CONSTELLATION MULTIPLE ELECTRODE PACING AND RECORDING SYSTEM
K983171 · Boston Scientific Corp · MTD · Mar 11, 1999 · Cardiovascular
Device Facts
Record ID
K983171
Device Name
CONSTELLATION MULTIPLE ELECTRODE PACING AND RECORDING SYSTEM
Applicant
Boston Scientific Corp
Product Code
MTD · Cardiovascular
Decision Date
Mar 11, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
For use in right atrial electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Constellation Multiple Electrode systems alone (i.e., linear mapping catheters), and manually generated pacing stimuli.
Device Story
Constellation Multiple Electrode Recording and Pacing Catheter System is a high-density array intracardiac mapping catheter. It is used in right atrial electrophysiology procedures to assist in diagnosing complex arrhythmias. The device functions by providing high-density mapping and manually generated pacing stimuli, which allows for better identification of arrhythmias compared to conventional linear mapping catheters. It is operated by physicians in a clinical setting. The system includes various catheter models (unipolar/bipolar, coated/uncoated), sheaths, a pacing switchbox, and cables. The output is used by the physician to interpret cardiac electrical activity and guide clinical decision-making regarding arrhythmia treatment. The device is intended to benefit patients by improving the diagnostic accuracy of complex cardiac rhythm disorders.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
High-density array intracardiac mapping catheter system. Components include unipolar and bipolar electrodes, coated and uncoated catheter options, sheaths, pacing switchbox, and cables. Operates as an electrophysiology mapping and pacing tool. No software or electronic processing described.
Indications for Use
Indicated for use in right atrial electrophysiology procedures to assist in the diagnosis of complex arrhythmias in patients where conventional linear mapping catheters are insufficient.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 1 1999
Boston Scientific/EP Technologies Steve Jwanouskos Senior Director Benior or Affairs and Quality Compliance 2710 Orchard Parkway San Jose, CA 95134
Re: K983171 K963171
Trade Name: Constellation Multiple Electrode Pacing and conscellations (see attached list of model numbers) Regulatory Class: II Regulatory Crass: II
Product Code: MTD – High density array intracardiac mapping catheter Dated: September 9, 1998 Received: September 10, 1998
Dear Mr. Jwanouskos:
We have reviewed your Section 510(k) notification of intent to we have levrewed your become bove and we have determined the market the device ferenced above afort afor the indications for use device is substanciary oqually marketed predicate devices Stated in the barstate commerce prior to May 28, 1976, the marketed in theorouse of the Medical Device Amendments, or to devices enatement date of che neaders
that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, the reachar rood, bray, subject to the general controls chercrore, marked and the limitations described below. The provisions of the rovisions of the Act include requirements for general concrolo provision of devices, good manufacturing annual regiberaing, and prohibitions against misbranding and adulteration.
The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that incended abound rate form. Therefore, in accordance with Section
{1}------------------------------------------------
Page 2 - Mr. Steve Jwanouskos
513(i)(1)(E) of the Act, the following limitation must appear in the Warnings section of the device's labeling:
The use of this device in conjunction with WARNING: radiofrequency ablation, as part of the diagnosis and treatment of cardiac arrhythmias, may pose an increased risk of adverse events, such as cardiac perforation, myocardial infarction, air embolism, and hematoma requiring surgical repair and/or blood transfusion.
The Warning must be presented within a black box, and the font size of the text should be at least 2 points larger than any surrounding text. The Warning must be present on the first page of your Operator's Manual, and on the packaging for each individual device.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.
Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.
{2}------------------------------------------------
#### Page 3 - Mr. Steve Jwanouskos
If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Susan Alpert Ph.D. M.D.
Susan Alpert, Ph.D., Director Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Device Description
Model Number
:
| Constellation Catheters (Uncoated) | |
|----------------------------------------|--------------|
| 48mm Constellation, unipolar, uncoated | 8048 U |
| 60mm Constellation, unipolar, uncoated | 8060 U |
| 75mm Constellation, unipolar, uncoated | 8075 U |
| 48mm Constellation, bipolar, uncoated | 8048 BU |
| 60mm Constellation, bipolar, uncoated | 8060 BU |
| 75mm Constellation, bipolar, uncoated | 8075 BU |
| Constellation Catheters (Coated) | |
| 48mm Constellation, unipolar, coated | 8048 C |
| 60mm Constellation, unipolar, coated | 8060 C |
| 75mm Constellation, unipolar, coated | 8075 C |
| 48mm Constellation, bipolar, coated | 8048 BC |
| 60mm Constellation, bipolar, coated | 8060 BC |
| 75mm Constellation, bipolar, coated | 8075 BC |
| Sheaths | |
| Straight | 961 |
| Curved | 960 962 |
| Pacing Switchbox | |
| Standard Switchbox | 951 |
| Cables | |
| Universal, primary | 900 |
| Universal, secondary | 901A<br>901B |
| Single spline, primary | 657 |
| Single spline, secondary | 626 |
--
:
ン
{4}------------------------------------------------
#### Attachment 4
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
k983171
Constellation ® Multiple Electrode Recording and Device Name: __ Pacing Catheter System, and Accessories
# Indications for Use:
For use in right atrial electrophysiology procedures to assist in the diagnosis of complex arrhythmias that may be difficult to identify using conventional mapping systems alone (i.e., linear mapping catheters). The Constellation Multiple Electrode systems alone (1.6., incal napping cannotes, and maily generated pacing stimuli.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K. Beth Telle
Prescription Use
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Pcr 21 CFR 801.109)
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.