The M-11 UltraClave is intended to be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat- and moisturestable, reusable equipment. Dental handpieces can be sterilized in the M-11 in the Pouches cycle. This device is not recommended for sterilization of liquids intended for direct patient contact. Refer to Standard Cycle Parameters below for detailed information.
Device Story
Steam sterilizer using saturated steam at high pressure/temperature to kill infectious bio-organisms. Electric heating element generates steam within chamber. Electronic control system manages pre-programmed cycles (Unwrapped, Pouches, Liquids, Packs) based on time, temperature, and pressure. Operator selects cycle and initiates via switch; device automatically fills, heats, sterilizes, vents, and cools. Used in clinical/laboratory settings by healthcare staff. Output is sterilized equipment; ensures instrument sterility for patient safety. Dental handpiece sterilization validated via specific test protocol.
Clinical Evidence
Bench testing only. Validation protocol performed to evaluate the M-11's ability to sterilize commercially available dental handpieces as a worst-case challenge, confirming acceptable sterility assurance levels (SAL).
Technological Characteristics
Steam sterilizer; electric heating element; electronic control system. Operates via saturated steam at 121°C-132°C and 104-186 kPa. Cycles: Unwrapped, Pouches, Liquids, Packs. No differences from predicate device.
Indications for Use
Indicated for sterilization of heat- and moisture-stable, reusable equipment in medical/dental offices, hospitals, clinics, nursing homes, and laboratories. Includes dental handpieces. Not for liquids intended for direct patient contact.
Regulatory Classification
Identification
A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.
{0}------------------------------------------------
K 998189
h) 510(k) SUMMARY ・ The information required by 21 CFR § 807.92 is listed below.
| Submitter: | Midmark Corporation<br>60 Vista Drive<br>Versailles, OH 45380<br>(800) 633-0508 | | | | | | |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|------------------------------------------|---------|------------------|-----------------|--------------------------|
| Contact Person: | James Rile, Regulatory Coordinator<br>(937) 526-8797 | | | | | | |
| Date Submitted: | January 15, 1999 | | | | | | |
| Device Names: | <table><tr><td>Proprietary:</td><td>Midmark M-11 Ultraclave Steam Sterilizer</td></tr><tr><td>Common:</td><td>Steam Sterilizer</td></tr><tr><td>Classification:</td><td>80FLE: Sterilizer, Steam</td></tr></table> | Proprietary: | Midmark M-11 Ultraclave Steam Sterilizer | Common: | Steam Sterilizer | Classification: | 80FLE: Sterilizer, Steam |
| Proprietary: | Midmark M-11 Ultraclave Steam Sterilizer | | | | | | |
| Common: | Steam Sterilizer | | | | | | |
| Classification: | 80FLE: Sterilizer, Steam | | | | | | |
## Device to which the Midmark M-11 Ultraclave Steam Sterilizer with dental handpiece indications for use claims equivalence:
Midmark M-11 Ultraclave Steam Sterilizer, cleared for marketing 11/30/94 on K936181
### Device Description:
The Midmark M-11 Ultraclave Steam Sterilizer uses saturated steam at high pressures and temperatures and to kill infectious bio-organisms. This steam is generated inside the sterilization chamber by an electric heating element. The sterilizer's electronic control system is pre-programmed to complete sterilization cycles according to established time, temperature, and pressure parameters, summarized below.
| CYCLE | TEMPERATURE | PRESSURE | TIME |
|-----------|-----------------|------------------|------------|
| Unwrapped | 132° C (270° F) | 186 kPa (27 psi) | 3 Minutes |
| Pouches | 132° C (270° F) | 186 kPa (27 psi) | 15 Minutes |
| Liquids | 121° C (250° F) | 104 kPa (15 psi) | 30 Minutes |
| Packs | 121° C (250° F) | 104 kPa (15 psi) | 30 Minutes |
Items to be sterilized are placed in the sterilization chamber. The operator chooses a sterilization cycle, and presses the appropriate switch. The sterilizer automatically fills the chamber, heats the water into steam, sterilizes the items, and automatically vents the steam and cools the chamber after sterilization is complete.
### Intended Use:
The M-11 UltraClave is intended to be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat- and moisturestable, reusable equipment. Dental handpieces can be sterilized in the M-11 in the Pouches cycle. This device is not recommended for sterilization of liquids intended for
{1}------------------------------------------------
direct patient contact. Refer to Standard Cycle Parameters below for detailed information.
