BIPOLAR FORCEPS, KLEPPINGER INSERT, MODEL #'S 8393.741,8393.771,8394.741,8394.771
K990147 · Richard Wolf Medical Instruments Corp. · HIN · Apr 16, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K990147
Device Name
BIPOLAR FORCEPS, KLEPPINGER INSERT, MODEL #'S 8393.741,8393.771,8394.741,8394.771
Applicant
Richard Wolf Medical Instruments Corp.
Product Code
HIN · Obstetrics/Gynecology
Decision Date
Apr 16, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The coagulation of tubes, tissue coagulation and vascular coagulation, and to stop bleedings under endoscopic vision, including coagulation of the fallopian tube and the mesoalpinx for female sterilization.
Device Story
Modular bipolar forceps used for tissue/vascular coagulation and hemostasis during endoscopic procedures. Device features jaws for grasping/holding tissue and an integrated cleaning port for hygiene. Operated by surgeons in clinical/endoscopic settings. Provides precise axial holding and coagulation under endoscopic visualization. Benefits include improved site access and hygiene without requiring additional instruments. Device utilizes electrical energy for coagulation; design incorporates durable coated insulation on electrodes.
Clinical Evidence
No clinical data performed. Bench testing conducted per ANSI/AAMI HF18 and IEC601-2-2 standards for high-frequency devices. Biocompatibility testing performed on new materials.
Indicated for patients undergoing endoscopic procedures requiring tissue or vascular coagulation, including female sterilization via coagulation of the fallopian tube and mesoalpinx.
Regulatory Classification
Identification
A bipolar endoscopic coagulator-cutter is a device used to perform female sterilization and other operative procedures under endoscopic observation. It destroys tissue with high temperatures by directing a high frequency electrical current through tissue between two electrical contacts of a probe. This generic type of device may include the following accessories: an electrical generator, probes, and electrical cables.
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance 2/12/90 (K-90),” and
(iii) “Guidance (‘Guidelines’) for Evaluation of Laproscopic Bipolar and Thermal Coagulators (and Accessories),”
(2) International Electrotechnical Commission's IEC 60601-1-AM2 (1995-03), Amendment 2, “Medical Electrical Equipment—Part 1: General Requirements for Safety,”
(3) American National Standards Institute/American Association for Medical Instrumentation's HF-18, 1993, “Electrosurgical Devices,”
(4) Labeling:
(i) Indication: For female tubal sterilization, and
(ii) Instructions for use:
(A) Destroy at least 2 centimeters of the fallopian tubes,
(B) Use a cut or undampened sinusoidal waveform,
(C) Use a minimum power of 25 watts, and
(D) For devices with ammeters: continue electrode activation for 5 seconds after the visual endpoint (tissue blanching) is reached or current flow ceases indicating adequate tissue destruction.
{0}------------------------------------------------
4/16/99
1
■
353 Corporate Woods Vernon I lills, Illinois 6006 I Phone: 847.913.4143 Fax: \$47.943.4488
RICHARD W MEDICAL INSTRUMENTS CORPORA
Image /page/0/Picture/9 description: The image shows a black square with a white drawing inside. The drawing appears to be a stylized representation of an animal, possibly a dog or wolf, standing on its hind legs. The animal has a long neck and a pointed head, and it is wearing some sort of garment or harness around its body.
