PACIFIC HEMOSTASIS IMMUNODEPLETED FACTOR VIII DEFICIENT PLASMA

K990046 · Pacific Hemostasis · GJT · Feb 17, 1999 · Hematology

Device Facts

Record IDK990046
Device NamePACIFIC HEMOSTASIS IMMUNODEPLETED FACTOR VIII DEFICIENT PLASMA
ApplicantPacific Hemostasis
Product CodeGJT · Hematology
Decision DateFeb 17, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.7290
Device ClassClass 2

Indications for Use

Pacific Hemostasis Immunodepleted Factor VIII (FVIII) Deficient Plasma is intended for use as a substrate in the quantitative determination of Factor VIII activity in citrated plasma.

Device Story

Pacific Hemostasis Immunodepleted Factor VIII Deficient Plasma is a lyophilized preparation of fresh human plasma depleted of Factor VIII via immobilized specific antibodies. Used in clinical laboratories by technicians/pathologists to quantify Factor VIII activity in patient plasma. The device acts as a substrate in an Activated Partial Thromboplastin Time (APTT) assay; the patient's plasma is mixed with the deficient plasma, and the resulting APTT correction is measured. The degree of correction is compared against a reference plasma with known Factor VIII activity to calculate the patient's Factor VIII levels. This quantitative measurement assists clinicians in diagnosing or monitoring coagulation disorders related to Factor VIII deficiency. The product is provided in 1.0mL vials.

Clinical Evidence

Bench testing only. Performance evaluated by comparing standard curves and Factor VIII recovery in control plasmas against the predicate device over 10 days. Correlation coefficient for recovered FVIII activity was 0.98 (y = 0.9911x + 0.8692). Combined data across multiple instruments (Amelung KC 4 A, MLA-700, MLA-1000C, ACL-3000PUS) yielded a correlation coefficient of 0.99 (y = 1.037x + 0.5603). Reconstituted stability studies confirmed no clinically significant differences between fresh and 8-hour aged plasma.

Technological Characteristics

Lyophilized human plasma; Factor VIII depleted via immobilized antibodies; Factor VIII activity <1%; other coagulation factors within normal range. Provided in 1.0mL vials. Compatible with standard clinical coagulation analyzers (e.g., Amelung KC 4 A, MLA-700, MLA-1000C, ACL-3000PUS).

