← Product Code [ITX](/productcode/ITX) · K982733

# OLYMPUS MH-246R BALLOON SHEATH, FOR BRONCHIAL USE (K982733)

_The Olympus Optical Co. · ITX · Oct 28, 1998 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K982733

## Device Facts

- **Applicant:** The Olympus Optical Co.
- **Product Code:** [ITX](/productcode/ITX.md)
- **Decision Date:** Oct 28, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1570
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The MH-246R Balloon Sheath for bronchial use is designed to be used with the Olympus Ultrasonic Probe UM 2R/3R for intraluminal ultrasonic imaging of the upper airways and tracheobronchial use.

## Device Story

The MH-246R Balloon Sheath is an accessory for ultrasonic probes used in bronchial imaging. It consists of an insertion section (balloon, light shielding cover, insertion tube, adapter) and a connector section (connector body, probe/sheath locking rings, irrigation port). The device is used by clinicians to facilitate intraluminal ultrasonic imaging of the upper airways and tracheobronchial tree. The ultrasonic probe is inserted through the sheath; a water-filled syringe is connected to the irrigation port to inflate the balloon, providing an acoustic coupling medium for the probe. The insertion section is sterile and single-use, while the connector section is reusable after cleaning and sterilization. The device enables high-quality imaging by maintaining proper probe positioning and acoustic contact within the airway, aiding in diagnostic visualization.

## Clinical Evidence

Bench testing only.

## Technological Characteristics

Consists of an insertion section (balloon, light shielding cover, insertion tube, adapter) and a connector section (connector body, locking rings, irrigation port). Operates as an acoustic coupling accessory for ultrasonic probes. Single-use sterile insertion section; reusable connector section. No electronic components; purely mechanical accessory.

## Regulatory Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

## Predicate Devices

- Olympus MH-246R (for GI use), [K961048](/device/K961048.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K982733

## 510(k) Summary Olympus MH-246R Balloon Sheath

| Device Name:                       | Olympus MH-246R Balloon Sheath (for bronchial use)                                                                                                                   |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name:                 | Balloon Sheath                                                                                                                                                       |
| Classification Number<br>and Name: | Class II, 21 CFR 892.1570, Diagnostic Ultrasound<br>Transducer<br>Class II, 21 CFR 874.4680<br>Bronchoscope and accessories                                          |
| Predicate Devices:                 | Olympus MH-246R (for GI use), K961048                                                                                                                                |
| Submitted by:                      | Laura Storms-Tyler<br>Olympus America Inc.<br>Regulatory Affairs<br>Two Corporate Drive<br>Melville, NY 11747-3157<br>phone (516) 844-5688, facsimile (516) 844-5416 |

Summary Preparation Date: August 4, 1998

## Statement of Intended Use

The MH-246R Balloon Sheath for bronchial use is designed to be used with the Olympus Ultrasonic Probe UM 2R/3R for intraluminal ultrasonic imaging of the upper airways and tracheobronchial use.

## Device Description

The MH-246R Balloon Sheath for bronchial use consists of two sections - insertion section and connector section. The insertion section is constructed of a balloon with the light shielding cover, insertion tube, and adapter. The connector section consists of a connector body, probe locking ring, sheath locking ring, and irrigation port.

The insertion section is connected to the connector body through a sheath locking ring, while the ultrasonic probe in inserted into the balloon sheath through a probe locking ring. The water filled syringe is connected to the irrigation port via an extension tube and three-way stopcock. The insertion section with the balloon will be provided sterile and intended for single use only. The connector section can be reused after proper cleaning and sterilization as outlined in the instruction manual.

## General Safety

When compared to the predicate device, the Olympus MH-246R Balloon Sheath does not incorporate any significant change in method or operation, material, or design that could affect the safety or effectiveness.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol with three abstract human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the symbol.

OCT 2 8 1998

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Laura Storms-Tyler Director, Regulatory Affairs Olympus America, Inc. Two Corporate Center Dr. Melville, NY 11747-3157

Re:

K982733 Olympus MH-246R Balloon Sheath, for Bronchial I Ise Dated: August 4, 1998 Received: August 5, 1998 Regulatory class: II 21 CFR 892.1570/Procode: 90 ITX

Dear Ms. Storms-Tyler:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976; the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act. . You may, therefore, market the device, subject to the general controls provisions of the general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR-Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613 Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97)." Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odri/dsmaldsmamain.html".

Sincerely yours,

William Yip, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal. Ear. Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known): _ K982733 Olympus MH-246R Balloon Sheath Device Name: (for bronchial use)

Indications for Use:

්‍යුං

Olympus MH-246R Balloon Sheath have been designed to be used with the Olympus Ultrasonic Probe UM-2R/UM-3R, for intraluminal ultrasonic imaging of the upper airways and the tracheobronchial tree.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) | ✓  |
|------------------------------------------|----|
|                                          | OR |
| Over-The-Counter Use                     |    |

| (Division Sign-Off)                                                |         |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices |         |
| 510(k) Number                                                      | K982733 |
| (Optional Format 1-2-96)                                           |         |

---

**Source:** [https://fda.innolitics.com/device/K982733](https://fda.innolitics.com/device/K982733)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
