NORTHEAST PROTON THERAPY CENTER

K983332 · Massachusetts General Hospital · LHN · Jul 20, 2001 · Radiology

Device Facts

Record IDK983332
Device NameNORTHEAST PROTON THERAPY CENTER
ApplicantMassachusetts General Hospital
Product CodeLHN · Radiology
Decision DateJul 20, 2001
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The NPTC is a facility intended to produce and deliver a proton beam of known energy, intensity and shape for treatment of a patient. It is indicated for use in the therapeutic application of a proton beam for the treatment of localized tumors or other diseases that are susceptible to treatment by radiation. The facility is designed so that it will: (1) create and direct (deliver) the proton beam appropriately to the patient treatment location: (2) produce a transverse and longitudinal dose distribution appropriate for the patient treatment; and (3) deliver the designated dose to the patient's treatment site.

Device Story

Northeast Proton Therapy Center (NPTC) is a facility-based medical charged-particle radiation therapy system. It utilizes beam production equipment (accelerator) to generate a proton beam and beam delivery equipment to shape, direct, and monitor the beam. The system creates a proton beam of specific energy, intensity, and shape to deliver a prescribed dose to a patient's treatment site. It includes a Therapy Safety System (TSS) and a computer-based Therapy Control System (TCS) to manage operations and ensure safety. The system is operated by clinical staff within a specialized facility to treat localized tumors or radiation-susceptible diseases. The output is a controlled proton beam; healthcare providers use this to deliver precise radiation doses, aiming to treat disease while sparing surrounding healthy tissue.

Clinical Evidence

Bench testing only. Performance testing was conducted to demonstrate that the device meets its performance specifications.

Technological Characteristics

Facility-based charged-particle radiation therapy system. Components: beam production equipment (accelerator), beam delivery equipment, Therapy Safety System (TSS), and computer-based Therapy Control System (TCS). System manages proton beam energy, intensity, and shape. Operates via beam delivery systems that shape, direct, and monitor protons.

