The VACUTAINER® Brand Blood Collection Syringe is intended to collect whole blood specimens for diagnostic testing which may include: pH, blood gases, electrolytes (including ionized calcium and magnesium), metabolytes, co-oximetry, and other tests.
Device Story
VACUTAINER® Brand Blood Collection Syringe is a sterile, single-use device for collecting whole blood specimens for diagnostic testing. It incorporates dry calcium-balanced lithium heparin (approx. 50 IU/mL) derived from porcine intestinal mucosa; calcium chloride is added to the heparin formulation to allow accurate measurement of ionized calcium and magnesium without heparin interference. Available in three configurations: A-LINE Kit (arterial line aspiration), PRESET™ Kit (preset volume with self-venting membrane), and DRIHEP® PLUS Kit (self-venting membrane, low plunger resistance, auto-stop). Needle options include standard hypodermic or SafetyGlide™ hinge-safety needles. Used by clinicians in healthcare settings to obtain blood samples; the syringe facilitates sample collection, which is then analyzed by laboratory equipment to inform clinical decision-making regarding patient metabolic and gas status.
Clinical Evidence
Three performance studies were conducted: anticoagulant & hemolysis, analytical, and clinical. Results demonstrated equivalence between the principal device (with calcium-balanced lithium heparin) and predicate devices currently marketed in the United States.
Technological Characteristics
Sterile, single-use blood collection syringe. Materials include dry calcium-balanced lithium heparin (porcine-derived) and calcium chloride. Features include self-venting membranes, low plunger resistance, and auto-stop mechanisms. Needle options include standard single lumen or SafetyGlide™ hinge-safety needles. No electronic components, software, or energy sources.
Indications for Use
Indicated for the collection of whole blood specimens for diagnostic testing, including pH, blood gases, electrolytes (ionized calcium and magnesium), metabolites, and co-oximetry. Intended for prescription use.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
Becton Dickinson Division Arterial Blood Gas Syringe (Exempt)
{0}------------------------------------------------
K981922
#### I. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
### 510(k) Summary Of Safety and Effectiveness
# I. General Information
This Summary of Safety and Effectiveness information is being submitted in accordance with the requirements of the SMDA of 1990 and 21 § 807.92
## Establishment:
| Address: | Becton Dickinson VACUTAINER Systems<br>1 Becton Drive<br>Franklin Lakes, NJ 07417-1885 |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Registration Number: | 2243072 |
| Contact Person: | Keith M. Smith<br>Senior Regulatory Affairs Specialist<br>Telephone no.: 201-847-6280<br>Fax No. 201-847-4858 |
| Date of Summary: | August 14, 1998 |
Device
| Trade Name: | VACUTAINER® Brand Blood Collection<br>Syringe |
|----------------------|--------------------------------------------------------------|
| Classification Name: | Tubes, Vials, Systems, Serum Separators,<br>Blood Collection |
- Class II Classification: . None Established under 514 of the Performance Standards: . Food, Drug and Cosmetic Act
{1}------------------------------------------------
### II. Safety and Effectiveness Information Supporting Substantial Equivalence
Substantial Equivalence Declaration:
The term "Substantial Equivalence" as used in this 510(k) Premarket Notification is limited to the definition of Substantial Equivalence found in the Federal Food, Drug, and Cosmetic Act, as amended and as applied under 21 CFR § 807, Subpart E, under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of, substantial equivalence berein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
· Device Description
The VACUTAINER® Brand Blood Collection Syringe is a sterile, single use device designed to collect whole blood specimens for diagnostic testing. The syringe contains dry calcium-balanced lithium heparin (approximately 50 IU per mL of whole blood) derived from porcine intestinal mucosa, as the anticoagulant. Calcium chloride has been added to the formulation to provide a calcium-balanced lithium heparin solution, specifically allowing the measurement of ionized calcium and magnesium. The calcium chloride solution binds to the heparin prior to blood collection allowing electrolytes such as ionized calcium and magnesium in the blood to be measured without interference from the heparin (calcium-balanced).
> VACUTAINER Brand A-LINE Kit
Contains a specifically designed syringe only for aspiration of blood samples from arterial lines.
