BEAR 1000, VEOLAR W LEONARDO GRAPHICS, BICORE CP-2000, BICORE CP-100, HI-LO JET ET TUBE
K983981 · Bear Medical Systems, Inc. · CBK · Nov 20, 1998 · Anesthesiology
Device Facts
Record ID
K983981
Device Name
BEAR 1000, VEOLAR W LEONARDO GRAPHICS, BICORE CP-2000, BICORE CP-100, HI-LO JET ET TUBE
Applicant
Bear Medical Systems, Inc.
Product Code
CBK · Anesthesiology
Decision Date
Nov 20, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5895
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The BEAR® 1000 Ventilator is fully capable of ventilating the entire spectrum of patients requiring such support-from the Emergency Room to the ICU, from pediatric patients to adults. The BEAR® 1000 Ventilator is available in a standard model as well as an upgraded model which includes the fully loaded enhancement package. The standard model BEAR® 1000 Ventilator offers the clinician the following ventilatory modes: Assist CMV, SIMV (with or without Pressure Support), and CPAP (with or without Pressure Support). In addition the Enhancement Package offers additional functions such as: Pressure Control Ventilation, Minimum Minute Volume capability, Pressure Augmentation, Smart Trigger, ES (Esophageal Monitoring).
Device Story
The BEAR 1000es is a volume-cycled/pressure-controlled ventilator. It processes inputs from pressure transducers (tracheal/esophageal catheters) and bi-directional flow sensors to monitor patient respiratory status. The Electronic Control System (ECS) computes flow rates, oxygen concentration, and valve states every 10ms, commanding the Electro/Pneumatic Interface (EPI) to adjust pneumatic valves. It provides various ventilation modes (CMV, SIMV, CPAP) and features like Pressure Augmentation and Smart Trigger to improve patient-ventilator synchrony. The device includes an integrated Graphic Display for real-time waveform/loop visualization and digital data monitoring. Used by clinicians in hospital settings (ICU/ER), the system provides automated leak compensation and alarm monitoring (e.g., prolonged esophageal pressure). The device benefits patients by reducing work of breathing and ensuring consistent support despite patient demand or circuit leaks.
Clinical Evidence
Bench testing only. Performance was verified through Fault Tree Analysis, FMEA of the Flow PCB, and formal qualification procedures for the Graphic Display software (Release 1.03) to ensure compliance with design specifications.
Technological Characteristics
Microprocessor/microcontroller-based control system; pneumatic valves; variable orifice flow sensing; esophageal/tracheal pressure transducers. Connectivity via RS-232 and analog ports for external data output. Software-controlled valve timing (10ms cycles). Modular design with replaceable control panel hood for upgrades.
Indications for Use
Indicated for pediatric and adult patients requiring ventilatory support in clinical settings including the Emergency Room and ICU.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K983981
NOV 20 1998
Image /page/0/Picture/2 description: The image shows the word "BEAR" in a stylized, bold, black font. The letters are slightly slanted to the right, giving the word a dynamic appearance. A registered trademark symbol is present to the bottom right of the word.
Bear Medical Systems, Inc.
2085 Rustin Avenue, Riverside, CA 92507-2437 Phone: (909) 788-)
November 3, 1998
## 510(k) Summary
# BEAR 1000es Ventilator
Bear Medical Systems, Inc. 1100 Bird Center Drive Palm Springs, California 92262
Contact:
Darryl L. Shelby
Phone Number (800) 232-7633 EX 614
FAX (760) 778-7274
{1}------------------------------------------------
| 510k Summary as required by section 807.92 | page 2 of 11 |
|--------------------------------------------|----------------------------------------------|
| TRADE/PROPRIETARY NAME: | BEAR 1000es Ventilator |
| COMMON/USUAL NAME: | Volume Cycled/Pressure-Controlled-Ventilator |
| CLASSIFICATION NAME: | Ventilator, Continuous (Respirator) |
| ESTABLISHMENT REGISTRATION NUMBER: | 2022747 |
| PRODUCT CLASSIFICATION | Class II |
| ANESTHESIOLOGY DEVICE CLASSIFICATION PANEL | (73 CBK) |
#### Comparison to Predicate Device;
| k912619/A | BEAR 1000® Ventilator |
|-----------|---------------------------------------------------------------------|
| k901448 | Hamilton Veolar, with Leonardo Graphics |
| k935788 | SMARTCATH INTRATRACHEAL CATHETER<br>now known as the BICORE CP 2000 |
