← Product Code [KQG](/productcode/KQG) · K983475

# HEMOCHRON RESPONSE (K983475)

_International Technidyne Corp. · KQG · May 3, 1999 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983475

## Device Facts

- **Applicant:** International Technidyne Corp.
- **Product Code:** [KQG](/productcode/KQG.md)
- **Decision Date:** May 3, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5400
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

The HEMOCHRON ® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays including the following HEMOCHRON Whole Blood Coagulation Assays Activated Clotting Time (ACT) -- All HEMOCHRON types Activated Partial Thromboplastin Time (APTT) - All HEMOCHRON types Prothrombin Time (PT) - All HEMOCHRON types Thrombin Time (TT) Heparin Neutralized Thrombin Time (HNTT) High Dose Thrombin Time (HiTT) Fibrinogen (FIB) Protamine Dose Assay (PDA) - All HEMOCHRON types Heparin Response Time (HRT) - All HEMOCHRON types Protamine Response Time (PRT) - All HEMOCHRON types For In Vitro Diagnostic Use Only

## Device Story

HEMOCHRON Response is an upgraded coagulation analyzer for professional clinical use. Input: whole blood or plasma samples in test tubes containing a magnet. Operation: magnet rotates within the sample; two solid-state Hall effect sensors detect magnet position. When a clot forms, the magnet is trapped, shifting out of the detection area; this electrical change triggers the timer to stop, signaling clot formation via an audible beep. Includes a UPC-E bar code reader for automated identification of test assay, expiration date, and lot number. Output: quantitative coagulation test results. Benefits: improved test well reliability, elimination of calibration drift, and advanced patient/QC data tracking. Healthcare providers use results for hemostasis management.

## Clinical Evidence

Clinical study used a split-sample design with 42 patients, yielding 242 comparative ACT results; correlation r=0.94. Laboratory studies compared four assays (ACT, APTT, PT, HiTT) against predicate HEMOCHRON instruments using split samples. ACT (n=66, r=0.96), APTT (n=41, r=0.99), PT (n=22, r=0.986), and HiTT (n=29, r=0.91) all demonstrated comparable sensitivity and performance across reportable ranges.

## Technological Characteristics

Mechanical clot detection system using a rotating magnet and two solid-state Hall effect sensors. Includes UPC-E bar code reader for assay identification. Features software/firmware upgrade for data tracking and computer interface. Operates on whole blood or plasma samples.

## Regulatory Identification

A coagulation instrument is an automated or semiautomated device used to determine the onset of clot formation for in vitro coagulation studies.

## Special Controls

*Classification.* Class II (special controls). A fibrometer or coagulation timer intended for use with a coagulation instrument is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- HEMOCHRON Model 8000 ([K930068](/device/K930068.md))
- HEMOCHRON Models 401
- HEMOCHRON Models 801

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

5/3/95

(983471

# T ORPO

Image /page/0/Picture/6 description: The image shows the word "Fax" in bold, black letters against a white background. The font is simple and sans-serif, making the word easily readable. The letters are closely spaced, creating a compact and impactful visual. The overall impression is clean and straightforward.

| To:    | Djuana Blagmon          | From:          | John Clay       |                |
|--------|-------------------------|----------------|-----------------|----------------|
| Fax:   | 301-594-5940            | Pages:         | Including cover |                |
| Phone: | 301-594-1243            | Date:          | April 27, 1999  |                |
| Re:    | K983475 Revised Summary | CC:            | F. LaDuca, Ph.D |                |
| Urgent | For Review              | Please Comment | Please Reply    | Please Recycle |

. Comments:

Dear Djuana:

Attached is the revised 510K Summary including the instrument correlations.

The original is being sent to the mail center via UPS. Please replace the previous version sent on

April 22, 1999.

If there are any questions, please contact me.

Regards,

John Clay

Regulatory Compliance Officer

{1}------------------------------------------------

Image /page/1/Picture/3 description: The image shows the logo and contact information for International Technidyne Corporation (ITC). The address is 8 Olsen Avenue, Edison, NJ 08820. The toll-free phone number is (800) 631-5945, the phone number is (732) 308-3700, and the fax number is (732) 632-9299.

# 510 (k) Summary HEMOCHRON® Response

April 27, 1999 Prenarcd:

Submitted by: John Clay International Technidyne Corp. 6 Olsen Ave. Edison, NJ 08820 (732-548-5700) Ext. 265 (732-548-2325) Fax

This summary of safety and effectiveness is heing submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

#### Device Name

Common / Usual Name: HEMOCHRON® Response Coagulation Analyzer. Product Name: HEMOCHRON® Response

#### Predicate Device

IIEMOCIIRON® Model 8000 / HFMOCHRON® Models 401 / 801

#### Device Description

The HEMOCIIRON® Response is an upgrade of the HEMOCHRON® Coagulation Instruments, HFMOCHRON® Model 8000 that was approved under 510(k) K930068 and the HEMOCHRON® Models 401 and 801. The HEMOCHRON® Response performs the same tests as the predicate HEMOCHRON@ Instruments. All test assays are previously 510(k) approved. The IIEMOCIIRON® Response employs the same mechanical clot detection system as the predicate HEMOCHRON® Instruments. Mechanical systems can be used to monitor clotling times in either whole blood or plasma samples.

