The Self Cath is a flexible tubular device to pass fluids to or from the urinary tract. The set is intended to allow for the catheterization of the individual.
Device Story
Self Cath Set is a urinary catheter kit for intermittent catheterization. Device consists of a flexible tube with a funnel, Tiemann tip, and two drainage eyes. Kit configurations include 'HomeSet' and 'Travelset', containing accessories such as drainage bags, antiseptic, gauze, gloves, and cleaning supplies. Used in home or clinical settings to facilitate fluid passage from the urinary tract. Operated by the patient or caregiver. Output is the physical drainage of urine into the provided bag. Benefits include enabling self-catheterization for individuals requiring bladder management.
Clinical Evidence
No clinical data provided; substantial equivalence is based on design, material, and intended use comparisons to the predicate device.
Technological Characteristics
Flexible tubular catheter with funnel and Tiemann tip; two drainage eyes. Kit includes various accessories (drainage bag, antiseptic, gauze, gloves). Materials are equivalent to predicate. No specific ASTM standards or software algorithms mentioned.
Indications for Use
Indicated for individuals requiring urinary tract catheterization to pass fluids to or from the urinary tract.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
Predicate Devices
Rüsch Catheter for intermittent catheterization
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows handwritten text that appears to be a code or identifier. The text includes the characters 'K18 3010' followed by '1 of 2'. The handwriting is somewhat stylized, with variations in stroke thickness and letter formation, giving it a casual yet distinct appearance.
Image /page/0/Picture/1 description: The image shows the logo for RUSCH, with the word "RUSCH" in large, bold, black letters. Below the logo, the words "INTERNATIONAL Group Regulatory Affairs" are printed in a smaller font. Underneath that, the text "A Subsidiary of Teleflex Incorporated (USA)" is printed in an even smaller font.
(603) 532-8211 or 6108
510 (k) Summary
#### Submitter Name, Address, and Date of Submission. 1.
Mr. James R. Whitney Group Regulatory Affairs Associate Rüsch International Tall Pines Park Jaffrey, New Hampshire 03452
Telephone: (603) 532-7706 Facsimile: (603) 532-8211 E-Mail: jrwhitney@compuserve.com
Contact: Same as above
## 2. Name of the Device, Common, Proprietary (if Known), and Classification.
Classification Name: Urological Catheter and Accessories
Common Name: Urinary Catheter
Proprietary Name: Self Cath Set
## 3. Identification of the legally marketed device to which the submitter claims equivalence.
The Self Cath Set is substantially equivalent to the Rüsch Catheter for intermittent catheterization.
#### 4 . Description of the Device.
The catheter consists of a funnel, a clear tube and a soft, Tiemann tip. The tube has two (2) drainage eyes at the distal end that are opposite and offset. The funnel, at the proximal end, is sized to connect to the drainage baq provided in the kit. For the 'HomeSet', the catheter is packed in an inner plastic bag and a Paper-Film outer pouch.
{1}------------------------------------------------
For the 'Travelset', the catheter is packed in the inner plastic bag and is in the sterile paper-film kit tray with the other contents.
1983010 20f2
The kit comes in two configurations. One is a 'HomeSet' and the other is a 'Travelset'. Each set comes in men's and woman's configurations which differs by the length of the catheter.
The 'HomeSet' consists of:
- (1) 1.5 l drainage catch bag with closure cap
- (3) Dab cloths
- (3) compression gauze
- (1) Tweezers
- (1) Table cloth (to set kit contents on)
- (1) disposal Bag
- (1) cleaning wipe
The 'Travelset' consists of:
- (1) Safety Cath
- (1) 1.5 l drainage catch bag with closure cap
- (2) Dab cloths
- (3) compression gauze
- (1) Table cloth (to set kit contents on)
- (1) Chair Cloth (for under the genital area)
- (1) disposal Baq
- (2) Rubber qloves
- (1) cleaning wipe
Attached to the back of the kit is a bottle of water soluble antiseptic (approx 15 ml).
#### 5. Intended Use of the Device.
The Self Cath is a flexible tubular device to pass fluids to or from the urinary tract. The set is intended to allow for the catheterization of the individual.
### Summary of Technological Characteristics. 6 .
The Self Cath is equivalent in design, use and materials to the predicate products.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle or bird with outstretched wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the bird image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 1 2 1999
Miss Karenann Brozowski Group Regulatory Affairs Associate Rüsch International Tall Pines Park Jaffrey, NH 03452
Re: K983010
Self Cath Set/Safety Cath (Urology Drainage Catheter) Regulatory Class: II 21 CFR 876.5130/Procode: 78 FCM, 78 KOD Dated: January 12, 1999 Received: January 13, 1999
Dear Ms. Brozowski:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the tray have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
In addition, we have determined that your device kit contains Povidone Iodine or Benzalkomium Chloride (BZK) which are subject to regulation as drugs.
{3}------------------------------------------------
# Page 2 - Miss Karenann Brozowski
Our substantially equivalent determination does not apply to the drug component of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug component. For information on applicable Agency requirements for marketing these drugs, we suggest you contact:
Director. Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'(21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Stind. A. Ayerm
CAPT. Daniel G. Schultz, M.D. Acting Director Division of Reproductive, Abdominal, Ear. Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name:__________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The Self Cath is a flexible tubular device to pass fluids to or from the urinary tract. The set is intended to allow for the catheterization of the individual.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) Division Sign-Ort)
> ivision of Reproductive. Abdominal, EN and Radiological D 510(k) Number
ption Use ~✓
Prescription Use_ ~ (Per 21 CFR 801.109) OR
Over-The-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.