FRESENIUS ATR40 AND ATR120 AUTOTRANSFUSION RESERVOIRS
K984586 · Fresenius Hemotechnology, Inc. · DTN · Jun 29, 1999 · Cardiovascular
Device Facts
Record ID
K984586
Device Name
FRESENIUS ATR40 AND ATR120 AUTOTRANSFUSION RESERVOIRS
Applicant
Fresenius Hemotechnology, Inc.
Product Code
DTN · Cardiovascular
Decision Date
Jun 29, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4400
Device Class
Class 2
Indications for Use
The ATR 40 and ATR 120 Autotransfusion Reservoirs are indicated as autotransfusion reservoirs to collect, defoam, filter, and store blood prior to processing.
Device Story
The Fresenius ATR 40 and ATR 120 are 3-liter sterile, disposable blood collection reservoirs. They function by applying vacuum suction to aspirate blood from an operative field into the reservoir lumen. The device features inlet connectors for suction lines and a filter holder containing either a 40-micron (ATR 40) or 120-micron (ATR 120) multi-layer filter. The filter media is treated with a defoaming agent to dissolve air bubbles introduced during collection. A pressure relief valve maintains internal pressure between -300 mmHg and +100 mmHg. Collected blood is stored in the reservoir and can be transferred to an autotransfusion device via an outlet line. The device is used in clinical surgical settings, operated by healthcare professionals. It benefits patients by enabling the collection and processing of autologous blood for reinfusion, reducing the need for donor blood.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility per modified ISO standards, structural integrity, shelf-life validation, and comparative performance testing against predicate devices. Results demonstrated equivalent filtration efficiency, defoaming capability, hemolysis levels, and cellular component retention.
Technological Characteristics
3-liter rigid plastic reservoir; vacuum-operated; multi-layer filter (40-micron for ATR 40, 120-micron for ATR 120) with defoaming agent; pressure relief valve (-300 to +100 mmHg); sterile (ETO); disposable; non-pyrogenic fluid path.
Indications for Use
Indicated for use as autotransfusion reservoirs to collect, defoam, filter, and store blood collected intra- or postoperatively in surgical procedures prior to processing. Not for use as a cardiotomy reservoir.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
### X. 510(k) Summary
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The contents of this 510(k) summary have been provided in conformance with 21 CFR \$807.3 (Federal Register, Vol. 59, No. 239, Dec. 14, 1994, pg. 64295).
Premarket Notification for the Fresenius ATR 40 and ATR 120 Autotransfusion Reservoirs: 106 Section X: 510(k) Summary of Safety and Effectiveness Information
{1}------------------------------------------------
# 510(k) Summary for the Fresenius ATR 40 and ATR 120 Autotransfusion Reservoirs
| Submitter's Name and Address: | Fresenius Hemotechnology, Inc.<br>110 Mason Circle, Suite A |
|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| | Concord, CA 94520 |
| Phone Number: | (925) 688-0990 |
| Telefax Number: | (925) 688-0999 |
| Contact Person: | Virginia Singer |
| Date Summary Prepared: | December 22, 1998 |
| Device Trade Name: | Fresenius ATR 40 and ATR 120 Autotransfusion<br>Reservoirs |
| Common name: | Autotransfusion Blood Collection Reservoir |
| Classification Name: | Autotransfusion Blood Collection Reservoir |
| Legally Marketed Devices to which<br>substantial equivalence is claimed: | Haemonetics Cell Saver Collection Reservoir (K802674)<br>Medtronic EL240 Blood Collection Reservoir (K915737 |
The Fresenius ATR 40 and ATR Autotransfusion Reservoirs are intended for Intended Use: use as autotransfusion reservoirs to collect, defoam, filter, and store blood prior to processing.
### Device Features:
The Fresenius ATR 40 and ATR 120 are 3 liter blood collection reservoirs, with 40 or 120 micron filters respectively. They are used for the collection of blood lost intra- or postoperalively in surgical procedures.
Collection is performed by vacuum suction via appropriate tubing systems e.g. double lumen suction lines for aspiration and simultaneous anticoagulation or drainage adapters connected to wound drains. They are connected to the respective inlet connectors on top of the reservoir: two ¼"-connectors (one vertical, one horizontal), luer female connector. All inlet connectors drain the fluid into a filter via a filter holder.
The ATR 40 contains a three-layer filter and the ATR 120 contains a two-layer filter that traps particulates greater than 40 and 120 microns, respectively. In both filters, the PUR foam is treated with a defoaming agent to dissolve bubbles generated in the blood collection process by mixing blood with air. Except for the filter ATR 40 and ATR 120 are identical. The filters are attached to the filter holder by a cable tie and hang like a pouch in the reservoir lumen.
