← Product Code [JJE](/productcode/JJE) · K983345

# BAYER IMMUNO 1 SYSTEM, UPGRADED FOR A LABORATORY (K983345)

_Bayer Corp. · JJE · Dec 7, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983345

## Device Facts

- **Applicant:** Bayer Corp.
- **Product Code:** [JJE](/productcode/JJE.md)
- **Decision Date:** Dec 7, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.2160
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Bayer Immuno 1™ System has been upgraded for Laboratory Automation capability. Software and Hardware will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation System (LAS) will coordinate the scheduling, sample identification, and physical transfer of samples, as demonstrated for the following representative methods: CEA, FSH, Free T4, PSA, T3, T4, TSH, and T-Uptake.

## Device Story

The Bayer Immuno 1 System Upgraded for Laboratory Automation (LAS) is an enhancement to the existing Bayer Immuno 1 diagnostic platform. It integrates hardware and software modifications to enable communication and physical interface with external laboratory automation systems. The system receives samples via robotic transfer from the LAS, which manages sample scheduling and identification. The device performs diagnostic assays on these samples and returns results to the laboratory information system. Used in clinical laboratory environments, it is operated by laboratory technicians. The upgrade ensures that the analytical performance of the Immuno 1 system remains consistent with the predicate device while allowing for high-throughput, automated workflow integration. This automation reduces manual handling, improves sample throughput, and maintains clinical decision-making accuracy for the specified analytes.

## Clinical Evidence

Bench testing only. Performance was evaluated by comparing the upgraded system against the predicate Bayer Immuno 1 System using 360 samples for most assays (200 for PSA). Primary endpoints included precision (Total CV%) and method correlation (regression analysis). Correlation coefficients (r) ranged from 0.987 to 1.000 across all eight representative analytes. Precision data demonstrated consistent performance between the upgraded and predicate systems.

## Technological Characteristics

The device is an automated in vitro diagnostic analyzer upgraded with hardware and software interfaces for robotic sample transfer. It utilizes the same analytical methods and reagents as the predicate Bayer Immuno 1 System. Connectivity is established via communication protocols for integration with Laboratory Automation Systems (LAS). The system is designed for clinical laboratory use, operating as a standalone analyzer integrated into a larger automated track system.

## Regulatory Identification

A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.

## Predicate Devices

- Bayer Immuno 1™ System ([K983345](/device/K983345.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Attachment 3

#### 7 1998 DEC

## Summary of Safety and Effectiveness

# Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. Below is a comparison of performance, using eight representative assays, between the Bayer Immuno 1 System upgraded for Laboratory Automation capability and the Bayer Immuno 1™ System, already granted clearance of substantial equivalence for these representative and other assays.

#### Intended Use

The Bayer Immuno 1 System upgraded for laboratory automation capability will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation system will coordinate the scheduling, sample identification, and physical transfer of samples.

#### Device

Bayer Immuno 1" System Upgraded for Laboratory Automation Capability

Gabriel J. Murace, Jr.

Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097

9/21/98
Date

{1}------------------------------------------------

| Part Numbers |           |                                                   |
|--------------|-----------|---------------------------------------------------|
|              | CEA:      | Reagents: T01-3184-51<br>Calibrators: T03-3188-01 |
|              | FSH:      | Reagents: T01-3086-51<br>Calibrators: T03-3148-01 |
|              | Free T4:  | Reagents: T01-3360-51<br>Calibrators: T03-3401-01 |
|              | PSA:      | Reagents: T01-3450-51<br>Calibrators: T03-3541-01 |
|              | T3:       | Reagents: T01-2949-01<br>Calibrators: T03-2872-01 |
|              | T4:       | Reagents: T01-3260-51<br>Calibrators: T03-3174-01 |
|              | TSH:      | Reagents: T01-2942-51<br>Calibrators: T03-3568-01 |
|              | T-Uptake: | Reagents: T01-3036-51<br>Calibrators: T03-3076-01 |

### Predicate Device

Bayer Immuno 1™ System

## Part Numbers

The part numbers are identical to those used for the Bayer Immuno 1 " System upgraded for laboratory automation capability.

.