# 510(k) Summary of Safety and Effectiveness
| Submitter: | Date of Preparation: | | |
|----------------------------------------|-------------------------------------------------------|----------|--------------------|
| Company / Institution name: | January 13, 1999 | | |
| RICHARD WOLF MEDICAL INSTRUMENTS CORP. | FDA establishment registration number: | | |
| Division name (if applicable): | 1418479 | | |
| N.A. | Phone number (include area code): | | |
| | (847) 913-1113 | | |
| Street address: | FAX number (include area code): | | |
| 353 Corporate Woods Parkway | (847) 913-0924 | | |
| City: | State/Province: | Country: | ZIP / Postal Code: |
| Vernon Hills | Illinois | USA | 60061 |
| Contact name: | | | |
| Mr. Robert L. Casarsa | | | |
| Contact title: | | | |
| Quality Assurance Manager | | | |
| Product Information: | | | |
| Trade name: | Model number: | | |
| Bipolar forcep set (modular) | 8393.741, 8393.771, 8394.741, 8394.771 | | |
| Common name: | Classification name: | | |
| Modular Bipolar forceps | Coagulator-Cutter, Endoscopic,<br>Bipolar Accessories | | |
Information on devices to which substantial equivalence is claimed:
| 510(k) Number | Trade or proprietary or model name | Manufacturer |
|---------------|-----------------------------------------------|------------------------------------|
| 1 pre-enact. | 1 Kleppinger bipolar forceps 8383.21, 8384.21 | 1 Richard Wolf GmbH |
| 2 K981321 | 2 Modular Bipolar Forcep Sets 8390/91/93/94 | 2 Richard Wolf Medical Instruments |
### 1.0 Description
The modular bipolar forcep set provides precise grasping and axial holding placement of the jaws, performs thorough hygiene, and site irrigation can be achieved without using additional instruments.
{1}------------------------------------------------
Image /page/1/Picture/3 description: The image contains a single Chinese character, '得', in a bold, sans-serif font. The character is black against a white background. The character appears to be centered in the image.
#### 2.0 Intended Use
The coagulation of tubes, tissue coagulation and vascular coagulation, and to stop bleedings under endoscopic vision, including coagulation of the fallopian tube and the mesoalpinx for female sterilization.
#### 3.0 Technological Characteristics
There are no significant technological characteristic changes to the new devices compared to the pre-enactment devices. The insulation of the electrodes is changed to a more durable coated material. A cleaning port is added. The instruments are identical to those cleared by K981321.
#### 4.0 Substantial Equivalence
The submitted devices pose the same type of questions about safety or effectiveness as the compared devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing pre-enactment devices sold by Richard Wolf. In addition, the submitted devices are substantially equivalent to devices sold by various competitors.
#### 5.0 Performance Data
No known FDA performance standard exists.
The bipolar forceps were tested to meet the appropriate sections of the ANSV AAMI standard on high frequency devices HF18 and IEC601-2-2, Device materials tests were performed to assure biocompatibility of the new material. The tests indicated that no irrigation would occur.
#### 6.0 Clinical Tests
No special clinical tests performed.
#### 7.0 Conclusions Drawn
These devices are designed and tested to guarantee the safety and effectiveness, when used according to the instructions manual.
By: Robert L. Pearson
Date: Apr 16, 99
Robert L. Casarsa Quality Assurance Manager
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The text is arranged in a circular fashion around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 16 1999
Mr. Robert L. Casarsa Quality Assurance Manager RICHARD WOLF MEDICAL INSTRUMENTS CORP. 353 Corporate Woods Parkway Vernon Hills, IL 60061
Re: K990147 Modular Bipolar Forcep Set, Kleppinger Principle Dated: January 13, 1999 Received: January 19, 1999 Regulatory Class: III 21 CFR 884.4150/Procode: 85 HIN
Dear Mr. Casarsa:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## * (*) : 10) :
510(k) Number (if known):
Modular Bipolar Forcep Set. Kleppinger Principle Device Name:__________________________________________________________________________________________________________________________________________________________________
### Intended Use:
The coagulation of tubes, tissue coagulation and vasculation, and to stop bleedings under endoscopic vision, including coagulation of the fallopian tube and the mesoalpinx for female sterilization.
(IT EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOXITIER PAGE IF NEEDED)
Prescription Use
Per 21 CFR 801.109
---
**(Division Sign-Off)**
and Radiological Devices
510(k) Number_
Division of Reproductive, Abdominal, ENT,
()R
Over-The Counter_
5 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.