{2}------------------------------------------------
The vacuum is applied via a separate vacuum line attached to a ¼" vacuum port. Positive and negative pressure level within the reservoir is limited by a pressure relief valve that opens in the range of approximately -300 mmHg and + 100 mmHg.
Up to 3 liters of aspirated blood can pass through the filter and be stored in the reservoir lumen at one The volume of collected blood can be controlled by calibration marks on both sides of the time. reservoir; the calibration marks are in increments of 100 ml to 3000 ml. Collected blood can be transferred to an autotransfusion device via the outlet line attached to the reservoir. The flow of blood through the outlet line can be controlled with a pinch clamp. A universal adapter, compatible with ¼" male adapters and 3/8" female connectors, it attached to the distal end of the outlet line for connection to autotransfusion tubing sets. The reservoir can be attached to an IV pole using a compatible reservoir holder or the hanger on top of the reservoir.
The ATR 40 and ATR 120 Autotransfusion Devices are supplied as sterile (ETO) disposable devices with a non-pyrogenic fluid path.
## Substantial Equivalence of the ATR 40
The Fresenius ATR 40 Autotransfusion Reservoir is substantially equivalent to the Medtronic EL240 Blood Collection Reservoir. The Medtronic EL240 was previously marketed under the trade name Electromedics EL40, cleared under K915737. For purposes of this discussion the Medtronic EL240 will be referred to as the "predicate device".
### Technological Characteristics of the Subject Device Compared with Predicate Device
The 510(k) "Substantial Equivalence Decision Making Process (Detailed)" decision tree (ODE Guidance Memo #K86-3) was used to make a determination of substantial equivalence. The answers to questions identified on this decision tree lead to a determination of substantial equivalence of the ATR 40 Autotransfusion Reservoir.
### 1.0 Does the New Device Have the Same Indication Statements?
Yes. The Fresenius ATR 40 and the Medtronic EL240 reservoirs are collection reservoirs intended to be used during autotransfusion procedures to collect, filter, defoam and store blood; both should not be used as cardiotomy reservoirs.
### 2.0 Does the New Device Have Same Technological Characteristics (e.g., design, materials etc.)?
Yes. Both reservoirs are designed for and operated by the same principles. Blood is drawn into the reservoir using a suction handpiece; vacuum applied to the reservoir aspirates the blood from the operative field into the cavity of the reservoir. The filter is capable of filtering particles greater than 40 microns in diameter. Filtered blood leaves the reservoir via an outlet port adapter at a rate established by the autotransfusion device that further processes the blood.
{3}------------------------------------------------
The materials used to manufacture the reservoir are rigid plastics that sustain sufficient negative and positive pressure to prevent implosion. The filter materials are virtually identical.
### Are the Descriptive Characteristics Precise Enough to Ensure Equivalence? 3.0
No. Although, the Fresenius ATR 40 and the Medtronic EL240 reservoirs are very similar, there may be differences in the design, materials used to manufacture the assemblies and in the manufacturing methods employed to build the assemblies. These differences could impact the biocompatibility, performance, structural integrity and shelf-life of the Fresenius ATR 40 Autotransfusion Reservoir.
#### 4.0 Are Performance Data Available to Assess Equivalence?
Yes. All blood/fluid-contacting materials in the Fresenius ATR 40 have been subjected to biocompatibility testing consistent with FDA's modified ISO standards for biological evaluation of medical devices. Information pertinent to the structural integrity of the ATR 40 and shelf-life validation has been provided.
Data is provided to the Agency that demonstrates that the blood collection, filtration, defoaming, level of hemolysis produced during use, and retention of cellular components during filtration using the Fresenius ATR 40 is substantially equivalent to the Medtronic EL240. Comparative testing was performed on the Medtronic EL240, the currently marketed version of the Electromedics EL40.
### 5.0 Does Performance Data Demonstrate Equivalence?
Yes. Based on the results of the testing cited above, Fresenius has demonstrated that:
- . The Fresenius ATR 40 satisfies requirements of the AAMI/ANSI standards for autotransfusion devices with respect to structural integrity and the materials used to manufacture the disposable set are suitable for the intended use of the device,
- . Shelf-life validation studies pertinent to the Fresenius ATR 40 have determined that the biocompatibility, structural integrity, packaging integrity and sterility of the Fresenius ATR 40 will be maintained for the labeled shelf-life.
- . Testing demonstrated that the blood products prepared using the Fresenius ATR 40 device are of equal quality with respect to filtration efficiency, defoaming, hemolysis, and retention of cellular components as the Medtronic EL240.
CONCLUSION: Based on the information and test results provided in this premarket notification, the Fresenius ATR 40 Autotransfusion Reservoir is substantially equivalent to the currently marketed Medtronic EL240 Blood Collection Reservoir.
