DEMOLIZER #47 ONE-GALLON POINT OF GENERATION SHARPS CONTAINER

K982781 · Thermal Waste Technology, Inc. · MMK · Sep 30, 1998 · General Hospital

Device Facts

Record IDK982781
Device NameDEMOLIZER #47 ONE-GALLON POINT OF GENERATION SHARPS CONTAINER
ApplicantThermal Waste Technology, Inc.
Product CodeMMK · General Hospital
Decision DateSep 30, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5570
Device ClassClass 2

Intended Use

The sharps container is intended for the collection of disposable sharps in both clinical and nonclinical healthcare settings, such as laboratories, dentists' offices, doctors' offices, veterinarian offices, corporate clinics, nursing homes, and other types of healthcare centers.

Device Story

The Demolizer #47 is a single-use, non-sterile, one-gallon metal sharps container. It is constructed of tin-plated steel with a friction-fit lid and black foam plastisol sealant. The device is used at the point of generation in clinical and non-clinical settings (e.g., doctors' offices, labs, nursing homes) to collect contaminated sharps, including hypodermic needles, syringes, scalpel blades, lancets, and broken glass. The container features a vertical inlet opening and a built-in vent port. Once closed, the lid is resistant to manual opening, providing a secure containment solution for hazardous medical waste. It is operated by healthcare staff or clinicians to safely dispose of sharps, reducing the risk of accidental injury during waste handling.

Clinical Evidence

No clinical data. Bench testing only.

Technological Characteristics

Material: Tin-plated steel; lap welded. Construction: Three-piece, one-gallon drum. Dimensions: 6-5/8" wide x 7-7/16" high; 2-3/8" inlet. Sealant: Black foam plastisol. Features: Friction-fit lid, 1 mm vent port. Sterilization: Non-sterile. Connectivity: None.

Indications for Use

Indicated for the disposal of contaminated medical sharps, including needles, syringes, slides, and ampoules, in clinical and laboratory healthcare settings.

Regulatory Classification

Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 3 0 1998 ### ATTACHMENT 9 ### (As required by 21 C.F.R. §§ 807.87(h), 807.92) - Manufacturer 1. Contact Person Thermal Waste Technologies, Inc. 19 Stony Hill Road Bethel, CT 06801 (203) 778-2210 Bill Ransone Vice President Sales & Marketing (203) 778-1039 - 3. Device 2. - Demolizer® #47 One-Gallon Point of Trade or Proprietary Name a. Generation Sharps Container Sharps Container Common Name b. Classification Name Accessory to Hypodermic Needle C. II Classification d. - Intended Use: The sharps container is intended for the collection of disposable sharps 4. in both clinical and nonclinical healthcare settings, such as laboratories, dentists' offices, doctors' offices, veterinarian offices, corporate clinics, nursing homes, and other types of healthcare centers. #### క్ Predicate Devices The Demolizer # 47 One-Gallon Point of Generation Sharps Container is substantially equivalent to the 1.0 Gallon Leaktight Locking Top Translucent Red Sharps Disposal Container marketed by Post Medical, Inc. (#K925816), and to the 5-Quart Round Sharps Container marketed by Sage Products, Inc. (#K943659). #### 6.. General Description The Demolizer # 47 One-Gallon Point of Generation Sharps Container is a single use, non-sterile, disposable, point of generation sharps container, constructed out of tin-plated steel. The device has a friction fit lid and black foam plastisol sealant for container closure. It is designed to hold contaminated disposable sharps, such as various-sized hypodermic, intravenous syringes or other medical needles, scalpel blades, disposable knives and lancets, and broken glass such as slides, slip covers and ampoules. The device is not a secondary container. {1}------------------------------------------------ ### 7. Detailed Description The sharps container is a three-piece construction, one-gallon metal drum with a friction fit lid and sealant used to close the container. The minimum nominal body wall thickness is .0094 inches, and the minimum nominal top/bottom thickness is .0097 inches. The container is opaque and measures 6-5/8 inches wide by 7-7/16 inches in height, with a 2-3/8 inch sharps inlet opening in the center of the top of the container. The lid is 2-3/4 inches in diameter, with a thickness of .0088 inches. Lining the perimeter of the posterior side of the lid is a black foam plastisol which acts as an additional safety seal; this foam ensures a closed container. A built-in vent port, approximately 1 mm in diameter, is located on the top of the sharps container. #### 8. Summary of Technological Characteristics The Demolizer # 47 One-Gallon Point of Generation Sharps Container differs slightly from its predicate devices, in that the Demolizer sharps container is constructed of tinplated steel and lap welded. The two predicate devices are constructed of plastic. Further, unlike the predicate devices, once the Demolizer sharps container is closed, it is resistant to manual opening. The Demolizer sharps container is similar to the two predicate devices, in that all are single use, disposable, nonsterile, opaque containers of roughly the same capacity, and all feature vertical inlet openings for the disposal of sharps. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the upper portion of the circle. # MAR 2 7 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Bill Ransone Vice President Sales & Marketing Thermal Waste Technologies, Incorporated 19 Stony Hill Road Bethel, Connecticut 06801 Re: K982781 Trade/Device Name: Demolizer #47 One-Gallon Point of Generation Sharps Container Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Codes: MMK Dated: August 26, 1998 Received: August 31, 1998 Dear Mr. Ransone: This letter corrects our substantially equivalent letter of September 30. 1998. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Ransone Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 http://www.fda.gov/cdrh/organiz.html#OC for OC organization structure). Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Syute Y. Mcheioms Chiu S. Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # ATTACHMENT 11 ### INDICATIONS FOR USE STATEMENT ## (REVISION) Page 1 of 1 3 510(k) Number: K982781 Device Name: Demolizer #47 One-Gallon Point of Generation Sharps Container Indications for Use: This device is intended for the disposal of various-sized contaminated medical sharps (including needles, syringes, slides, and ampoules) in clinical and laboratory healthcare settings. Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division Infection Control, and General Hospital Devices | 510(k) Number | K982781 | |---------------|---------| |---------------|---------| | Prescription Use | | |---------------------------|--| | (Per 21 C.F.R. § 801.109) | | OR | Over the Counter Use | | |----------------------|--| |----------------------|--|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...