← Product Code [FBP](/productcode/FBP) · K984607

# PANOVIEW PLUS OPTIC WITH 2.7MM DIAMETER, MODEL 8672.XXX, PANOVIEW PLUS OPTIC WITH 1.9MM DIAMETER, MODEL 8686.XXX (K984607)

_Richard Wolf Medical Instruments Corp. · FBP · Mar 5, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K984607

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corp.
- **Product Code:** [FBP](/productcode/FBP.md)
- **Decision Date:** Mar 5, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The endoscopes serve to visualize the inside of the patient via natural or surgically generated access.

## Device Story

Thin endoscopes (1.9 mm and 2.7 mm diameter) designed for visualization in infants and babies. Device provides optical access to internal body cavities via natural or surgical openings. Operated by trained clinicians in various medical disciplines (surgery, urology, gynecology, ENT). Used in conjunction with external light sources, flexible light cables, video/reflex cameras, objective lenses, and endoscopic accessories (trocar sleeves, forceps, electrodes). Provides enhanced image size and brightness compared to previous models. Facilitates clinical diagnosis and therapy by allowing direct visualization of internal anatomy.

## Clinical Evidence

No clinical data. Bench testing only.

## Technological Characteristics

Rigid endoscopes with 1.9 mm and 2.7 mm diameters. Features Panoview Plus optics for increased image size and brightness. Autoclavable at 134°C. Non-electronic, optical device.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Cysto-Urethrosocpe for children, Model 8670 and 8680 (pre-enactment)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

3/5/99

g 1 of 2

553 Corporate Woods Vernon I lills, Illinois 6006 i Phone: 847.913.4443 Fax: \$47.943.1488

510(k) Summary of Safety and Effectiveness

RICHARD WOLF

| Date of Preparation | December 23, 1998 |
|---------------------|-------------------|
|---------------------|-------------------|

| Submitter:                             |                                        |
|----------------------------------------|----------------------------------------|
| Company / Institution name:            | RICHARD WOLF MEDICAL INSTRUMENTS CORP. |
| FDA establishment registration number: | 14 184 79                              |
| Division name (if applicable):         | N.A.                                   |
| Phone number (include area code):      | (847) 913-1113                         |
| Street address:                        | 353 Corporate Woods Parkway            |
| FAX number (include area code):        | (847) 913-0924                         |
| City:                                  | Vernon Hills                           |
| State/Province:                        | Illinois                               |
| Country:                               | USA                                    |
| ZIP / Postal Code:                     | 60061                                  |
| Contact name:                          | Mr. Robert L. Casarsa                  |
| Contact title:                         | Quality Assurance Manager              |

Product Information:

| Trade name:          | Endoscopes with Panoview Plus Optics |
|----------------------|--------------------------------------|
| Model number:        | 8672.xxx and 8686.xxx                |
| Common name:         | Endoscopes                           |
| Classification name: | Endoscopes                           |

Information on devices to which substantial equivalence is claimed:

|   | 510(k) Number | Trade or proprietary or model name                   | Manufacturer |
|---|---------------|------------------------------------------------------|--------------|
| 1 | pre-enactment | Cysto-Urethrosocpe for children, Model 8670 and 8680 | Richard Wolf |
| 2 |               | 2                                                    | 2            |

#### 1.0 Description

An endoscope with 1.9 mm and 2.7 mm diameter, typically used in endoscopy in infants and babies.

### 2.0 Intended Use

The endoscopes serve to visualize the inside of the patient via natural or surgically generated access.

{1}------------------------------------------------

K 984607 - pg r of r

#### Technological Characteristics 3.0

- · increased image size and greater brightness
- sharp, brilliant quality over the entire image .
- autoclavable 134°C / 273°C .

### 4.0 Substantial Equivalence

The submitted devices pose the same type of questions about safety or effectiveness as the compared devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing devices sold by Richard Wolf.

#### 5.0 Performance Data

No performance standards are known. The devices conform to the relevant provisions of European Device Directive 93/42/EEC.

#### 6.0 Clinical Tests

Clinical tests performed were not performed.

### 7.0 Conclusions Drawn

These devices are designed and tested to assure their safety and effectiveness when used according to the instructions manual.

By: Robert L. Casares

Robert L. Casarsa Quality Assurance Manager

Date: Dec 23 98

{2}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 5 1999

Mr. Robert L. Casarsa Quality Assurance Manager RICHARD WOLF Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, IL 60061

Re: K984607

Thin Endoscopes with Panoview Plus Optics Dated: December 23, 1998 Received: December 28, 1998 Regulatory Class: II 21 CFR 876.1500/Procode: 78 FBP & GCM

Dear Mr. Casarsa:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

APT Daniel G. Schultz. M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

**Indications for Use**

510(k) Number (if known): __ K 98 46 07

Device Name: Thin Endoscopes with Panoview Plus Optics

# Intended Use:

The endoscopes serve to visualize the inside of the patient via natural or surgically generated access.

### Indications and Fields of Application:

For examination, diagnosis and/or therapy by personnel trained in the use of endoscopic instrumentation used in various medical disciplines, such as surgery, urology, gynecology, and ENT.

### Contraindications:

There are no known contradindications directly related to the product. The attending physician must determine the appropriateness of the application while considering the general condition of the patient.

## Combinations:

The endoscopes are used in connection with light sources and flexible light cables, video cameras or reflex cameras and objective lenses, as well as accessories for endoscopic use, e.g. trocar sleeves, forceps, electrodes.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devic 510(k) Number

Prescription Use
Per 21 CFR 801.109

5 - 1

Over-The Counter

---

**Source:** [https://fda.innolitics.com/device/K984607](https://fda.innolitics.com/device/K984607)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).