← Product Code [KMF](/productcode/KMF) · K982892

# 3M TEGASORB THIN HYDROCOLLOID DRESSING (K982892)

_3M Company · KMF · Oct 26, 1998 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K982892

## Device Facts

- **Applicant:** 3M Company
- **Product Code:** [KMF](/productcode/KMF.md)
- **Decision Date:** Oct 26, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5090
- **Device Class:** Class 1
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

3M Tegasorb™ THIN Hydrocolloid Dressing is intended for use as a dressing for partial thickness dermal ulcers, leg ulcers, superficial wounds, abrasions, first- and second-degree burns, and donor sites. It may also be used as a protective dressing on at-risk, undamaged skin or on skin beginning to show signs of damage from friction or shear.

## Device Story

Sterile wound dressing; consists of hypoallergenic hydrocolloid adhesive with outer clear moisture-vapor-permeable, water/bacteria-impermeable adhesive film. Applied over wound site; interacts with wound fluid to form soft, semi-transparent absorbent mass. Maintains moist wound environment to support healing. Used in clinical or home settings; applied by healthcare professionals or patients. Provides physical barrier and moisture management; aids in protection of compromised or at-risk skin.

## Clinical Evidence

Bench testing only; biocompatibility test results from K924280 support the safety of the dressing.

## Technological Characteristics

Sterile hydrocolloid wound dressing. Components: hypoallergenic hydrocolloid adhesive, outer clear adhesive film. Properties: moisture vapor permeable, water and bacteria impermeable. Form factor: thin dressing.

## Regulatory Identification

A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.

## Predicate Devices

- 3M Tegasorb™ Ulcer Dressing ([K924280](/device/K924280.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K982892

| OCT 26 | 510(k) Summary of Safety and Effectiveness |
|--------|--------------------------------------------|
|--------|--------------------------------------------|

- 3M Medical Products Manufacturer: 3M Center St. Paul, Minnesota 55144-1000
- Requlatory Affairs Anna E. McRight Contact: Requlatory Affairs Specialist 3M Medical Products Resource Division 3M Center, Building 275-3E-08 St. Paul, Minnesota 55144-1000
- Telephone: (651) 733-7948

Date Summary Prepared:

- 3M Tegasorb™ THIN Hydrocolloid Product Trade Name: Dressing
- Wound Dressing Common Name:
- Classification: Wound Dressing
- Predicate Devices: 3M Tegasorb™ Ulcer Dressing (see also, K924280, May 12, 1993, for competitive comparable products).

August 13, 1998

- [3M Tegasorb™ and] 3M Tegasorb™ Thin Description: Hydrocolloid Dressings are sterile wound dressings which consist of a hypoallergenic hydrocolloid adhesive with an outer clear adhesive cover film. The film is moisture vapor permeable and water and bacteria impermeable. The products are not designed, sold or intended for use except as indicated. Over the wound site, [3M Tegasorb and] 3M Tegasorb THIN Hydrocolloid dressings interact with wound fluid to create a soft, semi-transparent absorbent mass. The dressings maintain a moist wound environment, which has been shown to enhance wound healing. (continued)

{1}------------------------------------------------

| Intended Use:            | 3M Tegasorb™ THIN Hydrocolloid Dressing is intended for use as a dressing for partial thickness dermal ulcers, leg ulcers, superficial wounds, abrasions, first- and second-degree burns, and donor sites. It may also be used as a protective dressing on at-risk, undamaged skin or on skin beginning to show signs of damage from friction or shear. |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence: | This premarket notification was submitted for package insert changes only. The comparative products remain identical to those submitted in the premarket notification, K924280, cleared for marketing by FDA on May 12, 1993.                                                                                                                           |

See K924280 which provided Summary of Testing: biocompatibility test results to support 3M Tegasorb™ THIN Hydrocolloid Dressing as acceptable for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features an abstract image of an eagle with three lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the top of the circle.

## OCT 26 1998

Anna E. McRight. 3M Medical Products Division Building 275 - 3E- 08 St. Paul, Minnesota 55144

Re: K982892 Trade Name: 3M Tegasorb™ Thin Hydrocolloid Dressing K982893 Trade Name: 3M Tegasorb™ Hydrocolloid Dressing Regulatory Class: Unclassifed Product Code: KMF Dated: August 13 and 14. 1998 Received: August 17, 1998

Dear Ms. McRight:

We have reviewed your Section 510(k) notifications of intent to market the devices referenced above and we have determined the devices are substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your devices subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:

- These devices may not be labeled for use on third degree burns. 1 .
- 2. These devices may not be labeled as having any accelerating effect on the rate of wound healing or epithelization.
- 3. These devices may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
- These devices may not be labeled as a treatment or a cure for any type of wound. 4.

The labeling claims listed above would be considered a major modification in the intended use of the devices and would require a premarket notification submission (21 CFR 807.81).

The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

{3}------------------------------------------------

Page 2 - Ms. Anna McRight

If vour devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your devices can be found in the Code of Federal Regulations (CFR), Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to a legally marketed predicate device results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html.

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

22892

3M Tegasorb™ THIN Hydrocolloid Dressin 510(k) for Labeling Changes

## 982892

510(k) Number (if known): K924280; May 12, 1993

Device name: 3M Tegasorb™ THIN Hydrocolloid Dressing, Product Number series 90021-90025.

Indications for Use:

The following conditions are considered appropriate for OTC use by the lay person:

Superficial wounds such as cuts, lacerations, or abrasions

- First-degree burns
- Skin needing protection from friction or sheer

The following conditions are considered appropriate for OTC use under the supervision of a health care professional:

- Partial thickness dermal ulcers
- Leg ulcers

・・

Second-degree burns

- Donor sites
[PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED]

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use        |  |
|-------------------------|--|
| OR Over the Counter Use |  |

(Optional Format 1-2-96)
(Division Sign-Off)
of General Restorative Devices

| 510(k) Number | <982892 |
|---------------|---------|
|---------------|---------|

---

**Source:** [https://fda.innolitics.com/device/K982892](https://fda.innolitics.com/device/K982892)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
