CT BIOPSYVIEW
K983140 · Philips Medical Systems North America, Inc. · JAK · Nov 4, 1998 · Radiology
Device Facts
| Record ID | K983140 |
| Device Name | CT BIOPSYVIEW |
| Applicant | Philips Medical Systems North America, Inc. |
| Product Code | JAK · Radiology |
| Decision Date | Nov 4, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1750 |
| Device Class | Class 2 |
Intended Use
The CT BiopsyView option is intended to be used for instant display of CT images during interventional procedures The instant display of images allows the operating physician to watch the placement of an interventional instrument in real-time. CT Bropsylview is an option to the TOMOSCAN family of CT systems. TOMOSCAN systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional mages of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles They includes signal analysis and display equipment. patient and equipment supports. component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of anatomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment. The CT BiopsyView option extends the intended use of a TOMOSCAN CT system to include ts use by a physician in the interventional phases of patient conditions and treatment.
Device Story
CT BiopsyView is a software option for Philips TOMOSCAN CT systems; enables real-time CT fluoroscopy during interventional procedures. Input: X-ray transmission data from axial planes. Processing: Rapid image reconstruction at 6 images per second; scan parameters limited to 15-75 mA and 120-140 KV; max exposure 100 seconds. Output: Real-time cross-sectional images displayed on monitor. Used in clinical settings by physicians to guide interventional instruments (biopsy, drainage, therapy). Benefits: Allows physician to visualize instrument placement in real-time, improving accuracy and safety of interventional procedures.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and compliance with established safety standards for CT systems and software-controlled medical devices.
Technological Characteristics
Software option for TOMOSCAN CT systems. Real-time reconstruction capability (6 images/sec). Operates within defined scan parameters (15, 25, 50, 75 mA; 120, 140 KV). Complies with 21 CFR 1020.30/1020.33, UL-2601, UL-187, and IEC601-1. Hazard analysis indicates minor level of concern.
Indications for Use
Indicated for use by physicians during interventional procedures (e.g., biopsy, drainage, therapy control, functional studies) to provide real-time CT fluoroscopic imaging for instrument placement guidance in patients undergoing CT-guided interventions.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- Real-Time Reconstruction System option for Xpress/SX CT System (K950972)
- C.A.R.E. Vision option to the SOMOTOM Plus 4 CT System (K965004)
Related Devices
- K200917 — Azurion R2.1 · Philips Medical Systems Nederland B.V. · May 1, 2020
- K231482 — uCT ATLAS Astound with uWS-CT-Dual Energy Analysis, uCT ATLAS with uWS-CT-Dual Energy Analysis · Shanghai United Imaging Healthcare Co., Ltd. · Nov 29, 2023
- K131263 — XPERGUIDE · Philips Medical Systems Nederland B.V. · Jul 24, 2013
- K965004 — C.A.R.E. VISION FOR THE SOMATOM PLUS 4 CT SCANNER · Siemens Medical Solutions USA, Inc. · Mar 11, 1997
- K172822 — Azurion series R1.2 · Philips Medical Systems Nederland B.V. · Nov 22, 2017
Submission Summary (Full Text)
{0}------------------------------------------------
K983140
# Philips Medical Systems
# 510(K) Summary
In accordance with the requirements of the Safe Medical Device Act, Philips Medical Systems North America Compary herewith submits a 510(K) summary of safety and effectiveness for the following device
| SUBMITTER NAME / ADDRESS: | Philips Medical Systems North America Company<br>710 Bridgeport Avenue<br>Shelton. CT 06484-0917 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT PERSON / TEL NO: | Peter Altman Director of Regulatory Affairs<br>Tel No. (203) 926-7031 |
| DATE SUMMARY PREPARED: | September 4 1998 |
| ESTABLISHMENT NO.: | 1217116 |
| CLASSIFICATION NAME: | Computed Tomography X-ray System<br>(21 CFR 892.1750: Class II: Tier 2: 90JAK) |
| COMMON/USUAL NAME: | Option for CT X-ray System |
| TRADE/PROPRIETARY NAME: | CT BiopsyView |
| PREDICATE DEVICE(S): | (1) Real-Time Reconstruction System option for Xpress/SX<br>CT System Toshiba America Medical Systems, Inc.<br>(2) C.A.R.E. Vision option to the SOMOTOM Plus 4 CT<br>System; Siemens Medical Systems, Inc. |
#### DEVICE DESCRIPTION:
CT BiopsyView is a software option to the Philips TOMOSCAN product family of computed tomography systems.