| k900696B | CP-100 Pulmonary Monitor |
{2}------------------------------------------------
### 510k Summary as required by section 807.92
#### Statement of Indication for Use.
The BEAR® 1000 Ventilator employs both microprocessor and microcontroller technology to vield a simple, compact design. This full-featured package has powerful performance capabilities-in terms of output capability, Sensitivity to patient demand and response time. They render the BEAR® 1000 Ventilator fully capable of ventilating the entire spectrum of patients requiring such support-from the Emergency Room to the ICU, from pediatric patients to adults.
The BEAR® 1000 Ventilator is available in a standard model as well as an upgraded model which includes the fully loaded Enhancement Package. The standard model BEAR® 1000 Ventilator offers the clinician the following ventilatory modes:
- · Assist CMV
- · SIMV (with or without Pressure Support), and
- · CPAP (with or without Pressure Support)
In addition to greater flow and rate capabilities, Enhancement Package offers additional functions such as:
- · Pressure Control Ventilation
- · Minimum Minute Volume capability
- · Pressure Augmentation, and more.
- · Smart Trigger
- · ES (Esophageal Monitoring)
Any standard model can be upgraded to add the EPI features or any other new option; thus from the outset, the ventilator, has been designed for the 1990's, with future upgrades in mind. Consistent with Bear Medical Systems' philosophy of maintaining an easy-to-use operator interface, this ventilator can be upgraded by replacing the control panel hood. In this manner, not only will all new controls remain as logically arranged as those on the standard ventilator model, but control multiplexing can be held to a minimum. Beyond the above described ventilatory mode capabilities, BEAR® 1000 models offer important features which improve ventilator-patient synchrony and may reduce the work of breathing.
Important among these features is the "demand system" which has long been pioneered by Bear Medical Systems. Inc. This demand system provides additional gas flow in response to patient demand, even in volume-controlled breaths. Such response is termed "Flow end volume augmentation" and it serves to ensure that patients with high or asynchronous demand will not be limited by selected peak flow or tidal volume settings.
{3}------------------------------------------------
#### Statement of Indication for Use.
The BEAR® 1000 Ventilator expands the horizons of this responsiveness through its function called "Pressure Augmentation". Further improving ventilator-patient synchrony is the PRESSURE SLOPE control. This control enables the clinician to tailor, for each patient, the rate of pressure rise during inspiration in any pressure breath, and thereby more closely match ventilatory support to patient demand. Also available in the BEAR® 1000 Ventilator, again to improve ventilator-patient synchrony, is a leak makeup system. This system can detect and compensate for very small to rather large leaks. In small leaks, which are more clinically-relevant, this leak makeup system can not only maintain PEEP and avoid AUTOCYCLING, but also maintain consistently sensitive assist trigger levelseven when leaks are intermittent.
#### Device Design.
The modifications to the Bear 1000 % Ventilator are minor, and primarily affect only the monitoring capabilities of the system, with the exception of the addition of one alarm. There is no impact to the control functions of the ventilator. These modifications are:
The addition of esophageal and tracheal pressure monitoring. All monitoring is the result of pressure data that is obtained through the esophageal and tracheal catheters and communicated to the Graphic Display via the RS-232 communication port.
The addition of one alarm -- Prolonged Esophageal Pressure.
The addition of one membrane key switch for filling the esophageal catheter balloon with air.
The addition of a bi-directional flow sensor at the patient wve using a presently approved and marketed technology to enhance the flow and volume monitoring capabilities. The bidirectional flow sensor measures inspiratory and expiratory flow that is used in calculating and displaying digital data and various waveforms/loops on the Graphic Display. The user has the option of using either the bi-directional flow sensor or the exhaled flow sensor.
The digital data calculated by and displayed on the Graphic Display has been enhanced to include 17 additional respiratory parameters which are categorically displayed on four user selected pages.