## Indications for Use:
# Standard Cycle Parameters
| Cycle | Temp/Press/Time<br>(minimums) | Items to be Sterilized<br>(Always consult the item manufacturer's recommendations for<br>sterilization.) |
|-----------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Unwrapped | 132°C (270°F)/<br>186 kPa (27 psi)<br>3 minutes | Instruments loose on a tray. Open glass or metal canisters. Tubing not used in surgical procedures. Items manufacturers recommend for exposure at 132°C (270°F) for 3 minutes. The sterility of unwrapped items is compromised on exposure to a non-sterile environment. |
| Pouches | 132°C (270°F)/<br>186 kPa (27 psi)<br>15 minutes | Instruments and dental handpieces. Loosely wrapped individual instruments. Multiple layers of instruments separated by fabric. Wrapped trays of loose instruments. Tubing not used in surgical procedures. Items manufacturers recommend for exposure at 132°C (270°F) for 15 minutes. |
| Liquids | 121°C (250°F)/<br>104 kPa (15 psi)<br>30 minutes | Liquids or gels that could boil over with rapid depressurization. Cycle includes special vent and dry functions. Not recommended for sterilization of liquids intended for direct patient contact. |
| Packs | 121°C (250°F)/<br>104 kPa (15 psi)<br>30 minutes | Textiles and surgical packs wrapped for sterilization. Items, except liquids, manufacturers recommend for exposure at 121°C (250°F) for 30 minutes. |
To validate sterility claims for dental handpieces, which were not included in the predicate devices indications for use when it was cleared by the FDA, Midmark developed a new test protocol for the M-11. This protocol was designed to evaluate the M-11's ability to sterilize commercially available dental handpieces, when presented as a worst-case challenge. The resulting testing showed that, when used in accordance with Midmark's directions for use, the M-11 can sterilize dental handpieces to an acceptable SAL.
## Technological Characteristics that differ between the Midmark M-11 Ultraclave Steam Sterilizer and its predicate device:
None. The intended use, principles of operation, physical design, control system, safeguards, energy sources, target population, materials of construction, method of manufacture, and performance are the same for the proposed device and its predicate.
(Division Sign-off)
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number
10
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 9 1999
Mr. James Rile Regulatory Coordinator Midmark Corporation 60 Vista Drive Versailles, Ohio 45380
Re : K990189 Modification Of: Midmark M-11 Ultraclave Trade Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: January 15, 1999 January 20, 1999 Received:
Dear Mr. Rile:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in Please note: this response to your the Federal Register. premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Rile
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
> Sincerely yours, liferan
Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# i) STATEMENT OF INDICATIONS FOR USE: KGG C 189
Midmark Corporation Applicant:
Midmark M-11 Ultraclave Steam Sterilizer Device Name:
Indications for Use:
The M-11 UltraClave is intended to be used in medical and dental offices, hospitals, clinics, nursing homes, laboratories, and other facilities to sterilize heat- and moisturestable, reusable equipment. Dental handpieces can be sterilized in the M-11 in the Pouches cycle. This device is not recommended for sterilization of liquids intended for direct patient contact. Refer to Standard Cycle Parameters below for detailed information.
| Cycle | Temp/Press/Time (minimums) | Items to be Sterilized (Always consult the item manufacturer's recommendations for sterilization.) |
|-----------|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Unwrapped | 132°C (270°F)/<br>186 kPa (27 psi)<br>3 minutes | Instruments loose on a tray. Open glass or metal canisters. Tubing not used in surgical procedures. Items manufacturers recommend for exposure at 132°C (270°F) for 3 minutes. The sterility of unwrapped items is compromised on exposure to a non-sterile environment. |
| Pouches | 132°C (270°F)/<br>186 kPa (27 psi)<br>15 minutes | Instruments and dental handpieces. Loosely wrapped individual instruments. Multiple layers of instruments separated by fabric. Wrapped trays of loose instruments. Tubing not used in surgical procedures. Items manufacturers recommend for exposure at 132°C (270°F) for 15 minutes. |
| Liquids | 121°C (250°F)/<br>104 kPa (15 psi)<br>30 minutes | Liquids or gels that could boil over with rapid depressurization. Cycle includes special vent and dry functions. Not recommended for sterilization of liquids intended for direct patient contact. |
| Packs | 121°C (250°F)/<br>104 kPa (15 psi)<br>30 minutes | Textiles and surgical packs wrapped for sterilization. Items, except liquids, manufacturers recommend for exposure at 121°C (250°F) for 30 minutes. |
Standard Cycle Parameters
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Chine S. Lin
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use X
Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.