{2}------------------------------------------------

| BAYER IMMUNO 1™ UPGRADED SYSTEM |                 |                 | BAYER IMMUNO 1™ METHOD SHEET |                 |
|---------------------------------|-----------------|-----------------|------------------------------|-----------------|
| METHOD                          | MEAN<br>(UNITS) | TOTAL<br>CV (%) | MEAN<br>(UNITS)              | TOTAL<br>CV (%) |
| CEA                             | 1.54 ng/mL      | 3.3             | 2.9 ng/mL                    | 2.3             |
|                                 | 13.29 ng/mL     | 1.2             | 8.9 ng/mL                    | 2.2             |
|                                 | 63.43 ng/mL     | 1.4             | 18.2 ng/mL                   | 1.8             |
| FSH                             | 9.4 mIU/mL      | 1.6             | 5.5 mIU/mL                   | 3.2             |
|                                 | 15.26 mIU/mL    | 1               | 12.3 mIU/mL                  | 2.8             |
|                                 | 47.51 mIU/mL    | 0.7             | 30.3 mIU/mL                  | 2.8             |
| FREE T4                         | 0.83 ng/dL      | 4.1             | 0.36 ng/dL                   | 15.4            |
|                                 | 1.43 ng/dL      | 4.3             | 1.25 ng/dL                   | 6.0             |
|                                 | 3.06 ng/dL      | 4.7             | 4.35 ng/dL                   | 3.1             |
| PSA                             | 2.89 ng/mL      | 3.1             | 0.05 ng/mL                   | 0.007 ng/mL     |
|                                 | 6.9 ng/mL       | 2.4             | 0.92 ng/mL                   | 2.9             |
|                                 | 26.41 ng/mL     | 2.4             | 2.68 ng/mL                   | 3.1             |
|                                 |                 |                 | 9.87 ng/mL                   | 2.2             |
|                                 |                 |                 | 23.02 ng/mL                  | 3.4             |
|                                 |                 |                 | 49.13 ng/mL                  | 2.0             |
|                                 |                 |                 | 96.38 ng/mL                  | 2.5             |
| T3                              | 0.66 ng/mL      | 5.1             | 0.46 ng/mL                   | 13.3            |
|                                 | 1.78 ng/mL      | 2.4             | 1.34 ng/mL                   | 6.0             |
|                                 | 2.9 ng/mL       | 2               | 3.43 ng/mL                   | 3.9             |
| T4                              | 3.31 µg/dL      | 1.6             | 4.7 µg/dL                    | 3.6             |
|                                 | 7.57 µg/dL      | 2.2             | 8.2 µg/dL                    | 2.6             |
|                                 | 14.12 µg/dL     | 1.2             | 15.7 µg/dL                   | 2.5             |
| TSH                             | 0.55 µIU/mL     | 6.4             | 0.1 µIU/mL                   | 13.0            |
|                                 | 5.08 µIU/mL     | 1               | 1.3 µIU/mL                   | 6.3             |
|                                 | 32.73 µIU/mL    | 1.2             | 9.0 µIU/mL                   | 2.0             |
|                                 |                 |                 | 22.5 µIU/mL                  | 1.8             |
| T-UPTAKE                        | 0.97            | 3.3             | 0.71                         | 2.8             |
|                                 | 1.03            | 3.1             | 1.03                         | 2.6             |
|                                 | 1.24            | 2.2             | 1.41                         | 2.4             |

.

- 1) Precision for Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability:
* Standard deviation

:

.

:

ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﺴﺎﻋﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟ

{3}------------------------------------------------

## 2) Correlation:

| COMPARISON<br>METHOD | NUMBER<br>OF<br>SAMPLES | REGRESSION<br>EQUATION<br>y = | CORRELATION<br>COEFFICIENT<br>r | STANDARD<br>ERROR<br>Sy.x | RANGE OF<br>ANALYTE<br>CONCENTRATION |
|----------------------|-------------------------|-------------------------------|---------------------------------|---------------------------|--------------------------------------|
| CEA                  | 360                     | 1.01x - 0.01                  | 0.998                           | 0.116                     | 0.4 - 16.4 ng/mL                     |
| FSH                  | 360                     | 0.99x + 0.02                  | 0.999                           | 0.841                     | 0 - 145 mIU/mL                       |
| Free T4              | 355                     | 0.99x + 0.04                  | 0.988                           | 0.042                     | 0.26 - 1.98 ng/dL                    |
| PSA                  | 200                     | 1.00x + 0.01                  | 1.000                           | 0.066                     | 0 - 20.3 ng/mL                       |
| T3                   | 360                     | 0.99x + 0.02                  | 0.987                           | 0.046                     | 0.48 - 2.16 ng/mL                    |
| T4                   | 356                     | 0.98x + 0.07                  | 0.997                           | 0.141                     | 3.7 - 16.4 µg/dL                     |
| TSH                  | 360                     | 0.99x + 0.02                  | 1.000                           | 0.108                     | 0.23 - 77 µIU/mL                     |
| T-Uptake             | 358                     | 0.98x + 0.03                  | 0.991                           | 0.019                     | 0.61 - 1.41                          |

:

.

:

:

.

y = Bayer Immuno 1™ System upgraded for laboratory automation capability

x = Bayer Immuno 1™ System

: :

: .

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a single snake winding around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 7 1998

Gabriel J. Muraca, Jr. Manager Requlatory Affairs BAYER CORPORATION 511 Benedict Avenue Tarrytown, NY 10591-5097

Re: K983345 Trade Name: Bayer Immuno 1™ System Upgraded for Laboratory Automation Capability Requlatory Class: II Product Code: JJE September 22, 1998 Dated: Received: September 23, 1998

Dear Mr. Muraca:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls: Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

#### Page 2

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page 1 of 1 1. 1. 1. 1.

### 510(k) Number (if known): K983345

# Device Name: Bayer Immuno 1™ System Upgrade for LAS

Indications For Use:

The Bayer Immuno 1™ System has been upgraded for Laboratory Automation capability. Software and Hardware will facilitate communication and robotic transfer of samples to and from a Laboratory Automation system. The Laboratory Automation System (LAS) will coordinate the scheduling, sample identification, and physical transfer of samples, as demonstrated for the following representative methods: CEA, FSH, Free T4, PSA, T3, T4, TSH, and T-Uptake.

_ .

. .. . .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

|  | Concurrence of CDRH, Office of Device Evaluation (ODE) |  |
|--|--------------------------------------------------------|--|
|--|--------------------------------------------------------|--|

| Prescription Use     | ✓ |
|----------------------|---|
| (Per 21 CFR 801.109) |   |

OR

| Over-The-Counter Use |                         |
|----------------------|-------------------------|
|                      | Optional Format 1-2-96) |

(Division Sign-Off)

Division of Clinical Laboratos ! 510(k) Number ________________________________________________________________________________________________________________________________________________________________ 983345 1-2

---

**Source:** [https://fda.innolitics.com/device/K983345](https://fda.innolitics.com/device/K983345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