{4}------------------------------------------------
#### Electronic System Overview:
Electronic Control System (ECS)
The ECS consists of the Control and Monitor PCB's. The Control PCB performs all of the ventilator's higher level functions, while the Monitor PCB scans the membrane touch panels and the seven segment displays. The Control PCB communicates new display information to the Monitor and the Monitor PCB communicates changes in the membrane touch panels to the Control PCB over a parallel eight bit data bus. The ECS also has an RS-232 port, and three analog ports. The RS-232 port is used to communicate the monitor, alarm and control settings to an external device, and send flow and pressure signals in real time to an external graphics display. The analog ports are used to output real time flow, volume and pressure signals.
#### Control PCB
The Control PCB performs all of the ventilators high level control functions. This is accomplished by computing the desired flow rate, oxygen concentration, and desired valve states every 10 msec and sending them to the Electro/Pneumatic Interface (EPI) via the ECS/EPI-Communication Channel (ECS/EPI-CC). The Control PCB's second major function is to monitor, alarm and display all clinical data, based on the transducer data received from the EPI.
#### Electro/Pneumatic Interface (EPI)
The major purpose of the EPI is to provide a clear interface from the ECS to the pneumatic hardware. The EPI receives the desired flow rate, oxygen concentration, and valve states every 10 msec from the ECS. Based on this information, the EPI commands each individual valve to its correct position or state.
The second function of the EPI is to read the transducers, digitize their data, and send it to the ECS upon request. The pressure transducers are physically mounted on the EPI to improve signal quality by minimizing the transmission distance of their analog signals.
#### Power Supply
The power supply converts AC line voltage into the DC voltages used by the system (+5 VDC, +12 VDC & -5 VDC + 7 VDC). It will also provide approximately 150 milliseconds of hold up power in the event of a temporary AC power loss.
{5}------------------------------------------------
510k Summary as required by section 807.92
page 6 of 11
#### Pneumatic System Overview:
Inlet Pneumatic System
The inlet pneumatic system conditions and monitors the air and oxygen supplies entering the ventilator.
Gas conditioning:
The air and 02 inlet filters remove aerosol and particulate contaminants from the incoming gas supplies.
The air & 02 regulator-relay combination enables this system to provide balanced supply pressure to the gas blending system. The air regulator reduces the air supply pressure to 18.0 psig and pilots the 02 relay to track at this same pressure.
In the event the supply air pressure falls below an acceptable level, the Crossover Solenoid opens delivering high pressure oxygen to the air regulator, allowing the 02 Relay to function. In this mode, 02 flowing through the air regulator will also allow the inlet pneumatic system to provide sufficient flow of 100% 02 to the blending system. In the event of an oxygen supply pressure drop below the threshold, the Crossover Solenoid stavs closed. The blender slews to 21 % O2 and regulated air pressure provides sufficient flow of 100% air to the blending system.
#### Oxygen Blender
The Oxygen Blender (O2B) is comprised of three major sub-systems, the valve, the stepper motor and the drive electronics. The oxygen blender receives the system air/O2 from the Inlet Pneumatic System and blends the two gases to the user selected value. The user setting is received by the ECS which commands the EPI to position the blender stepper motor to a pre-calculated position.
#### PEEP/Exhalation Control System & Exhalation Valve
The exhalation valve in tandem with the PEEP/Exhalation control system closes off the expiratory limb of the patient circuit during an inspiration, so that all gas delivered by the machine will be diverted into the patient. Once the breath is completed the valve is opened so that the patient may exhale. Its second function is to maintain a positive end expiratory pressure (PEEP) during exhalation to a set value. The exhalation valve also acts as a check valve so that the patient can not inspire gas through the expiratory leg of the circuit.
{6}------------------------------------------------
### 510k Summary as required by section 807.92
#### Pneumatic System Overview:
#### Exhalation Valve:
The exhalation valve regulates all flow out of the patient circuit. Exhaled gases from the patient flow through the patient circuit and into the exhalation valve. If the valve is open, gases flow past the valve seat and continue on to the exhaled flow sensor. Pilot pressure from the PEEP control pump is applied to the balloon diaphragm. The exhalation valve acts as a piloted relief valve, i.e. when patient pressure exceeds the pilot pressure, the diaphragm lifts off the seat and "relieves" the patient pressure. The pressure ratio (Proximal pressure)/(Control pressure) of the diaphragm is approximately 1:1.3.
