← Product Code [LPL](/productcode/LPL) · K983278

# OXYLENS TINTED, PROSTHETIC (HIOXIFILCON B) TINTED, SOFT DAILY WEAR CONTACT LENS (SIGHTED AN NON-SIGHTED EYES) (K983278)

_Alden Optical Labs., Inc. · LPL · Nov 25, 1998 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983278

## Device Facts

- **Applicant:** Alden Optical Labs., Inc.
- **Product Code:** [LPL](/productcode/LPL.md)
- **Decision Date:** Nov 25, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic
- **Attributes:** Therapeutic

## Indications for Use

The OXYLENS PROSTHETIC (hioxifilcon B), Tinted Soft Contact Lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prostheric contact leases for cosmetic management of conditions such as comeal, ins or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system. The OXYLENS TINTED (hioxifilcon B) Spherical Soft Contact Lenses for daily wear, are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected with either a chemical or a heat disinfection system. The OXYLENS TINTED (hioxifileon B) Toric Soft Contact Leases for daily wear, are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters. The lens may be disinfected with either a chemical or a heat disinfection system. The leases are available within a planned replacement program with frequency of replacement determined by the practitioner.

## Device Story

Lathe-cut soft contact lens fabricated from hioxifilcon B polymer; hydrophilic material becomes pliable when hydrated. Lens functions as refractive medium to correct ametropia or as prosthetic device for ocular masking/occlusion. Available in various tint patterns including clear, black pupil, black occluder, and black annular designs. Prescribed by eye care practitioners for daily wear; requires maintenance in compatible storage solution. Benefits include visual acuity correction and cosmetic management of ocular abnormalities. Patient visibility may be reduced in low illumination with dark shades.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, manufacturing/chemistry data, and established safety profiles of predicate devices.

## Technological Characteristics

Hydrophilic polymer (hioxifilcon B); 48% water content; 15 Fatt units oxygen permeability; 1.404 refractive index (hydrated); 1.136 specific gravity (hydrated); >70% light transmission. Lathe-cut manufacturing. Color additives: Vat Brown 1, Vat Blue 6, Vat Green 1, Vat Yellow 3, Vat Orange 1. Compatible with chemical or heat disinfection.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- ALDEN CLASSIC PROSTHETIC ([K980554](/device/K980554.md))
- OXYLENS ([K981252](/device/K981252.md))
- BENZ-G 3X ([K964528](/device/K964528.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K983278

#### Applicant information:

| Date Prepared:  | September 14, 1998                      |
|-----------------|-----------------------------------------|
| Name:           | Alden Optical Laboratories              |
| Address         | 13295 Broadway<br>Alden, New York 14004 |
| Contact Person: | Charles H. Creighton                    |
| Phone Number:   | 800.253.3669                            |
| USA Consultant: | MedVice Consulting, Inc.                |
| Phone Number:   | Martin Dalsing<br>(970) 243-5490        |

#### Device Information:

| Device Classification: | Class II                                                                                        |
|------------------------|-------------------------------------------------------------------------------------------------|
| Classification Number: | LPL                                                                                             |
| Trade Name:            | OXYLENS TINTED, PROSTHETIC (hioxifilcon B), Soft<br>Daily Wear Contact Lens, Tinted (lathe-cut) |
| Classification Name:   | Lenses, Soft Contact, Daily Wear                                                                |

#### Substantially Equivalent Devices:

The OXYLENS TINTED, PROSTHETIC (hioxifilcon B), Soft Daily Wear Contact Lens, Tinted (lathecut) is substantially equivalent to the "ALDEN CLASSIC PROSTHETIC" (polymacon) Tinted, Soft Contact lens (lathe-cut) and the BENZ-G 3X (hioxifilcon B) Soft (Spherical and Toric) Daily Wear Contact Lens, the predicate devices.

{1}------------------------------------------------

#### Device Descriptive Characteristics:

The OXYLENS TINTED, PROSTHETIC (hioxifilcon B), Soft Contact Lens is fabricated from hioxifilcon B, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. A tint mixture containing; Blue 7,16-Dichloro-6, 15-dihydro-5,9,14,18-anthrazinetetrone, Green 16,17 -Dimethoxydinaphtho [1.2.3 -cd:3'.2'.1' -lm] pervlene-5,10-dione, and Brown 16,23 -Dihydrodinaphtho [2,3-a.2 ] 3 ' -i] napth [2',3' ; 6,7] indolo [2,3-c] carbazole-5,10,15,17,22,24-hexone is added to the lens.

The tint mixture (BLACK) is processed into the contact lens to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia.

