← Product Code [LHN](/productcode/LHN) · K983024

# PROTON THERAPY SYSTEM (K983024)

_Ion Beam Applications S.A. · LHN · Jul 12, 2001 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K983024

## Device Facts

- **Applicant:** Ion Beam Applications S.A.
- **Product Code:** [LHN](/productcode/LHN.md)
- **Decision Date:** Jul 12, 2001
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The PTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

## Device Story

Proton Therapy System (PTS) generates/delivers proton beams for radiation therapy of localized tumors. System components: cyclotron (beam production); beam delivery equipment (directs beam to treatment site); patient positioning system; Therapy Safety System (automatic/manual shutdown); computer-based Therapy Control System. Used in clinical facilities; operated by trained personnel. Beam delivery utilizes nozzles for scattering/scanning, collimators, and range verifiers. System supports fixed beam stations and isocentric/rotatable gantries. Output: controlled proton beam dose distribution. Clinical benefit: targeted radiation delivery to tumor sites.

## Clinical Evidence

No clinical data provided; substantial equivalence based on technological characteristics and performance comparisons to legally marketed predicate devices.

## Technological Characteristics

System includes cyclotron, beam delivery equipment, patient positioning system, and computer-based Therapy Control System. Features include nozzles for beam scattering/scanning, beam-limiting collimators, and range verifiers. Employs isocentric/rotatable gantries and fixed beam stations. Operates as a networked medical system with integrated safety controls.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Loma Linda University Medical Center Proton Beam Therapy device ([K872369](/device/K872369.md))
- Harvard University Cyclotron Laboratory Proton Beam Therapy device

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUL 1 2 2001

Image /page/0/Picture/1 description: The image contains the word "CONFIDENTIAL" in bold, sans-serif font. Below the word "CONFIDENTIAL" is the alphanumeric string "K983024" in a cursive font. The text is black and the background is white.

### 510(k) SUMMARY

### Ion Beam Applications S.A.

### Applicant

Ion Beam Applications S.A. Chemin du Cyclotron, 3 B-1348 Louvain-la-Neuve, Belgium Attention: Mr. Yves Jongen Phone: 32-10-47-58-90 Facsimile: 32-10-47-58-10 E-mail: jongen@iba.be

### Contact Person and Agent for Ion Beam Applications S.A.

John B. Reiss, Ph.D., J.D. Saul, Ewing, Remick & Saul Centre Square West, 38th Floor Philadelphia, PA 19102 Phone: (215) 972-7124 Facsimile: (215) 972-1906 E-mail: jreiss@saul.com

### Classification Name

Medical charged-particle radiation therapy systems. (21 C.F.R. §892.5050).

### Predicate Devices

The PTS is substantially equivalent to the previously cleared Loma Linda University Medical Center ("Loma Linda") Proton Beam Therapy device (K872369) and the Harvard University Cyclotron Laboratory Proton Beam Therapy device, a pre-1976 device. The PTS and its predicate devices have the same intended use and principles of operation, and are substantially equivalent in terms of performance and technological characteristics.

### Intended Use

The PTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

{1}------------------------------------------------

## JNF JENTIAl

### Technological Characteristics

The device is designed to: (1) create and deliver the proton beam to the patient treatment location; (2) produce a transverse and longitudinal dose distribution appropriate for the patient's treatment; and (3) deliver the designated dose to the patient's treatment site. The PTS has two primary components: (1) the beam delivery equipment, which directs the proton beam to the patient's treatment site within the patient treatment location and ensures the patient critical functions are properly and safely accomplished; and (2) the beam production equipment, which includes a cyclotron and delivery system to produce the proton beam and deliver it to the patient treatment locations. In addition to these primary components, the PTS includes a Therapy Safety System to protect against unsafe conditions, having both automatic and manual controls to shut down the PTS in the event problems occur; and a computer-based Therapy Control System which controls the parameters of the proton beam.

### Substantial Equivalence Discussion

The PTS is substantially equivalent to both the Loma Linda (K872369) and the Harvard Cyclotron Laboratory ("HCL") proton therapy devices. The HCL is a pre-1976 device that was constructed in 1949.

Like its predicate devices, the PTS is a device designed to produce and deliver a proton beam for treatment of a patient. Also like its predicate devices, it is intended for use in the therapeutic application of a proton beam for the treatment of localized tumors or other diseases that are susceptible to treatment by radiation.

The predicate devices also provide the same or substantially equivalent functions, characteristics, and accessories as does the PTS. All these devices are comprised of beam delivery systems which shape, direct, and monitor the protons delivered to the patient. They are also comprised of beam production equipment which generates the beam used by the beam delivery systems.

The technological aspects of a patient treatment consist of protons generated by the beam production equipment, directed to the patient's treatment site by the beam shaping system which is either mounted on a rotatable gantry, or in a fixed position. The patient is put into the correct position relative to the beam by a positioning system.

The facilities include patient treatment rooms, with each having a different number of rooms. The PTS device may service three to seven rooms, the Loma Linda predicate has four rooms and the HCL predicate has two. Like the predicate Loma Linda and HCL devices, the PTS provides fixed beam treatment stations. The PTS also includes treatment rooms which have isocentric/rotatable gantries similar to those used in the Loma Linda facility, but the space enclosed by the gantry is larger than at Loma Linda so that the patient can be rotated horizontally, as at HCL, allowing more choice of treatment direction.

{2}------------------------------------------------

# CONFIDENTIAL

The PTS and predicate Loma Linda devices are equipped with nozzles that provide beam scattering and beam scanning; the nozzles for the HCL predicate use beam scattering. All three devices have beam-limiting collimators and range verifiers.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with three curved lines above them, resembling a bird in flight.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 2 2001

Ion Beam Applications % John B. Reiss, Ph.D., J.D. Saul, Ewing, Remick & Saul Center Square West 1500 Market Street, 38th Floor PHILADELPHIA PA 19102-2186 Re: K983024

Proton Therapy System Dated: April 13, 2001 Received: April 16, 2001 Regulatory Class: II 21 CFR 892.5050/Procode: 90 LHN

### Dear Dr. Reiss:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that keve been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substanially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part I ractive requirements, as sol the missections, the Food and Drug Administration (FDA) will verify such assumptions. Failure o comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro If you don't devices), please contact the Office of Compliance at (301) 594-4639. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely vours

Nancy Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure (s)

{4}------------------------------------------------

CONFIDENT

#### STATEMENT OF INDICATIONS FOR INTENDED USE XI.

510(k) Number (if known): K983024

Page 1 of 1

Device Name:

Proton Therapy System.

Indications for Use:

The PTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Concurrence of CDRH, Office of Device Evaluations (ODE) |                                                                                                               |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Prescription Use                                        | <div style="display:inline-block;"> <div style="display:inline-block;">✓</div> </div> OR Over-the-Counter Use |
| (Per 21 C.F.R. 801.109)<br>(Optional Format 1-2-96)     |                                                                                                               |

  

(Division Sign-Off)

Division of Reproductive, Abdominal,

and Radiological Devices

| 510(k) Number | K983024 |
|---------------|---------|
|---------------|---------|

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

---

**Source:** [https://fda.innolitics.com/device/K983024](https://fda.innolitics.com/device/K983024)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