Safety System (Shut Off Solenoid and Low Pressure Regulator)
The safety system and SOPR valve allow the patient to breath "room air" in the event of a ventilator or power failure. In addition the safety system sets the exhalation valve reference pressure to zero, allowing the patient to exhale unimpeded. Under normal operating conditions the low pressure regulator generates the 2 psig reference pressure necessary to shut the exhalation valve during inspiration, and close the relief port of the SOPR valve. PEEP drops to zero during a power failure.
#### Advanced Monitoring System overview:
Exhaled Flow Sensor and Advanced Monitoring
Three types of flow sensors are supported: flow sensor at the exhalation valve (also referred to as exhaled flow sensor), adult flow sensor at the wye, and infant flow sensor at the wye. The flow sensor at the exhalation valve is designed to detect only flow in the exhalation direction while both types of flow sensors at the wye are designed to detect flow in both directions. Advanced monitoring consists of tracheal and esophageal pressure sensing.
The exhaled flow sensor measures exhaled flow at the exhalation manifold while both types of flow sensor at the wye measure flow at the circuit wye. The technology used to measure flow with all three types of Low sensors are variable orifice flow sensing. All three types of sensors are based on the Bicore Monitoring Systems disposable flow sensor and are called Varflex type flow sensors.
{7}------------------------------------------------
#### 510k Summary as required by section 807.92 page 8 of 11
#### Advanced Monitoring System overview:
Tracheal monitoring monitors airway pressure on the far side of the endotracheal tube eliminating the resistance of the endotracheal tube. A tracheal catheter is used to measure the tracheal pressure.
Esophageal monitoring monitors pressures in the patient esophagus. Esophageal monitoring provides an accurate representation of pleural pressure and therefore patient effort. The BEAR 1000 monitors both adult and pediatric esophageal pressures. Bicore adult and pediatric esophageal catheters (Balloons) are used for monitoring the esophageal pressure. The BEAR 1000 determines which catheter is connected each time the extension tube is mated to the ventilator. The volume of the balloon is larger for the adult catheter. The BEAR 1000 performs a fill for a duration that would distend the pediatric balloon but not fill the adult balloon. After this fill, the esophageal pressure will be at a maximum for the pediatric catheter but not for the adult.
The system consists of one of any of the three types of flow sensors and a tracheal catheter, an esophageal catheter, internal harness assemblies, pneumatic components, and a printed circuit board (PCB). The Flow Sensor and Advanced Monitoring (FSAM PCB) contains all of the electronics required to perform the monitoring functions. This includes pressure transducers, signal conditioning, analog to digital conversion, solenoids, and support electronics required for data communication.
The FSAM PCB contains a pressure transducer, which converts the pressure drop across the flow sensor to a voltage. The ADC converts this voltage to digital data and communicates it to the EPI.
Two solenoids are used to support flow sensing. The auto zero solenoid is used to cancel time and temperature related drift of the pressure transducer output. Enabling the auto zero solenoid provides the same pressure to both sides of the differential pressure transducer. The analog output value generated at this time is used to offset the zero value stored in the ventilator's memory (EPI characterization EEPROM) at calibration time.
{8}------------------------------------------------
#### 510k Summary as required by section 807.92 page 9 of 11
#### Advanced Monitoring System overview:
The second solenoid used for flow sensing is the purge solenoid, unless the flow sensor is an infant wye type. Purging does not occur if the flow sensor is an infant type at the wye. For the other two types of Varflex flow sensors, periodic purging of the sensor pneumatic lines utilizing fixed volumes of air is used to keep them clear of contamination. The purge solenoid is provided with a 2 psig source of air which is metered into the sense lines as determined by the ventilator control software. Prior to activation of the purge solenoid, the auto-zero solenoid is enabled to protect the differential pressure transducer from most pressure spikes due to an occlusion of the pneumatic lines.
Two solenoids are used to support esophageal pressure monitoring. The source solenoid switches between Negative flow (for balloon evacuation) and positive flow (for balloon filling). An isolated solenoid is used to isolate the balloon from the flow source.