The HEMOCHRON® employs a mechanical clot detection system. The principle of operation is based on the electrical field generated by a magnet contained within a glass test tube when the magnet is in close proximity to the detector located within the test well.

{2}------------------------------------------------

To perform a test, blood is added to the test tube and placed in the test well. The magnet freely rotates within the tube, in a non-clotted sample. The magnet position is detected by two solid-state Hall effect sensors. When a clot forms the magnet is caught within the clot and is shifted out of the detection area. The electrical change that occurs due to the magnet rotation triggers the timer to stop with an audible beep signaling clot formation to the user.

# Statement of Intended Use

The HEMOCIIRON® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays.

For In Vitro Diagnostic Use Only

# Summary of Technological Characteristics and Performance Data

The HEMOCHRON® Response is a sollware / firmware and mechanical upgrade of the HEMOCHRON® Instruments designed to perform the same tests as the predicate instruments. The upgrade provides the end user with additional quality features not currently available in the predicate HEMOCHRON® Instruments.

The HEMOCHRON@ Response is a modification of the HEMOCHRON® Instruments with improved test well operation and reliability through the use of two Hall Effect solidstate detectors. This provides for full magnet position tracking within the test tube and eliminates the calibration drift of well parameters.

In addition a UPC-E bar code detector has been added to automatically read the affixed bar code label and identify the test assay, expiration date and lot number of the test assay, The instrument provides advanced patient and OC data tracking and streamlined computer interface capabilities which provide essential quality features for the end users.

# The following is a summary of Clinical and Laboratory data demonstrating substantial equivalence between the HEMOCHRON Response and the predicate HEMOCHRON instrument.

#### Clinical

Using a split sample design, the data was collected from 42 patients, yielding a total of 242 comparative ACT results. The test results were highly correlated (r = 0.94). y = (),912x +24.77

#### Laboratory

Four representative assays were selected to demonstrate substantial equivalence. (AC)', APTT, PT(citrate) and HiTT) In all studies split samples were obtained to run on both the HEMOCHRON Response and the predicate HEMOCHRON.

{3}------------------------------------------------

### ACT

An in vitro heparin dose response employing normal donor blood demonstrated the comparable sensitivity of the ACT through the entire reportable range (up to 1000 seconds) of the assay. (n = 66, r = 0.96) y=0.95x + 18.00

#### APTT

An in vitro heparin dose response employing normal donor blood demonstrated the comparable sensitivity of the APTT through the catire reportable range of the assay. (n = 41, r = 0.99) y=1.03x - 4.51

#### PT (citrate)

Comparable PT results using the HEMOCHRON Response and the predicate HEMOCHRON were generated using freshly obtained blood specimens from patients receiving low doses of oral anticoagulant. (n = 22, r = 0.986) y= 1.00x -3.876

#### HITT

An in vitro heparin dose response employing normal donor blood demonstrated the comparable sensitivity of the HiTT through the reportable range of the assay. (n=29,r=0.91) y=0.89x +20.58

#### Conclusion:

The HEMOCHRON Response is substantially equivalent to the predicate HEMOCHRON instrument. The HEMOCHRON Response system has the same intended use and employs the same assays as the predicate HEMOCHRON.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

#### 3 1000 MAY

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. John Clay Regulatory Compliance Officer International Technidyne Corporation 6 Olsen Avenue Edison, New Jersey 08820

Re: K983475 Trade Name: HEMOCHRON® Response Regulatory Class: II Product Code: KQG, JPA Dated: February 23, 1999 Received: February 24, 1999

Dear Mr. Clay:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Image /page/4/Picture/8 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized image of a bird-like figure with three wavy lines representing its body. The logo is black and white.

{5}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# 510(k) Number (If Known): K983475

Device name: HEMOCHRON® Response

Indications for Use:

6

The HEMOCHRON ® Response is intended for professional use for hemostasis management in a variety of clinical settings for the quantitative determination of an assortment of coagulation test assays including the following HEMOCHRON Whole Blood Coagulation Assays

Activated Clotting Time (ACT) -- All HEMOCHRON types Activated Partial Thromboplastin Time (APTT) - All HEMOCHRON types Prothrombin Time (PT) - All HEMOCHRON types Thrombin Time (TT) Heparin Neutralized Thrombin Time (HNTT) High Dose Thrombin Time (HiTT) Fibrinogen (FIB) Protamine Dose Assay (PDA) - All HEMOCHRON types Heparin Response Time (HRT) - All HEMOCHRON types Protamine Response Time (PRT) - All HEMOCHRON types

For In Vitro Diagnostic Use Only

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Clinical Laboratory Devices/L983475
510(k) Number **__**

Prescription Use ***_*** or Over- the- Counter Use **_**

Prescription Use Per 21 CFR 801.109

(Optional Format 1-2-96)

---

**Source:** [https://fda.innolitics.com/device/K983475](https://fda.innolitics.com/device/K983475)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
