CYBER-LASE 2000 CURING AND WHITENING LASER

K983654 · Ics of North America, Inc. · GEX · Jan 12, 1999 · General, Plastic Surgery

Device Facts

Record IDK983654
Device NameCYBER-LASE 2000 CURING AND WHITENING LASER
ApplicantIcs of North America, Inc.
Product CodeGEX · General, Plastic Surgery
Decision DateJan 12, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Source of illumination for curing dental restorative materials, and for tooth whitening activities.

Device Story

The CYBER-LASE 2000 is an air-cooled argon laser system; emits energy at 457 and 502nm; output energy <450 mW. Delivery system consists of a fused silica fiber optic cable and stainless steel handpiece. Operated by dental professionals in a clinical setting for curing dental restorative materials and tooth whitening. Device connects to standard 120V AC power. Provides controlled light energy to target dental materials or whitening agents; clinical benefit includes efficient material polymerization and accelerated whitening procedures.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Air-cooled argon laser; wavelengths 457nm and 502nm; output <450 mW. Delivery via fused silica fiber optic and stainless steel handpiece. Dimensions: 16" x 10" x 6.7"; weight: 20 lbs. Power: 120V AC, 20A. Complies with 21 CFR 1040.10 and 1040.11(a).

Indications for Use

Indicated for use as a source of illumination for curing dental restorative materials and assisting in the whitening process in a dental office setting.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JAN 1 2 1999 983654 XII, 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (Separate Page) A. Submitter: Alan A. Creamer, President, ICS of North American, Inc., 10054 Mesa Ridge Ct. San Diego. CA 92121 I. Classification: Class II. II, Common or usual name: Argon Laser system III. Proprietary Name: CYBER-LASE 2000™ IV. Registration No .: 61073 V. Classification Name: Argon laser system, powered, 79GEX, CFR 878.4810, Class II. VI. Performance standards: None established (as a medical device) under section 514. Meets Performance Standards for Light-Emitting Products, 21 CFR 1040.10 (Laser Products) and 21 CFR 1040.11(a) (Medical Laser Products). VII. Description: The device is a standard air-cooled argon laser emitting energy at 457 and 502nm. Output energy is less than 450 mW. The delivery system is a fused silica fiber optic system with a stainless steel handpiece. The laser operates on any 120 volt, alternating current, 20 ampere circuit. The unit is about 16" by 10" by 6.7" in dimension and weighs 20 lbs. VIII. Labels and Labeling: Labels and Instructions for Use are provided. Competitive labels and labeling are provided and the products are compared. IX. Indications for Use: source of illumination for curing dental restorative materials, and for tooth whitening activities. X. Substantial Equivalence: The CYBER-LASE 2000™ is substantially equivalent to lasers cleared by Premier Laser Systems, Inc., as the product, "Arago" under 510(k) K971118, and other Premier lasers including Argon Curing Laser cleared under K961682. It also is equivalent to similar lasers cleared by other laser companies such as LaserMed, Inc. (Accucure 1000, cleared under K970637) and American Medical laser. The "510(k) Substantial Equivalence Decision-making Process (Detailed) from ODE Guidance Memorandum #86-3 was followed. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The logo is presented in black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 1 2 1999 Mr. Alan A. Creamer President ICS of North America. Inc. 10054 Mesa Ridge Court San Diego, California 92121 Re: K983654 Trade Name: Cyber-Lase 2000™ Regulatory Class: II Product Code: GEX Dated: October 16, 1998 Received: October 19, 1998 Dear Mr. Creamer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Mr. Alan A. Creamer This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, f Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ IX. Indications for Use: [Separate Page] K983654 510(k) Number: NA ## Device Name: CYBER-LASE 2000™ This device is intended for use in: 1. Source of illumination for curing dental restorative materials, 2. Assisting in the whitening process in dental office. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) or (Division Sign-Off) Division of General Restorative Deviges 510(k) Number 1983654 3 Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...