← Product Code [DXZ](/productcode/DXZ) · K984526

# ACUSEAL CARDIOVASCULAR PATCH (K984526)

_W. L. Gore & Associates, Inc. · DXZ · Apr 8, 1999 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K984526

## Device Facts

- **Applicant:** W. L. Gore & Associates, Inc.
- **Product Code:** [DXZ](/productcode/DXZ.md)
- **Decision Date:** Apr 8, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3470
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The ACUSEAL™ Cardiovascular Patch is indicated for use in cardiovascular patching.

## Device Story

ACUSEAL™ Cardiovascular Patch is an ePTFE-based surgical implant designed for cardiovascular patching. It incorporates an additional interpositional layer of fluoropolymer material to reduce bleeding through suture holes during surgical implantation. The device is intended for use by surgeons in clinical settings. By minimizing suture hole bleeding, the patch aims to improve surgical outcomes during cardiovascular procedures. It is applied using standard surgical techniques.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by mechanical bench testing and in vivo testing demonstrating that the reduced bleeding feature does not negatively impact material characteristics or tissue response.

## Technological Characteristics

Material: ePTFE with an additional interpositional layer of fluoropolymer copolymer. Form factor: Cardiovascular patch. Principle: Mechanical patch for cardiovascular repair with modified design to reduce suture hole bleeding. Sterilization: Not specified.

## Regulatory Identification

An intracardiac patch or pledget made of polypropylene, polyethylene terephthalate, or polytetrafluoroethylene is a fabric device placed in the heart that is used to repair septal defects, for patch grafting, to repair tissue, and to buttress sutures.

## Predicate Devices

- GORE-TEX® Cardiovascular Patch ([K811841](/device/K811841.md) and [K912107](/device/K912107.md))
- CVPro™ Cardiovascular Patch ([K943736](/device/K943736.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4/8/99

Image /page/0/Picture/1 description: The image shows a handwritten string of characters. The string starts with the letter 'K', followed by the numbers '9', '8', '4', '5', '2', and '6'. The characters are written in a dark ink, and the image has a white background.

Tab 8

#### Summary of Safety and Effectiveness Page 1 of 2

## Applicant:

W.L. Gore & Associates, Inc. 3450 West. Kiltie Lane P.O. Box 500 Flagstaff, AZ 86002-0500

## Contact

Timothy J. Rynn

## Date Prepared

December 11, 1998

## Trade or Proprietary Name

ACUSEAL™ Cardiovascular Patch

## Common or Usual Name

Cardiovascular Patch

## Classification Name

Intracardiac patch or pledget made of polyethylene terephthalate, or polytetrafluoroethylene (21 CFR § 870.3470).

### Device Predicates

GORE-TEX® Cardiovascular Patch (K811841 and K912107); CVPro™ Cardiovascular Patch (K943736).

## Device Description

The ACUSEAL™ Cardiovascular Patch is an ePTFE cardiovascular patch with an optional manufacturing modification consisting of an additional interpositional layer or layers of a fluoropolymer material. The additional material is intended to reduce suture hole bleeding through suture holes when stured into place using standard surgical techniques.

### Statement of Intended Use

Intended for use in cardiovascular patching; reduces bleeding through suture holes.

{1}------------------------------------------------

#### Summary of Safety and Effectiveness Page 2 of 2

# Substantial Equivalence

The applicant device is substantially equivalent in materials to currently marketed ePTFE cardiovascular patch devices. Mechanical testing data demonstrate the applicant device has strength values which are substantially equivalent to the predicate devices. In vivo testing demonstrates that the reduced bleeding feature of the applicant device does not adversely affect its material characteristics or its tissue response properties.

The applicant device is composed of the same inert ePTFE biomaterial as the GORE-TEX® Cardiovascular Patch, plus an additional interpositional layer of a copolymer material.

No new types of safety and effectiveness issues are raised by the modification of product design or materials to reduce suture hole bleeding.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing to the right, with the profiles overlapping each other. The profiles are enclosed within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

APR 8 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Timothy J. Rynn Requlatory Associate Medical Products Division W.L. Gore & Associates, Inc. 3450 West Kiltie Lane P.O. Box 500 Flaqstaff, AZ 86002-0500

Re: K984526 Acuseal™ Cardiovascular Patch Regulatory Class: II (Two) Product Code: DXZ Dated: March 22, 1999 Received: March 26, 1999

Dear Mr. Rynn:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. Be advised that all promotional materials must also contain the warning regarding lack of clinical data to assess enhanced ingrowth and associated adverse effects.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{3}------------------------------------------------

Page 2 - Mr. Timothy J. Rynn

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Thomas J. Callahan

Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K984526 510(k) NUMBER (IF KNOWN): ACUSEALTM Cardiovascular Patch DEVICE NAME:

INDICATIONS FOR USE:

The ACUSEAL" Cardiovascular Patch is indicated for use in cardiovascular patching.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ﻟﮯ Prescription Use (Per 21 CFR 801.109)

OR

・・

Over-The-Counter-Use (Optional Format 1-2-96)

Bera L. Campell Ce
(Division Sign-Off)

Division of Cardiovascular, Respiratory, and Neurological Devices

510(k) Number_ 484

---

**Source:** [https://fda.innolitics.com/device/K984526](https://fda.innolitics.com/device/K984526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