Indications for Use

Indicated for the therapeutic application of a proton beam for the treatment of localized tumors or other diseases susceptible to radiation therapy.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## JUL 2 0 2001 983332 ### 510(k) SUMMARY Massachusetts General Hospital Northeast Proton Therapy Center Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Jonathan S. Kahan, Esq. Hogan & Hartson, L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004-1109 Phone: (202) 637-5794 Facsimile: (202) 637-5910 Contact Person: same as above September 21, 1998 Date Prepared: #### Name of Device and Name/Address of Sponsor Northeast Proton Therapy Center Massachusetts General Hospital Northeast Proton Therapy Center 30 Fruit Street Boston, MA 02114 #### Classification Name Medical Charged-Particle Radiation Therapy System (21 C.F.R. § 892.5050) #### Predicate Devices - Loma Linda University Medical Center's Proton Beam Therapy Center (1) (K872369) - Harvard University Cyclotron Laboratory's Proton Beam Therapy (2) Center (preamendments device) #### Intended Use The Northeast Proton Therapy Center ("NPTC") is a facility designed to produce and deliver a proton beam of known energy, intensity and shape for treatment of a patient. It is indicated for use in the therapeutic application of a DC - 57171/5 - 0363497.07 {1}------------------------------------------------ proton beam for the treatment of localized turnors or other diseases that are susceptible to treatment by radiation. The facility is designed so that it will: (1) create and direct (deliver) the proton beam appropriately to the patient treatment location: (2) produce a transverse and longitudinal dose distribution appropriate for the patient treatment; and (3) deliver the designated dose to the patient's treatment site. #### Technological Characteristics The equipment installed in the NPTC is comprised of the following two main components: (1) the beam delivery equipment and (2) the beam production equipment. The primary responsibility of the beam delivery equipment is to direct the proton beam to the patient's treatment site within the patient treatment location and to ensure that the patient critical functions are properly and safely carried out. The beam production equipment is necessary to produce the proton beam and direct it to the appropriate treatment room. In addition to the main components, the equipment in the NPTC also includes: (1) a Therapy Safety System ("TSS"); and (2) a computer-based Therapy Control System ("TCS"). #### Substantial Equivalence Discussion The NPTC is substantially equivalent to both the Loma Linda (K872369) and the Harvard Cyclotron Laboratory proton therapy devices, the latter of which is a preamendments device. Like its predicate devices, the NPTC is a facility designed to produce and deliver a proton beam of known energy, intensity and shape for treatment of a patient. It is indicated for use in the therapeutic application of a proton beam for the treatment of localized tumors or other diseases that are susceptible to treatment by radiation. The NPTC and its predicate devices provide the same or substantially equivalent functions, characteristics, and accessories as the NPTC. All of these devices are comprised of beam delivery systems that shape, direct, and monitor the protons delivered to the patient. They are also comprised of beam production equipment that generates the beam used by the beam delivery systems. All of the facilities include patient treatment rooms, but each has a different number of rooms. The beam range in both the NPTC and Loma Linda facilities is similar. Even though the NPTC and its predicate devices use different accelerators. the engineering principles underlying their design are the same. The safety and control systems for the NPTC, Loma Linda, and HCL facilities are equivalent. Although there are some differences between the NPTC and its predicate devices, these differences are minor and raise no new questions of safety and effectiveness. {2}------------------------------------------------ ### Performance Data The submission includes performance testing that Massachusetts General Hospital conducted to demonstrate that the device meets its performance specifications. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ### JUL 2 0 2001 Massachusetts General Hospital % Mr. Jonathan S. Kahan Hogan & Hartson, L.L.P. Columbia Square 555 Thirteenth Street, NW WASHINGTON DC 20004-1109 Re: K983332 Northeast Proton Therapy Center (Proton Therapy Center) Dated: June 4, 2001 Received: June 5, 2001 Regulatory Class: II 21 CFR 892.5050/Procode: 90 LHN #### Dear Mr. Kahan: Down has a market to a finent to market the device referenced above and we have deemning the We have reviewed your Section 5 (1){ } houtleation of much onclosure) to legally marketed prochisat devices device is substantially equivalent (for the indications of the enc device is substantially equivalent (of the macantist to tas sation and the Medical Device Amendments, or to devices of to devices of to devices of to devices of to devices of marketed in interstate commerce prior of the Federal Food, Drug, and Comments Act (Act.). Tou may, that have been reclassified in accordance with the provisions of the are that have been reclassified in accordance will the provisions of the reason on the provisions of the Act . therefore, market the device, subject to the general controls provi therefore, market the device, subject of the general controls of the rear magazine of the ground of the mainst misbranding and adulteration. If your device is classified (see above) into either class II (Premaket Approval), it may be subject If your device is classified (see above) into ellier coass of (optude control in the Code of Federal Regulations, to such additional controls. Existing major regulations arectives with the Current Good Manufacturing Title 21, Parts 800 to 895. A substantialiy equivalent decembriation C Title 21, Parts 800 to 895. A substantialion (QS) for Medical Devices: General regulation (21 CFR Part Practice requirements, as set forth in the Qualine (Device (DVA) Vecif Practice requirements, as set forth in the Qualified (40) for and Drup (402) will bently such assumptions. Failure 820) and that, through periodic QS inspections, the Food an 820) and that, through periodic QS inspections, the Programs Program Prohish further announcements to comply with the GMP regulation may result in regulatory action. In acce to comply with the GMP regulation may result in regulations on more notification sumission doss not concerning your device in the Federal Register. Please not of the Alectio concerning your device in the Federal Register. Ticase noter in the respections under the Electronic Product arroot any obtigations, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in the PDA finding for your device and This letter will allow you to begally marketing your device to real procession in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 01 and additionally for questions on the promotions on the promotions on the promotions If you desire specific advice to your device of Compliance at (301) 584-4639. Additionally, for questions on the promotion diagnostic devices), please contact the Office of Compliation in (301) 594-4639. Also, please not the regulation of version of version on version on version on version on ve entitled, "Misbranding by reference to premarket notification". Other general information on your a entitled, "Misbranding by reterence to premaint notine of Sconel Manufactures assistance at its toll-free number (800) responsibilities under the Act may be obtained from the responsibilities under the Act may be oblamed from the Drilston of Birlines of Suites 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsma Sincerely yours, Nancy Crogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure (s) {4}------------------------------------------------ 510(k) Number (if known): 1 98 333 Z Device Name: Northeast Proton Therapy Center Indications for Use: The NPTC is a facility intended to produce and deliver a proton beam of known The NPTC is a factify intended to product of a patient. It is indicated for use in the energy, intensity and shape for treatment or a parcelor of I localized tumors or therapeutic application of a proton beam for the treatment of the facility is ther diseases that are susceptible to treatment by radiation. The facility is other diseases that are susceptible to treatment by proton heam other diseases that are susceptions to dreated direct (deliver) the proton beam designed so that it will. (1) create and includes a transverse and appropriately to the patient Libaton locator, (or process treatment; and (3) deliver longitudinal dose distribution appropriate for the patient treatment; and (3) deliver the designated dose to the patient's treatment site. # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluations (ODE) | Prescription Use | OR | Over-the-Counter Use | |------------------|----|----------------------| |------------------|----|----------------------| (Per 21 C.F.R. 801.109) (Optional Format 1-2-96) ![Signature](signature.png) (Division Sign-Off) Division of Reproductive, Abd and Radiological Devices 510(k) Number . \\DC . 57171/5 . 0383487.08
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...