HEMOSLEEVE, MODEL 4.0MM, 5.5MM

K984018 · Surgical Laser Technologies, Inc. · GEI · Mar 15, 1999 · General, Plastic Surgery

Device Facts

Record IDK984018
Device NameHEMOSLEEVE, MODEL 4.0MM, 5.5MM
ApplicantSurgical Laser Technologies, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateMar 15, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SLT bipolar sheath is indicated for use with currently marketed cutter systems used in Orthopedic Surgery. The SLT bipolar sheath is intended to provide coagulation of tissue at the distal end of the cutter tip. This sheath is designed with a standard interface to allow use with most electrosurgical generators.

Device Story

SLT HemoSleeve Bipolar Sheath is an accessory for orthopedic cutter systems; provides tissue coagulation at distal end of cutter tip. Device features standard interface for compatibility with most electrosurgical generators. Used by surgeons in clinical settings during general surgery procedures. Device functions as an electrosurgical accessory to achieve hemostasis; benefits patient by reducing bleeding during debridement.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Bipolar electrosurgical sheath; standard interface for electrosurgical generators; designed for use with tissue debrider tips.

Indications for Use

Indicated for use in general surgery procedures to provide coagulation at the distal end of a tissue debrider tip. Prescription use only.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Submission Summary (Full Text)

{0}------------------------------------------------ 3/15/99 K984018 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS The SLT bipolar sheath is indicated for use with currently marketed cutter systems used in Orthopedic Surgery. The SLT bipolar sheath is intended to provide coagulation of tissue at the distal end of the cutter tip. This sheath is designed with a standard interface to allow use with most electrosurgical generators. Description statements were not relied on to show substantial equivalence to legally marketed devices; instead, performance data from device validation is used. The comparison of intended use and technological features of this device to other legally marketed devices taken together with validation results indicate that this device is substantially equivalent to legally marketed predicate devices with regards to safety, effectiveness and intended use. The intended use of this device is the same as the intended use of other bipolar electrocautery devices marketed to provide the same tissue effects. TTherefore, all aspect of this device have predicates which are well accepted in the clinical community. This product simply provides an alternative to those currently marketed devices. ଚ {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 5 1999 Ms. Monica Ferrante Regulatory Affairs Surgical Laser Technologies 147 Keystone Drive Montgomeryville, Pennsylvania 18936 Re: K984018 Trade Name: SLT HemoSleeve Bipolar Sheath Regulatory Class: II Product Code: GEI Dated: January 29, 1999 Received: February 1, 1999 Dear Ms. Ferrante: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ ## Page 2 - Ms. Monica Ferrante This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ K984018 ## INDICATIONS FOR USE The SLT HemoSleeve Bipolar Sheath is indicated for use in general surgery procedures to provide coagulation at the distal end of a tissue debrider tip. This device is a prescription device. Prescription Use (Per 21 CFR 801.109) Acoolefa (Division Sign-Off) Division of General and Restorative Devices 510(k) Number __ K984018
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...