> VACUTAINER Brand PRESET™ Kit
Contains a specifically designed syringe (may include needle) that can be preset to a desired volume, but permits aspiration when necessary. Includes a venting system that expels residual air through the self-venting membrane (as blood fills the syringe), which ensures rapid filling.
> VACUTAINER Brand DRIHEP® PLUS Kit
Contains a specifically designed syringe (may include needle) with a self-venting membrane, (which seals automatically upon blood contact), and low plunger resistance which permits preset, aspiration, and natural fill sampling. Plunger auto-stop avoids overfill and leakage.
#### Please Note:
The needle component may be one of two configurations. The needle may be a single lumen hypodermic needle or a SafetyGlide™ Needle, a single lumen hypodermic needle with a hinge safety mechanism. The single lumen hypodermic needle is a pre-amendment device for Becton Dickinson and the SafetyGlide™ Needle was cleared on 10/10/95 via 510(k) K951254.
{2}------------------------------------------------
- Intended Use
The VACUTAINER® Brand Blood Collection Syringe is intended to collect whole blood specimens for diagnostic testing which may include: pH, blood gases, electrolytes (including ionized calcium and magnesium), metabolytes, co-oximetry, and other tests.
#### · Synopsis of Performance Study Results
Three separate performance studies (anticoagulant & hemolysis, analytical, and clinical) were done to show the performance and equivalence of the principal device with the new calcium-balanced lithium heparin anticoagulant to the predicate devices currently marketed in the United States.
The VACUTAINER® Blood Collection Syringes are substantially equivalent to the performance of the currently marketed predicate devices as presented in this submission. All results from the three studies show equivalence between the principal device and the predicate device. Therefore, the performance of the VACUTAINER Blood Collection Syringes are substantially equivalent to the predicate devices.
{3}------------------------------------------------
#### · Substantial Equivalence
Based on comparison of the device features, materials, intended use and performance, the VACUTAINER® Blood Collection Syringe can be shown to be substantially equivalent to the commercially available predicate devices indicated in the table below. The predicate devices, K number, and clearance date are also identified in the table below.
| Manufacturer | Predicate Device | K-Number | Clearance Date |
|------------------------------------------------------------|------------------------------------------------------------|----------|---------------------------------------|
| Becton Dickinson<br>Division | Becton Dickinson<br>Division Arterial<br>Blood Gas Syringe | N/A | Exempt from premarket<br>notification |
| RADIOMETER<br>AMERICA INC. | Smooth ETM<br>Arterial Blood<br>Sampler | K896514 | unknown |
| CHIRON/Diagnostics<br>RapidLyte™ Arterial<br>Blood Sampler | VACUTAINER<br>Brand Safety<br>Blood Collection<br>Assembly | K875117 | 2/29/88 |
Kim Reid
8/14/98
---
Date
Keith M. Smith Senior Regulatory Affairs Specialist Becton Dickinson VACUTAINER Systems Becton Dickinson and Company
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle-like symbol with three wavy lines representing the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.
SEP 2 2 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Keith M. Smith Senior Regulatory Affairs Specialist Becton Dickinson VACUTAINER Systems 1 Becton Drive Franklin Lakes, New Jersey 07417-1885
Re : K982922 VACUTAINER® Brand Blood Collection Syringes Regulatory Class: II Product Code: JKA Dated: August 14, 1998 August 19, 1998 Received:
Dear Mr. Smith:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਸ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 -through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification™ (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Dietman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# B. INDICATIONS FOR USE
510(k) Number (if known): k892922
- .---------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Device Name: VACUTAINER® Brand Blood Collection Syringe
Indications for Use:
The VACUTAINER Brand Blood Collection Syringe is intended to collect whole blood specimens for diagnostic testing which may include: pH, blood gases, electrolytes (including ionized calcium and magnesium), metabolytes, co-oximetry, and other tests.
(Please do not Write below this line-continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
Over-the-Counter Use__________________________________________________________________________________________________________________________________________________________
(Per 21 CFR § 801.109)
ef 29
(Optional format 1-2-96)
(Division Sign-Off)
Or
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K982822
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.