Indications for Use

Indicated for use in clinical laboratories as a substrate for the quantitative measurement of Factor VIII activity in citrated human plasma samples.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ 2/17/99 # Premarket Notification 510 Factor VIII Immunodepieted Plasma | 7.0 | PREMARKET NOTIFICATION 510(K) SUMMARY | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------| | Applicant: | Laura A. Worfolk, Ph.D.<br>Pacific Hemostasis<br>11515 Vanstory Drive<br>Huntersville, NC 28078<br>704-875-0494<br>Fax # 704-875-2092 | | Contact: | Mark Ellis, Regulatory Affairs Manager<br>704-948-3279<br>Fax # 704-875-2092 | | Date: | January 6, 1999 | | Trade Name: | Pacific Hemostasis Immunodepleted Factor VIII Deficient Plasma | | Common Name: | Factor VIII Deficient Plasma | | Classification Name: | Plasma, Coagulation Factor Deficient<br>(per 21 CFR section 864.7290) | | Comparison Device: | Biopool Factor VIII Deficient Plasma, K893525 | #### Description of Immunodepleted Factor VIII Deficient Plasma Pacific Hemostasis (PH) Immunodepleted Factor VIII Deficient Plasma is a lyophilized preparation of fresh human plasma with added stabilizer. The product is prepared from pooled normal citrated plasma, and then depleted of FVIII by immobilized highly specific antibodies. Factor VIII activity is less than 1%, all other coagulation Factors are within the normal range. Each unit of source material used in the preparation of this product has been tested and found negative for HBsAg (Hepatitus B Surface antigen) and negative for antibodies to HIV and HCV. The product is provided in 1.0mL vials, 10 vials per package. #### Intended Use of Immunodepleted Factor VIII Deficient Plasma This product is intended for use in a clinical laboratory for the quantitative measurement of Factor VIII activity. Factor VIII activity in patient or control plasma is assayed by the amount of Activated Partial Thromboplastin Time (APTT) correction produced by the test plasma when mixed with Factor VIII deficient plasma. Results are compared to the degree of APTT correction of a reference plasma with known Factor VIII activity. {2}------------------------------------------------ #### Summary of Performance Data for Substantial Equivalence Comparisons Pacific Hemostasis (PH) Immunodepleted Factor VIII Deficient Plasma was compared to Biopool (BP) Factor VIII Deficient Plasma. Both products are lyophilized preparations of normal human plasma, the PH product contains added stabilizer. The Factor VIII level in both is less than 1%; all other coagulation factors are within the normal range. The intended use for both products is identical; for the quantitative measurement of Factor VIII activity in patient plasma. Biopool FVIII deficient plasma may also be used as a negative control in von Willebrand Factor assays, however this claim is not made for the PH product. Comparison studies performed for this application support the claim of substantial equivalence. Dav-to-day precision studies were performed by preparing standard curves using both plasmas, and measuring FVIII activity contained in control plasmas over a 10 day period. The control plasmas tested contained FVIII activity levels ranging from normal to markedly abnormal low. The standard curves obtained using each plasma were indistinguishable over the 10 day testing period (R2 = 0.98-0.99). Further, recovery of Factor VIII activity contained in six control plasmas was equivalent for both (Table 1). The correlation coefficient for recovered FVIII activity contained in the control plasmas was 0.98, with a regression line equation of y = 0.9911x + 0.8692. Table 1. Recovery of FVIII Activity (%) Contained in Control Samples. Testing with PH or BP FVIII Deficient Substrate Plasmas** | | PNP* | | Control 1 | | Control 2 | | Control 3 | | Control 4 | | Control 5 | | |------------------------|-------|-------|-----------|-------|-----------|------|-----------|------|-----------|------|-----------|------| | | PH | BP | PH | BP | PH | BP | PH | BP | PH | BP | PH | BP | | mean | 114.7 | 117.0 | 130.9 | 127.1 | 95.5 | 96.6 | 65.6 | 65.7 | 33.6 | 34.1 | 18.0 | 18.1 | | 1 SD | 16.1 | 21.29 | 6.46 | 8.73 | 4.63 | 5.35 | 1.82 | 4.82 | 2.13 | 1.57 | 1.32 | 0.98 | | % CV | 14.0% | 18.2% | 4.9% | 6.9% | 4.8% | 5.5% | 2.8% | 7.3% | 6.3% | 4.6% | 7.3% | 5.4% | | n (days of<br>testing) | 10 | 10 | 8 | 8 | 8 | 8 | 10 | 10 | 8 | 8 | 8 | 8 | *PNP = pooled normal plasma, ** Testing on the MLA - 1000C™. Several different instruments were used to assess the performance of the deficient plasmas, these included the Amelung KC 4 A™, MLA®-700, MLA®-1000C™ and the ACL-3000PUS. Standard curves were prepared, and the recovery of FVIII activity contained in control plasmas was determined using both products. The combined instrument data vielded a correlation coefficient of 0.99, with a regression line equation of y = 1.037x + 0.5603. Last, reconstituted stability studies were performed. Standard curves prepared with fresh or 8-hour aged deficient plasma were indistinguishable for both products. Further, there were no clinically significant differences observed in FVIII activity recovered in control plasmas between fresh and aged deficient plasmas, for PH and Biopool substrate plasmas, {3}------------------------------------------------ In summary, the similar intended use, technological characteristics and combined performance data support the substantial equivalence claim for Pacific Hemostasis Immunodepleted Factor VIII Deficient Plasma to Biopool Factor VIII Deficient Plasma. Therefore based on the data provided, it is our conclusion that Pacific Hemostasis Immunodepleted Factor VIII Deficient Plasma is substantially equivalent to Biopool Factor VIII Deficient Plasma. / {4}------------------------------------------------ ## PREMARKET NOTIFICATION TRUTHFUL AND ACCURATE STATEMENT [As required by 21 CFR 807.87(j)] I certify that, in my capacity as a Research Scientist at Pacific Hemostasis, a Fisher Scientific Company, I believe to the best of my knowledge, that all data and information submitted in the premarket notification are truthful and accurate and that no material fact has been omitted. Laura A. Worfolk 1/6/99 **Laura A. Worfolk, Ph.D.** K990046 *(Premarket Notification [510(k)] Number) {5}------------------------------------------------ FEB 1 7 1999 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Mr. Mark Ellis Regulatory Affairs Manager Pacific Hemostasis 11515 Vanstory Drive, Suite 125 Huntersville, North Carolina 28078-8144 > K990046 Trade Name: Pacific Hemostasis Immunodepleted Factor VIII Deficient Plasma Regulatory Class: II Product Code: GJT Dated: January 6, 1999 Received: January 7, 1999 Dear Mr. Ellis: Re: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {6}------------------------------------------------ ### Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Butman Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ Page_ of I 510(k) Number (if known): _ (99004/2 Device Name:_ Indications For Use: ### Statement of Indications for Use Pacific Hemostasis Immunodepleted Factor VIII (FVIII) Deficient Plasma is intended for use as a substrate in the quantitative determination of Factor VIII activity in citrated plasma. יי הייני היינו Peter E. Madlem (Division Sign-Off) Division of Clinical Laboratory Devices K190046 510(k) Number (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **Prescription Use** (Per 21 CFR 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...