MODULUS COMPATIBLE STABILITY (MCS)-HA TOTAL HIP SYSTEM
K990197 · Exactech, Inc. · MEH · Feb 4, 1999 · Orthopedic
Device Facts
Record ID
K990197
Device Name
MODULUS COMPATIBLE STABILITY (MCS)-HA TOTAL HIP SYSTEM
Applicant
Exactech, Inc.
Product Code
MEH · Orthopedic
Decision Date
Feb 4, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Exactech Hip Systems are also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.
Device Story
The Exactech MCS - HA Porous Coated Total Hip System is a prosthetic hip replacement device. It consists of a metallic substrate (Ti-6Al-4V or Co-Cr-Mo) with a hydroxyapatite (HA) coating. The device is implanted by a surgeon during total hip arthroplasty in a hospital setting. The HA coating is designed to facilitate fixation to the bone. The surgeon selects the appropriate system components based on patient requirements and clinical judgment. The device is intended to replace the hip joint to restore function and alleviate pain associated with degenerative conditions or fractures. It is not labeled for enhanced clinical or radiographic performance, enhanced fixation, or long-term stable fixation compared to conventional porous coated hip prostheses.
Clinical Evidence
Bench testing only. Testing characterized HA coating integrity, including interfacial shear strength on Ti-6Al-4V and Co-Cr-Mo substrates (34.5-44.8 MPa) and shear fatigue strength (13.8 MPa at 10 million cycles). Bonding of the beaded surface to the substrate was evaluated, showing interfacial shear strengths exceeding 27 N/sq.mm.
Technological Characteristics
Materials: Ti-6Al-4V and Co-Cr-Mo substrates with hydroxyapatite (HA) coating. Sterilization: Gamma irradiation (SAL 10^4) per AAMI guidelines. Design: Porous coated total hip prosthesis. No software or electronic components.
Indications for Use
Indicated for skeletally mature individuals undergoing primary total hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative hip problems, or proximal femoral fractures. Also indicated for revision of failed previous reconstructions with sufficient bone stock. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip joint control, or patients whose weight, age, or activity level suggests early system failure.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
Predicate Devices
Exactech MCS Total Hip System
Charnley
Zimmer Harris-Galante
Depuy AML
Osteonics Secur Fit HA
Smith & Nephew Synergy HA
OTI
Submission Summary (Full Text)
{1}------------------------------------------------
# K990197
# SUMMARY OF THE SAFETY AND EFFECTIVENESS INFORMATION FOR THE PREMARKET NOTIFICATION FOR THE MODULUS COMPATIBLE STABILITY (MCS) - HA TOTAL HIP SYSTEM
# Exactech, Inc. Established Registration Number 1038671
The Exactech MCS - HA Porous Coated Total Hip System is made of similar materials and is of a similar design to prostheses that were on the market before May 28, 1976. Additionally, this Porous Coated Total Hip System is of similar design to other components on the market that have been determined to be equivalent to devices on the market prior to May 28, 1976. These predicates include but are not limited to:
- 1. Exactech MCS Total Hip System
- 2. Charnley
- 3. Zimmer Harris-Galante
- 4. Depuy AML
- 5. Osteonics Secur Fit HA
- 6. Smith & Nephew Synergy HA
- 7. OTI
#### Material Integrity:
Overall integrity has been previously determined in the initial 510(k) submissions K921113 and K921114. HA coating performance characteristics are further defined in the FDA Master File MAF339 (BioCoat, Inc.) Data obtained meets or exceeds predicate device data. A summary is presented below.
#### Summary:
The evaluation of the interfacial shear strength of the uncoated sintered bead surface of the Exactech Porous System indicates excellent bonding of the beaded surface to the T1-6A1-4V substrate. Interfacial shear strengths in excess of 27 N/sq.mm can be expected with little variability based on the standard deviations observed for ten coating samples tested.