{4}------------------------------------------------
## Substantial Equivalence of the ATR 120
The Fresenius ATR 120 Autotransfusion Reservoir is substantially equivalent to the Haemonetics Cell Saver ® Collection Reservoir (K802674). For purposes of this discussion the Haemonetic Cell Saver ® is referred to as the "predicate device".
The 510(k) "Substantial Equivalence Decision Making Process (Detailed)" decision tree (ODE Guidance Memo #K86-3) was used to make a determination of substantial equivalence. The answers to questions identified on this decision tree lead to a determination of substantial equivalence for the ATR 120 Autotransfusion Reservoir.
#### 1.0 Does the New Device Have the Same Indication Statements?
Yes. The Fresenius ATR 120 and the Haemonetics Cell Saver ® are collection reservoirs intended to be used during autotransfusion procedures to collect, filter, defoam and store blood; neither should be used as a cardiotomy reservoir.
### 2.0 Does the New Device Have Same Technological Characteristics (e.g., design, materials etc.)?
Yes. Both reservoirs and designed for are operated by the same principles. Blood is drawn into the reservoir using a suction handpiece; vacuum applied to the reservoir aspirates the blood from the operative field into the cavity of the reservoir. The filter is capable of filtering particles greater than 120 microns in diameter. Filtered blood leaves the reservoir via an outlet port adapter at a rate established by the autotransfusion device that further processes the blood.
The materials used to manufacture the reservoir are rigid plastics that sustain sufficient negative and positive pressure to prevent implosion. The filiter materials are virtually identical.
### 3.0 Are the Descriptive Characteristics Precise Enough to Ensure Equivalence?
No. Although, the Fresenius ATR 120 and the Haemonetics Cell Saver ® reservoirs are very similar, there may be differences in the design, materials used to manufacture the assemblies and in the manufacturing methods employed to build the assemblies. These differences could impact the biocompatibility, performance, structural integrity and shelf-life of the Fresenius ATR 120 Autotransfusion Reservoir.
### 4.0 Are Performance Data Available to Assess Equivalence?
Yes. All blood/fluid-contacting materials in the Fresenius ATR 120 have been subjected to biocompatibility testing consistent with FDA's modified ISO standards for biological evaluation of medical devices. Information pertinent to the structural integrity of the ATR 120 and shelf-life validation has been provided.
{5}------------------------------------------------
Data is provided to the Agency that demonstrates that the blood collection, filtration, defoaming, level of hemolysis produced during use, and retention of cellular components during filtration using the Fresenius ATR 120 is substantially equivalent to the Haemonetics Cell Saver ®.
### 5.0 Does Performance Data Demonstrate Equivalence?
Yes. Based on the results of the testing cited above, Fresenius has demonstrated that:
- . The Fresenius ATR 120 satisfies requirements of the AAMI/ANSI standards for autotransfusion devices with respect to structural integrity and the materials used to manufacture the disposable set are suitable for the intended use of the device,
- . Shelf-life validation studies pertinent to the Fresenius ATR 120 have determined that the biocompatibility, structural integrity, packaging integrity and sterility of the Fresenius ATR 120 will be maintained for the labeled shelf-life.
- . Testing demonstrated that the blood products prepared using the Fresenius ATR 120 device are of equal quality with respect to filtration efficiency, defoaming, hemolysis, and retention of cellular components as the Haemonetics Cell Saver ®.
CONCLUSION: Based on the information and test results provided in this premarket notification, The Fresenius ATR 120 Autotransfusion Reservoir is substantially equivalent to the Haemonetics Cell Saver ® Collection Reservoir.
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/10 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of a bird or eagle, with three curved lines forming its body and wings, and two curved lines forming its tail.
Public Health Service
JUN 2 9 1999
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Virginia Singer Manager, Regulatory Affairs Fresenius Hemotechnology, Inc. 110 Mason Circle, Suite A Concord, CA 94520-1238
K984586 Re: Fresenius ATR 40 and ATR 120 Autotransfusion Reservoirs Regulatory Class: II (Two) Product Code: DTN May 14, 1999 Dated: Received: May 17, 1999
Dear Ms. Singer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Please note: this response to your premarket Federal Register. notification submission does not affect any obliqation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Indications for Use Statement
Page 1 of 1
510(k) Number (if known);
Device Name:
Indications for Use:
K984586
Fresenius ATR 40 and ATR 120 Autotransfusion Reservoirs
The ATR 40 and ATR 120 Autotransfusion Reservoirs are indicated as autotransfusion reservoirs to collect, defoam, filter, and store blood prior to processing.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Bethe R. Bemperele
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number_k984566
Prescription Use
(Per 21 CFR 801.109)
Over The Counter Use _________________________________________________________________________________________________________________________________________________________
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.