# CERALAS DIODE LASER SYSTEM (MODEL H) (K983950)

_Ceram Optec, Inc. · GEX · Feb 3, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983950

## Device Facts

- **Applicant:** Ceram Optec, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Feb 3, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Ceralas H Laser System is a surgical instrument indicated to treat leg veins in dermatology and plastic surgery procedures. Specific indications for the Ceralas H Laser System include the treatment of leg veins.

## Device Story

Ceralas H 808nm Diode Laser System is a compact surgical laser for treating leg veins. Device utilizes gallium aluminum arsenide (GaAlAs) semiconductor diodes to generate near-infrared laser radiation. System includes console, footswitch, and handpiece. Operator (physician) places handpiece against patient skin; triggers pulse via footswitch and handpiece button. Laser delivers pulsed light (5-30ms duration, 10-40J/cm2 fluence) to skin surface. Output energy is 8-32 Joules per pulse. Used in clinical settings (dermatology/plastic surgery). Device provides controlled thermal energy to target leg veins.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

GaAlAs semiconductor diode laser; 808 ± 20nm wavelength; 9x9mm handpiece; 5-30ms pulse duration; 10-40J/cm2 fluence; 1Hz max repetition frequency. Self-contained console with footswitch. Software developed per FDA guidance for computer-controlled medical devices.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Star Medical Technologies, Inc.'s StarLight™ Pulsed Diode Array Laser System ([K974346](/device/K974346.md) and [K973324](/device/K973324.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 983950

FEB 3 1339

## 510(k) Summary Ceralas Diode Laser System

# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

CeramOptec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028 (413) 525-0600 Phone: Facsimile: (413) 525-0611

Carol J. Morello, V.M.D. Contact Person: February 4, 1998 Date prepared:

## Name of Device and Name/Address of Sponsor

Ceralas H 808nm Diode Laser System CeramOptec, Inc. 515 Shaker Road East Longmeadow, Massachusetts 01028

### Classification Name

Laser, Surgical Diode Laser System

#### Predicate Device

Star Medical Technologies, Inc.'s StarLight™ Pulsed Diode Array Laser System (K974346 and K973324).

## Intended Use

The Ceralas H Laser System is a surgical instrument indicated to treat leg veins in dermatology and plastic surgery procedures. Specific indications for the Ceralas H Laser System include the treatment of leg veins. As explained below, the Ceralas H Laser System is substantially equivalent to Star Medical Technologies, Inc.'s StarLight™ Pulsed Diode Array Laser System (K974346 and K973324).

{1}------------------------------------------------

# Technological Characteristics and Substantial Equivalence

The Ceralas H Laser System is a complete, self-contained compact surgical laser that utilizes gallium alumnum arsenide (GaAlAs) semiconductor diodes to generate near-infrared laser radiation. The laser system consists of a console, footswitch, and handpiece which are connected to the device's console. The diode laser is enclosed in a rugged, factory aligned, environmentally protective module. The handpiece is placed against the patient's skin and a pulse of light is delivered to the skin surface when the footswitch and handpiece trigger are depressed.

The wavelength for the Ceralas H Laser System is 808 ± 20nm. The laser delivers pulsed light with a selectable pulse duration of 5-30 millsseconds and a fluence of 10-40J/cm2. The corresponding pulse energy delivered through the 9x9mm handpiece is 8-32 Joules. The laser pulses are generated at a maximum pulse repetition frequency of up to 1 Hz. Additionally, the software and software development activities are in accordance with FDA's guidance document. "Reviewer Guidance For Computer Controlled Medical Devices Undergoing 510(k) Review."

The Ceralas H Laser System and the StarLight Diode Laser System have equivalent technological characteristics. For instance, both devices are diode lasers which operate at a similar beam mode, fluence, reperition rate, pulse width, and pulse energy. The principle technological distinction between the two devices is that the StarLight Diode Laser System's wavelength is 800nm and the Ceralas H Laser System's wavelength is 808+20nm. However, it is not believed that the difference between the devices' wavelength raises any new questions of safety and effectiveness. In summary, although there are minimal differences between the Ceralas H Laser System and its predicate device, these differences are minor and raise no new questions of safety and efficacy.

#### Performance Data

None required.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines. The logo is black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

3 1039 FEB

Mr. Jonathan S. Kahan c/o Hogan & Hartson, L.L.P. 555 13th Street, NW Washington, D.C. 20004-1109

Re: K983950 Trade Name: Ceralas Diode Laser System (Model H) Regulatory Class: II Product Code: GEX Dated: November 4, 1998 Received: November 5, 1998

Dear Mr. Kahan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action, In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

### Page 2 -- Mr. Jonathan Kahan

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Hyst Aurelius

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Page_of

x983950 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Ceralas H 808nm Diode Laser System

Indications For Use:

The Ceralas H 808nm Diode Laser System is indicated to treat leg veins in dermatology and plastic surgery procedures.

+

Hok Diodes

(Division Sign-Off) Division of General Restorative Devices x983950 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH. Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

X Prescription Use . Prescription Sul.109)
(Per 21 CFR 801.109)

---

**Source:** [https://fda.innolitics.com/device/K983950](https://fda.innolitics.com/device/K983950)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).