PROFILE 2000 MAMMOGRAPHY SYSTEM, PRO-2000
K984092 · Trex Medical Corp. · IZH · Feb 5, 1999 · Radiology
Device Facts
| Record ID | K984092 |
| Device Name | PROFILE 2000 MAMMOGRAPHY SYSTEM, PRO-2000 |
| Applicant | Trex Medical Corp. |
| Product Code | IZH · Radiology |
| Decision Date | Feb 5, 1999 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1710 |
| Device Class | Class 2 |
Intended Use
THE PROFILE 2000 MAMMOGRAPHY SysTEM, MODEL PRO-2000, 15 A DEDICATED MANMOGRAPHIC IMAGING SYSTEM USED FOR MAMMOGRAPHIC EXAMINATIONS.
Device Story
Dedicated mammographic imaging system; functions as x-ray source for screening or biopsy procedures. Components: x-ray generator cabinet (single or three-phase), tube stand with non-tilting c-arm, mammographic collimator, operator control panel, and image receptor (film cassette or bucky). Operated by clinical staff in a radiology setting. System is microprocessor-controlled; c-arm is counterbalanced with electro-mechanical locks; fixed source-to-image distance (SID) of 76 cm. Provides x-ray output for image acquisition; supports both film-based and digital-based receptors. Clinicians use images to perform mammographic examinations for diagnostic or screening purposes.
Clinical Evidence
No clinical data. Safety and effectiveness established through bench testing and analysis of system functions.
Technological Characteristics
Microprocessor-controlled x-ray source; single-phase (200-240 VAC) or three-phase (208-440 VAC) power input. C-arm with electro-mechanical locks; 76 cm fixed SID. Dedicated mammographic imaging system. Compatible with film or digital image receptors.
Indications for Use
Indicated for use as a dedicated mammographic imaging system for mammographic examinations. Patient population not specified.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- Bennett Profile Mammography System, Model M-PRO (K964576)
- LORAD M-IV
- Instrumentarium Alpha IQ
- GE Senographe 800T
- Planmed Sophie Classic
Related Devices
- K964576 — PROFILE MAMMOGRAPHY SYSTEM (M-PRO) · Bennett X-Ray Corp. · Jan 10, 1997
- K984091 — CONTOUR 2000 MAMMOGRAPHY SYSTEM, MODEL CTR- 2000 · Trex Medical Corp. · Feb 5, 1999
- K013290 — MODIFIED 650 MAMMOGRAPHY SYSTEM · Lorad, A Hologic Co. · Oct 24, 2001
- K151645 — MAMMOMAT Fusion · Siemens Medical Solutions USA, Inc. · Sep 14, 2015
- K012953 — GIOTTO IMAGE · Medx, Inc. · Nov 30, 2001
Submission Summary (Full Text)
{0}------------------------------------------------
K984092
FEB 5 1999
## SECTION 2
## SUMMARY AND CERTIFICATION
## 510(k) SUMMARY
| Submitted by: | Trex Medical Corporation, Bennett Division |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 445 Oak Street, Copiague, NY 11726 |
| Phone: | (516) 691-6100, (516) 691-6103 (fax) |
| Contact: | Lim Cheung |
| Device Name: | Bennett Profile 2000 Mammography System, Model PRO-2000 |
| Equivalent Device: | Bennett Profile Mammography System, Model M-PRO (#K964576),<br>LORAD M-IV, Instrumentarium Alpha IQ, GE Senographe 800T, and<br>Planmed Sophie Classic |
| Date: | November 13, 1998 |
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 as promulgated in 59 FR 64287-01 for §807.92.
- Description of the Device: The Bennett Profile 2000 Mammographic System, Model PRO-1. 2000, is a dedicated mammographic imaging system used as an x-ray source in the performance of mammographic examinations. The PRO-2000 is adaptable to performing screening and/or biopsy procedures with the addition of either a film-based or digital-based image receptor device attached. It consists of the following components; an x-ray generator cabinet, Model M-2000G (single-phase input) or Model M-2000G-3P (three-phase input); a tube stand with a non-tilting c-arm; mammographic collimator (DM-2000) and tube; an operator control panel, Model M-2000C; and image receptor (film cassette holder or bucky).
The Profile 2000 Mammography System is a microprocessor-controlled x-ray source requiring single-phase 200-240 VAC, 50/60 Hz or three-phase 208-440 VAC, 50/60 Hz for operation. The Profile 2000 system's c-arm is fully counterbalanced and is locked in position using electro-mechanical locks. The c-arm has a fixed source-to-image distance (SID) of 76 cm.
{1}------------------------------------------------
- Same Intended Use: The Bennett Profile 2000 Mammography System (Model PRO-2000), 2. like the Bennett Profile M-PRO, LORAD M-IV, Instrumentarium Alpha IQ, GE Senographe 800T, and Planmed Sophie Classic mammographic imaging systems, is a dedicated mammographic imaging system used for mammographic examinations.
The Bennett Profile 2000 Mammography System is substantially equivalent to currently marketed devices in terms of basic features, operation, and functionality. The safety functions of the Profile 2000 Mammography System have been rigorously tested and analyzed for conformance to requirements. In each case, these functions have performed as required, ensuring that each identified hazardous condition would not occur under simulated conditions. The Profile 2000 Mammography System fulfills its design requirements by providing the operator with the ability to perform safe and effective mammographic examinations.
{2}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 5 1999
Walter F. Schneider President Trex Medical Corporation Bennett Division 445 Oak Street Copiague, NY 11726
Re:
K984092 Bennett Profile 2000 Mammography System Model PRO-2000 Dated: November 13, 1998 Received: November 16, 1998 Regulatory class: II 21 CFR 892.1710/Procode: 90 IZH
Dear Mr. Schneider:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely vours.
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page
510(k) Number (if known):
Device Name: PROFILE 2000 MAMMOGRAPHY SYSTEM,
MODEL PRO-2000
Indications for Use:
THE PROFILE 2000 MAMMOGRAPHY SysTEM, MODEL PRO-2000, 15 A DEDICATED MANMOGRAPHIC IMAGING SYSTEM USED FOR MAMMOGRAPHIC EXAMINATIONS.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED!
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
ਿੱ
Over-The-Counter Use_
(Optional Format 1-2-95)
David A. Seaman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Device 510(k) Number