The TOMOSCAN products are whole body scanners. These systems are already provided with the RapidView Reconstructor (512 x 512 image matrix) option which allows the reconstruction of one image per second.
The CT BiopsyView option provides real-time CT fluoroscopy (6 images per second) for a TOMOSCAN CT system equipped with the RapidView reconstructor. The scan parameters are imited to 15. 25. 50 and 75 mA and 120 and 140 KV. The maximum exposure time is 100 seconds and the remaining time is displayed on the monitor for quick reference.
The CT BiopsyView option is used for interventional procedures, such as biopsy, drainage procedures, therapy control and functional studies.
{1}------------------------------------------------
## INTENDED USE:
The CT BiopsyView option is intended to be used for instant display of CT images during rterventional procedures The instant display of images allows the operating physician to watch the clacement of an interventional instrument in real-time.
CT Bropsylview is an option to the TOMOSCAN family of CT systems. TOMOSCAN systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional mages of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles They includes signal analysis and display equipment. patient and equipment supports. component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of anatomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment.
The CT BiopsyView option extends the intended use of a TOMOSCAN CT system to include ts use by a physician in the interventional phases of patient conditions and treatment.
## SUBSTANTIAL EQUIVALENCE INFORMATION:
The CT BiopsyView option is considered comparable and substantially equivalent to the following predicate devices:
- · Real-Time Reconstruction System option for the Toshiba Xpress/SX CT System (ref: K950972;
- · C A.R.E. Vision option to the Siemens SOMOTOM Plus 4 CT System (ref: K965004)
Each predicate device has been cleared for commercial distribution via its referenced 510 × submission.
#### SAFETY INFORMATION:
The CT BiopsyView option introduces no new safety issues to the TOMOSCAN family of CT systems other then those already known with the TOMOSCAN system in which this option is used. Computed tomography is a mature technology with which industry and users have many years of experience. These devices must comply with the appropriate sections of the Radiation Control for Health and Safety Act. The CT BiopsyView option as part of the TOMOSCAN system and its associated labeling complies with the applicable requirements of the Federal X-ray Performance standards 21CFR 1020.30, 1020.33.
The Philips TOMOSCAN family of CT systems with which the CT BiopsyView option is used are designed to comply with the requirements of Underwriters Laboratories (UL) Standard for Safety of Medical Electrical Equipment (UL-2601) or UL Standard for Safety of X-ray Equipment (UL-187) and be classified by Underwriters Laboratories or an equivalent test laboratory. They are also designed to comply with the requirements of IEC601-1 (Medical Electrical Equipment).
The results of the hazard analysis, combined with the appropriate preventive measures taken ndicate that the device is of minor level of concern as per the August 29, 1991 issue of the "Reviewer's Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review"
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.
Re:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
4 1998 NOV
Frank Gianelli Senior of Regulatory Affairs Specialist Philips Medical Systems 710 Bridgeport Avenue Shelton, CT 06484
K983140 BiopsyView CT Option Dated: September 4, 1998 Received: September 8, 1998 Regulatory class: II 21 CFR 892.1750/Procode: 90 JAK
Dear Mr. Gianelli:
We have reviewed your Section 510(k) notification of intent to markets the referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments; or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good-manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Fecteral Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitn diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".
Sincerely yours,
William Yi
Lillian Yin, Ph.D. Director, Division of Reproductiv Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):
Philips CT BiopsyView Device Name :
Indications For Use :
The CT BiopsyView option is intended to be used for instant display of CT images during interventional procedures. The instant display of images allows the operating physician to watch the placement of an interventional instrument in real-time.
CT BiopsyView is an option to the TOMOSCAN family of CT systems. TOMOSCAN systems are whole body Computed Tomography (CT) systems which are diagnostic X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission of data from the same axial plane taken at different angles. They includes signal analysis and display equipment, patient and equipment supports, component parts and accessories which are used in combination with signal and image processing software to facilitate the relative localization of anatomy with gray-scale representation of density relative to water utilizing Houndsfield indices with or without contrast mediums. They are used for the display, storage and analysis of digital diagnostic CT images. They are intended for use by a physician in the diagnosis and planning phases of patient conditions and treatment.
The CT BiopsyView option extends the intended use of a TOMOSCAN CT system to include its use by a physician in the interventional phases of patient conditions and treatment.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Chid le. leym
OR
ivision Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number
ﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ ﺑﻪ
Prescription Use
( Per 21 CFR 801.109 )
Over-The-Counter Use
(Optional Format 1-2-96)