Tint Patterns Available:

- 1. Clear lens with Black Pupil. Pupil sizes available in 2.0 mm to 12.5 mm.
- 2. Black Occluder Lens. A Central Black area that occludes light. Available to full lens diameter in 0.5 mm increments.
- 3. Black Annular with clear pupil. Black Annular diameter range 7.5 mm to full lens diameter in 0.5 mm increments. Clear Pupil diameter range 2.0 mm to 7.5 mm in 0.5 mm increments.
- 4. Tinted lens with Black Pupil. Uses the Alden Classic Tinted (polymacon 38%) contact lens with black pupil. Pupil sizes available in 2.0 mm to 12.5 mm diameter.

The OXYLENS TINTED may contain a single listed color additive, or may contain a combination of color additives in amounts not to exceed the minimum reasonably required to accomplish the intended coloring effect. The color additives are added to the contact lens for enhancing and/or altering the apparent color of the eve.

In the hydrated state, the OXYLENS TINTED, PROSTHETIC lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a color altering optical surface. The (hioxifilcon B) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped, however, it will return to its proper configuration when completely rehydrated in the proper storage solution.

{2}------------------------------------------------

The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 48% water by weight. The physical properties of the lens are:

| Refractive Index    | 1.515 (dry) 1.404 (hydrated)                                                    |
|---------------------|---------------------------------------------------------------------------------|
| Light Transmission: | greater than 70% T *                                                            |
| Water Content       | 48 %                                                                            |
| Specific Gravity    | 1.308 (dry) 1.136 (hydrated)                                                    |
| Color additives     | Vat Brown 1, Vat Blue 6, Vat Green 1, Vat Yellow3, Vat Orange 1                 |
| Oxygen Permeability | 15 X 10-11 Fatt Units (cm2/sec)(ml O2/ml x mm Hg @ 35°C), (revised Fatt method) |

- * Patients may experience a reduction in visibility while wearing a dark shade of lens in conditions of low illumination.
#### Intended Uses:

The OXYLENS PROSTHETIC (hioxifilcon B), Tinted Soft Contact Lens is indicated for daily wear to enhance or alter the apparent eve color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, in's or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system.

The OXYLENS TINTED (hioxifilcon B) Spherical Soft Contact Lenses for daily wear, are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected with either a chemical or a heat disinfection system.

The OXYLENS TINTED (hioxifilcon B) Toric Soft Contact Lenses for daily wear, are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eves with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters. The lens may be disinfected with either a chemical or a heat disinfection system.

The lenses are available within a planned replacement program with frequency of replacement determined by the practitioner.

{3}------------------------------------------------

#### Substantial Equivalence:

The devices will be manufactured according to specified process controls and a quality assurance program. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Alden Optical Laboratories. The established safety profile (pre-clinical toxicology and manufacturing/chemistry data) of the devices is equivalent to the Alden Classic Tinted (polymacon) 510(k) #K980554, the Oxylens (hioxifilcon B) 510(k) # K981252 and the Benz-G 3X (hioxifilcon B) 510(k) # K964528. Being similar with respect to materials, physical construction and safety & effectiveness to the predicate devices, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above.

The following matrix illustrates the production methods, lens functions and material characteristics of the OXYLENS TINTED, PROSTHETIC (hioxifilcon B), Soft Daily Wear Contact Lens (lathe-cut), as well as the predicate devices.

