescent dye and argon optical laser bench for reticulocyte analysis. Microprocessor-controlled. In vitro diagnostic use.

Indications for Use

Indicated for in vitro diagnostic use in clinical laboratories to provide complete blood counts, differential leucocyte counts, and reticulocyte counts. Assists users in identifying abnormal cell populations via flags and alarms.

Regulatory Classification

Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows a logo for ABX Diagnostics. The logo is a black square with the letters ABX in white. Above the logo is the date 3/31/99. The text "DIAGNOSTICS" is below the logo. K990311 ## Premarket Notification [510(k)] Summary Company: ABX Diagnostics, Inc. 34 Bunsen Drive Irvine, CA 92618 Phone: (949) 453-0500 FAX: (949) 453-0600 Contact Person: Franck Brouzes Date Prepared: January 18, 1999 #### Trade Name: VEGA RETIC Automated Hematology Analyzer | Common Name: | Automated cell counter and<br>Automated differential cell counter | |--------------|-------------------------------------------------------------------| |--------------|-------------------------------------------------------------------| | Classification Name: | Automated cell counter (864.5200) and<br>Automated differential cell counter (864.5220) | |----------------------|-----------------------------------------------------------------------------------------| |----------------------|-----------------------------------------------------------------------------------------| #### Substantial Equivalence: The VEGA RETIC Automated Hematology Analyzer is substantially equivalent to the following devices: - VEGA Hematology Analyzer (K962633, cleared to market November 4, 1996). . - SYSMEX R-3000 Automated Reticulocyte Analyzer (K912494, cleared for market . September 10, 1991). #### Description : The VEGA RETIC Automated Hematology Analyzer is a benchtop, clinical laboratory instrument which analyzes in-vitro samples of whole blood to provide complete blood count, leucocyte differential count and reticulocyte count using principles of cytochemistry, focused flow impedance, light scattering, and fluorescence. The instrument is microprocessor driven. {1}------------------------------------------------ Image /page/1/Picture/0 description: The image is a logo for a company called "ABX DIAGNOSTICS". The logo is a black square with the letters "ABX" in white. The letters are stylized and overlapping. The word "DIAGNOSTICS" is written in smaller letters below the square. The logo is simple and modern. #### Indications For Use: The VEGA RETIC Automated Hematology Analyzer is an automated hematology analyzer providing complete blood count, differential leucocyte count as well as reticulocyte count for in vitro diagnostic use in clinical laboratories. The VEGA RETIC Automated Hematology Analyzer provides appropriate flags and alarms to assist users in final checking and identification of abnormal cell populations. #### Comparison to Predicate Devices: VEGA RETIC Automated Hematology Analyzer is essentially identical to the VEGA Hematology Analyzer with respect to the indications for use for the hematological parameters for complete blood count and differential leucocyte count. The reagents and controls and the technological characteristics (cytochemistry, focused flow impedance and light scattering) are the same. VEGA RETIC Automated Hematology Analyzer is similar to the SYSMEX R-3000 with respect to the indications for use for the reticulocyte count. The reagents and controls are similar. The technological characteristics are similar: both systems use a fluorescent dye and argon optical laser bench to measure forward scattered light and forward fluorescence. #### Discussion of Performance Data: The determination of substantial equivalence is based on an assessment of clinical performance data comparing results from 200 human whole blood samples processed in the VEGA RETIC Automated Hematology Analyzer and in the SYSMEX R-3000 with respect to correlation between the two procedures. In addition the within-run, between run, and between-day precision of the VEGA RETIC Automated Hematology Analyzer was evaluate using three commercially available control specimens having low, normal and high reticulocyte counts. These results were compared to publicly available data for the SYSMEX R-3000. The following additional performance assessments were made: assay linearity, sample stability over time at ambient and low temperatures, and the carry-over effect from a high reticulocyte count sample to a low reticulocyte count sample. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a logo for ABX Diagnostics. The logo features the letters "ABX" in a stylized, interconnected design, with the "A" and "B" sharing a common vertical stroke. The letters are white against a black square background. Below the logo, the word "IAGNOSTICS" is written in a simple, sans-serif font. ### Conclusions: Performance data comparing the VEGA RETIC Automated Hematology Analyzer and the SYSMEX R-3000 demonstrated that the absolute/proportional reticulocyte counts and absolute erythrocyte counts generated on clinical specimens tested with each procedure are highly correlated. Additionally performance data demonstrated that the within-run, between run, and between-day precision characteristics of the two procedure are very similar. The VEGA RETIC Automated Hematology Analyzer provides linear absolute/proportional reticulocyte count data over a broad range of values. Performance data indicate that very little sample carry-over occurs and that samples stored up to 48 hours at 4°C provide reliable estimates of absolute/proportional reticulocyte counts and samples stored up to 72 hours at ambient or 4°C provide reliable estimates of erythrocyte counts. Prepared By: Patricia Amtower Consultant, ProMedica International {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of a bird-like figure, composed of three curved lines. The seal is black and white. Public Health Service MAR 3 1 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Franck Brouzes Vice President ABX Diagnostics, Inc. 34 Bunsen Drive Irvine, California 92618 Re: K990311 > Trade Name: ABX™ PENTRA 120 RETIC Automated Hematology Analyzer Regulatory Class: II Product Code: GKZ Dated: January 28, 1999 Received: February 1, 1999 Dear Mr. Brouzes: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # INDICATIONS FOR USE STATEMENT 510(k) Number (if known): K 990311 Device Name: VEGA RETIC Automated Hematology Analyzer Indications For Use: The VEGA RETIC Automated Hematology Analyzer is an automated hematology analyzer providing complete blood count (CBC), differential leucocyte count (DIFF) as well as reticulocyte count for in vitro diagnostic use in clinical laboratories. The VEGA RETIC Automated Hematology Analyzer provides appropriate flags and alarms to assist users in final checking and identification of abnormal cell populations. #### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) -Prescription Use (Per 21 CRFR 801.109) OR Over-The-Counter Use Peter E. Mhum (Division Sign-Off) Division of Clinical Laboratory Devices K990311 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...