← Product Code [DZL](/productcode/DZL) · K983728

# LORENZ IMF SCREW (K983728)

_Walter Lorenz Surgical, Inc. · DZL · May 13, 1999 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983728

## Device Facts

- **Applicant:** Walter Lorenz Surgical, Inc.
- **Product Code:** [DZL](/productcode/DZL.md)
- **Decision Date:** May 13, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4880
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Lorenz IMF Screw is intended for use as a bone screw in temporary fixation of the maxilla and mandible, providing indirect stabilization of fractures of the maxilla and/or the mandible.

## Device Story

Lorenz IMF Screw is a bone screw for maxillomandibular fixation; used for temporary stabilization of maxillary and mandibular fractures. Device features 1.7mm diameter; lengths 8mm-20mm; head includes relief groove with optional hole for wire or elastic band attachment. Self-drilling tip design allows placement with or without predrilled holes. Implanted temporarily in maxilla and mandible by surgeons. Output is mechanical fixation of bone segments; facilitates fracture healing by providing indirect stabilization. Device is non-sterile; requires user sterilization.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Bone screw; 1.7mm diameter; 8mm-20mm length; self-drilling tip; head with relief groove/hole for wire/elastic attachment. Non-sterile; requires user sterilization.

## Regulatory Identification

An intraosseous fixation screw or wire is a metal device intended to be inserted into fractured jaw bone segments to prevent their movement.

## Predicate Devices

- Leibinger IMF Screw ([K963030](/device/K963030.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its wings spread, surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA". The eagle is a common symbol of the United States, and its presence on the seal signifies the department's role in protecting the health and well-being of the American people.

MAY 2 5 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Diana Preston Regulatory Affairs Specialist Walter Lorenz Surgical, Incorporated 1520 Tradeport Drive Jacksonville, Florida 32218-2480

Re: K983728

Trade/Device Name: Lorenz IMF Screw Regulation Number: 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: February 12, 1999 Received: February 16,1999

Dear Ms. Preston:

This letter corrects our substantially equivalent letter of February 12, 1999 regarding the incorrect product code of the Lorenz IMF Screw.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2- Ms. Preston

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Pact 807); labeling (21 CFR Pact 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

MAY 1 3 1999

K983728

## Summary of Safety and Effectiveness

Device Name: Lorenz IMF Screw

## Classification Name and Reference: Screw, Fixation, Intra osseous (21 CFR 872.4880

- Intended Use: The Lorenz IMF Screw is intended for use as a bone screw in temporary fixation of the maxilla and mandible, providing indirect stabilization of fractures of the maxilla and/or the mandible.
- Device Description: The bone screw for maxillomandibular fixation is 1.7mm in diameter and lengths may range from 8mm - 20mm. The head has a relief groove which may or may not have a hole in which a wire or elastic bands can be wrapped around the screws which are temporarily implanted in the maxilla and mandible. The tip of the screw is designed so that a predrilled hole is not required, but may be used.

## Potential Risks:

- Nonunion or delayed union which may lead to breakage of device.
- . Metal sensitivity or allergic reaction to a foreign body.
- . Pain, discomfort or abnormal sensations due to the presence of the device.
- Nerve damage due to trauma or improper placement of the device.
- Other conditions brought on by the surgical procedure including skin irritation and infection.
- . The device may bend, loosen, or fracture while implanted.
- Biomechanical complications due to improper positioning of the mandibular condyle.

Substantially Equivalent Devices: The device is believed to be substantially equivalent to Leibinger IMF Screw K963030.

Sterility Information: The device is being labeled nonsterile. Sterility recommendations are included in the package insert.

---

**Source:** [https://fda.innolitics.com/device/K983728](https://fda.innolitics.com/device/K983728)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