{4}------------------------------------------------

### Substantial Equivalence Matrix

|     | Characteristic       | OXYLENS<br>PROSTHETIC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | OXYLENS TINTED                                                                                                                                                                                                                               | PREDICATE<br>DEVICE(S)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|-----|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | PRODUCTION<br>METHOD | Lathe-Cut                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Lathe-Cut                                                                                                                                                                                                                                    | Lathe-Cut                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| 2.) | LENS FUNCTION        | Enhance or alter the<br>apparent eye color,<br>including ocular masking,<br>either in sighted or non-<br>sighted eyes that require a<br>prosthetic contact lenses for<br>cosmetic management of<br>conditions such as corneal,<br>iris or lens abnormalities.<br>The lens may also be<br>prescribed for the correction<br>of refractive ametropia<br>(myopia, hyperopia and<br>astigmatism) in aphakic and<br>not aphakic persons or for<br>occlusive therapy for<br>conditions such as diplopia,<br>amblyopia or extreme<br>photophobia. | Refractive medium that<br>focuses light rays from near<br>and distant objects on the<br>retina, while compensating<br>for refractive error,<br>including (astigmatism),<br>including enhancing and/or<br>altering the apparent eye<br>color. | Enhance or alter the<br>apparent eye color,<br>including ocular masking,<br>either in sighted or non-<br>sighted eyes that require a<br>prosthetic contact lenses for<br>cosmetic management of<br>conditions such as corneal,<br>iris or lens abnormalities.<br>The lens may also be<br>prescribed for the correction<br>of refractive ametropia<br>(myopia, hyperopia and<br>astigmatism) in aphakic and<br>not aphakic persons or for<br>occlusive therapy for<br>conditions such as diplopia,<br>amblyopia or extreme<br>photophobia. |
| 3.) | MATERIAL             | Hydrophilic Polymer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Hydrophilic Polymer                                                                                                                                                                                                                          | Hydrophilic Polymer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| a.  | Water Content        | 48%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 48%                                                                                                                                                                                                                                          | 38%, 48%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| b   | Polymer Content      | 52%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 52%                                                                                                                                                                                                                                          | 62%, 52%                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| C   | Polymer              | hioxifilcon B                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | hioxifilcon B                                                                                                                                                                                                                                | Polymacon, hioxifilcon B                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| d.  | DK Value             | 15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 15                                                                                                                                                                                                                                           | 9, 15                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| e   | Refractive Index     | 1.404 (hydrated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1.404 (hydrated)                                                                                                                                                                                                                             | 1.430, 1.404 (hydrated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| f   | Specific Gravity     | 1.136 (hydrated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 1.136 (hydrated)                                                                                                                                                                                                                             | 1.18, 1.136 (hydrated)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| g   | Light Transmission   | greater than 70 % T<br>(N/A<br>for non-sighted eyes)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | greater than 70 % T                                                                                                                                                                                                                          | greater than 70 % T                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| h.  | Color Additives      | Vat Green 1, Green (21<br>CFR 73.3120) CI# 59825                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Vat Green 1, Green (21<br>CFR 73.3120) CI# 59825                                                                                                                                                                                             | Vat Blue 6, Blue (21 CFR<br>73.3119) CI# 69825                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|     |                      | Vat Brown 1, Brown (21<br>CFR 73.3117) CI# 70800                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Vat Brown 1, Brown (21<br>CFR 73.3117) CI# 70800                                                                                                                                                                                             | Not Applicable for Clear<br>Contact Lens                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|     |                      | Vat Blue 6, Blue (21 CFR<br>73.3119) CI# 69825                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Vat Yellow 3, Yellow (21<br>CFR 73.3118) CI# 61725                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|     |                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Vat Orange 1, Orange (21<br>CFR 73.3112) CI# 59105                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|     |                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Vat Blue 6, Blue (21 CFR<br>73.3119) CI# 69825                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of four stylized human profiles facing right, arranged in a stacked formation, with the top profile being the most prominent and the others gradually diminishing in size.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 25 1998

Alden Optical Laboratories, Inc. c/o Mr. Martin Dalsing 623 Glacier Drive Grand Junction, CO 81503

Re: K983278

Trade Name: OXYLENS TINTED PROSTHETIC (hioxifilcon B) Tinted Soft Daily Wear Contact Lens (lathe-cut) Regulatory Class: II Product Code: 86 LPL Dated: September 14, 1998 Received: September 17, 1998

Dear Mr. Dalsing :

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .

{6}------------------------------------------------

Page 2 - Mr. Martin Dalsing

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

ARoey L. Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

#### OXYLENS TINTED, PROSTHETIC (hioxifilcon B), Soft Daily Wear Device Name: Contact Lens, Tinted (lathe cut).

#### INDICATIONS FOR USE:

The OXYLENS PROSTHETIC (hioxifilcon B), Tinted Soft Contact Lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prostheric contact leases for cosmetic management of conditions such as comeal, ins or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system.

The OXYLENS TINTED (hioxifilcon B) Spherical Soft Contact Lenses for daily wear, are indicated for the correction of visual acuity in aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 1.50 diopters or less where the astigmatism does not interfere with visual acuity. The lens may be disinfected with either a chemical or a heat disinfection system.

The OXYLENS TINTED (hioxifileon B) Toric Soft Contact Leases for daily wear, are indicated for the correction of visual acuity in aphakic and not-aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 10 diopters. The lens may be disinfected with either a chemical or a heat disinfection system.

The leases are available within a planned replacement program with frequency of replacement determined by the practitioner.

(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |         |                           |
|--------------------------------------------------------|---------|---------------------------|
|                                                        |         |                           |
|                                                        |         |                           |
| <u></u>                                                |         |                           |
| (Division Sign-Off)                                    |         |                           |
| Division of Ophthalmic Devices                         |         |                           |
| Prescription Use (Per 21 CFR 801.109)                  | X       |                           |
| 510(k) Number                                          | K983278 |                           |
|                                                        | or      |                           |
|                                                        |         | Over-The-Counter Use ____ |

(Optional Format 1-2-96)

OXYLENS TINTED, PROSTHETIC (hionifiloon B), Soft Deily Wear Corned Lans, Ticked (latho ask)

--- 17

---

**Source:** [https://fda.innolitics.com/device/K983278](https://fda.innolitics.com/device/K983278)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