The integrity of the HA coating has been characterized. The manufacturer acknowledges the importance of coating quality to the success of the device and has collaborated with BioCoat, Inc., a well know coating specialty company, for the application and characterization of its coatings. (Additional information regarding the HA coating has been provided to the FDA as MAF339 by BioCoat)
The HA coating tensile strength on porous coated Ti alloy has been shown to be approximately 55 MPa. The shear strength of the HA on Ti-6Al-4V and Co-Cr-Mo solid
{2}------------------------------------------------
substrates has been shown to be between 5300 and 6500 psi. (36.6-44.8 MPa) while on porous coated Ti-6A1-4V and Co-Cr-Mo substrates the shear strength of the HA is between 5000 and 5900 psi. (34.5-40.7 MPa). Shear fatigue testing of the HA on Ti-6A1-4V shows a fatigue strength of 2000 psi. (13.8 MPa.) at 10 million cycles.
## Sterilization
The Exactech HA - MCS Total Hip System will be sterilized by gamma irradiation. The Sterility Assurance Level (SAL) is 104. Exactech utilizes Method 3, Protocol B from the "AAMI Guideline for gamma radiation sterilization" for the sterility dose setting and validation procedure.
## Utilization and Implantation
Selection of the Exactech HA - MCS Hip System depends on the judgment of the surgeon in relationship to the requirements of the patient. The surgeon should become thoroughly familiar with the technique of implantation by appropriate reading of the literature, and training in the operative skills and techniques required for total hip Arthoplasty surgery.
#### Indications
The Exactech HA - MCS Total Hip System is indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, osteonecrosis, congenital hip dysplasia, rhematoid arthritis, ankvlosing spondylitis and /or posttraumatic degenerative problems. It is also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.
### Contraindications
Use of the Exactech HA - MCS Total Hip System is contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure to the system.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white circular seal. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the seal is a stylized symbol consisting of three curved lines that resemble a person embracing another person.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1999 FEB
Gary J. Miller, Ph.D. Vice President of Research and Development Exactech® 2320 NW 66th Court Gainesville, Florida 32653
Re : K990197 Ha Coated MCS Total Hip System Requlatory Class: II Product Code: MEH Dated: January 15, 1999 Received: January 20, 1999
Dear Dr. Miller:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). This decision is based on consideration of the specific design of stem and coating composition detailed in this application. You may, therefore, market the device, subject to the general controls provisions of the Act and the following limitation:
You may not label or in any way promote these devices for "enhanced clinical or radiographic performance, enhanced fixation and/or long-term stable fixation." The data presented support equivalence with no additional claims over a conventional porous coated hip prosthesis.
Additional limitations for more specific claims of safety and effectiveness may be forthcoming. Should additional limitations be applied you will be contacted in writing to inform you of the additional labeling limitations.
The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{4}------------------------------------------------
Page 2 - Gary J. Miller, Ph.D.
You may market your device under the above limitations as class II devices. These devices would be considered not substantially equivalent to a legally marketed predicate device if labeled with other intended uses and/or claims of safety or effectiveness. Any other intended uses or claims may cause the device to be classified into Class III under Section 513 (f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing.
Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to A substantially equivalent determination assumes 895. compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Requlations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to Other general information on your responsibilities under the Act may be
{5}------------------------------------------------
Page 3 - Gary J. Miller, Ph.D.
obtained from the Division of Small Manufacturers Assistance obcathed from the bivibror. or 638-2041 or (301) 443-6597 or at at Its Coll-free namber (600)- 050-2811-02-28 11:52 PM dsmamain.html".
Sincerely yours,
Neef R.P. Oyler
Celia M. Witten, Ph.D., M . Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## "INDICATIONS FOR USE"
510(k) Number (if known) _ k990197 7 ___
Device Name:
Indications for Use:
All Exactech Hip Systems are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Exactech Hip Systems are also potentially indicated for revision of failed previous reconstruction's where sufficient bone stock is present.
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X
OR
Over the Counter Use
MRO
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number. K990197
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.