Tracheal Catheter:
The tracheal catheter is a single lumen 4 FR catheter. The catheter is inserted into the side port adapter attached to the patient. The catheter contains a luer lock for securing the catheter in place with the side port adapter. The catheter mates with a tracheal catheter extension tube which mates with the ventilator. The extension tube connector contains two
Esophageal Catheter:
The esophageal catheter is a single lumen 8 FR catheter end contains several holes to allow filling and evacuation of the balloon. The catheter mates with a esophageal catheter extension tube which mates with the ventilator. The extension tube connector contains two pins which are electrically shorted allowing the FSAM PCB to detect when the extension tube is connected to the ventilator.
{9}------------------------------------------------
page 10 of 11
#### Comparison to Predicate Device
The BEAR 1000es Ventilator is the same as the BEAR 1000 Ventilator (k912619A) with the BICORE monitoring (k935788) (k900696b) technology added. The external additions to the ventilator are three connections on the front bezel and one button to fill the esophageal balloon located on the lower front panel. Internal change are the addition of a Circuit board with the pressure transducers. Software and 17 added features to the Graphic Display Monitor. There is an addition of a prolong esophageal balloon pressure alarm. External connection are used to connect Flow Transducers and various size Catheters identical to that used with the BICORE monitoring systems (k935788) (k900696b).
CLEARED 510(k),
BEAR 7 now know as the BEAR 1000 (UNMODIFIED) device. K912619/A
Comparison to Predicate Device; Hamilton Veolar, with Leonardo Graphics.
k901448 Hamilton Veolar, with Leonardo Graphics
Comparison to Predicate Device: BEAR 1000 Ventilator and Bicore monitors CP-100 and CP-2000.
- K935788 SMARTCATH INTRATRACHEAL CATHETER now known as the BICORE CP 2000
- K900696B CP-100 Pulmonary Monitor
{10}------------------------------------------------
#### Summary of Performance Testing
As a result of the review of the Fault Tree Analysis, a FMEA was done for the Flow PCB and concluded that there are no risk associated with any potential failures of components on the PCB.
The software for the ventilator was qualified by performing formal qualification procedures on a test ventilator. These qualification tests consisted of a Mini-qual, which verifies the ventilator's performance in accordance with the approved Design and Performance Specification, as well as a number of independent tests associated with the verification of the software changes.
Although much of the Graphic Display software is unchanged, many new features have been added. As a result, a complete formal qualification was performed on the Graphic Display. The qualification consisted of a complete verification of all features as specified in the Design and Performance Specification, resulting in software Release 1.03.
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the seal of the Department of Health & Human Services (USA). The seal features a stylized eagle with three heads in profile, facing right. The eagle is surrounded by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 20 1998
Mr. Darryl L. Shelby Bear Medical Systems, Inc. 1100 Bird Center Drive Palm Springs, CA 92262
Re: K983981 BEAR 1000es Ventilator Requlatory Class: II (two) Product Code: 73 CBK November 3, 1998 Dated: Received: November 9, 1998
Dear Mr. Shelby:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{12}------------------------------------------------
Page 2 - Mr. Darryl L. Shelby
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
Bear Medical Systems, Inc.
Image /page/13/Picture/1 description: The image shows the word "BEAR" in a stylized, bold font. The letters are black with horizontal white lines running through them, creating a textured effect. A registered trademark symbol is present to the right of the letter "R".
Rustia Avenue, Hiverside, CA 92507-2437
Phone: (909) 788-2460
(800) 232-7633
Fax: (909) 351-4609
## Statement of Intended use: [As required by 21 CFR 807.87(e)]
The BEAR® 1000 Ventilator is fully capable of ventilating the entire spectrum of patients requiring such support-from the Emergency Room to the ICU, from pediatric patients to adults.
The BEAR® 1000 Ventilator is available in a standard model as well as an upgraded model which includes the fully loaded enhancement package. The standard model BEAR® 1000 Ventilator offers the clinician the following ventilatory modes:
- · Assist CMV
- · SIMV (with or without Pressure Support), and
- · CPAP (with or without Pressure Support)
In addition the Enhancement Package offers additional functions such as:
- · Pressure Control Ventilation
- · Minimum Minute Volume capability
- · Pressure Augmentation
- Smart Trigger
- ES (Esophageal Monitoring)
Kramer
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 1983481 510(k) Number .
prescriptions use ✓
OTC***_***_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.