MIDAS TOUCH BIPOLAR FORCEPS

K982705 · Iti Medical Technologies, Inc. · GEI · Sep 8, 1998 · General, Plastic Surgery

Device Facts

Record IDK982705
Device NameMIDAS TOUCH BIPOLAR FORCEPS
ApplicantIti Medical Technologies, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateSep 8, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ITI Medical Technologies, Inc. Midas Touch™ Bipolar Forceps are intended for use attached by cable to a standard electrosurgical bipolar general surgery procedure that requires coagulation and or tissue cauterization to achieve hemostasis.

Device Story

Midas Touch™ Bipolar Forceps are surgical instruments used for tissue coagulation and cauterization; device connects via cable to standard electrosurgical bipolar generators. Forceps feature various tip lengths/configurations; constructed from stainless steel or titanium alloy with proprietary non-stick coating; proximal end includes protective plastic insulator sheath and industry-standard connection pins. Operated by trained medical practitioners in clinical settings. Device delivers electrical energy to target tissue to achieve hemostasis. Benefits include non-stick properties reducing tissue adhesion during surgery. Sold non-sterile; compatible with steam and EtO sterilization.

Clinical Evidence

No clinical data. Bench testing only. Performance evaluations conducted using written protocols compared subject device to predicate devices, confirming compliance with ANSI/AAMI HF18-1993 standards and demonstrating equivalent or superior performance.

Technological Characteristics

Materials: Stainless steel or titanium alloy with proprietary non-stick coating. Energy source: Standard electrosurgical bipolar generator. Form factor: Various tip lengths/configurations with plastic insulator sheath. Connectivity: Industry standard electrosurgical cable pins. Standards: ANSI/AAMI HF18-1993. Sterilization: Steam and EtO.

Indications for Use

Indicated for use by trained medical practitioners in general surgery procedures requiring tissue coagulation or cauterization to achieve hemostasis.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Rev2 8 1998 SEP ITI Medical Technologies, Inc. 2452 Armstrong Street Livermore, CA 94550 (925) 371-8305 FAX (925) 371-8222 1982702 # SUMMARY OF SAFETY AND EFFECTIVENESS July 31, 1998 | Contact Person: | Roger W. Werne, Ph.D., President | | |-----------------------|-----------------------------------------------------------------------------|---------| | Common or Usual Name: | Bipolar Forceps | | | Proprietary Name: | Midas Touch™ Bipolar Forceps | | | Classification Name: | Unit, Electrosurgical and Coagulation with accessories<br>21 CFR § 878.4400 | | | Class and Panel: | Regulatory Class II<br>Product Code: GEI General and Plastic Surgery 79 | | | Predicate Devices: | Ethicon Inc. Non-Stick Bipolar Forceps | K973384 | The ITI Medical Technologies, Inc. Midas Touch™ Bipolar Description of Device: Forceps family are similar in design to the predicate devices. They have various tip lengths and configurations, and are available in insulated and non-insulated versions. They are made from a stainless steel or Titanium alloy with a proprietary non-stick coaling. The proximal end has a protective plastic insulator sheath with industry standard clectrosurgery cable connecting pins. Forceps are packaged individually, sold non-sterile, and intended for steam, and EtO sterilization. Statement of intended use: The ITI Medical Technologies, Inc. Midas Touch™ Bipolar Forceps are intended for use attached by cable to a standard electrosurgical bipolar generator in any general surgery procedure that requires coagulation and or tissue cauterization to achieve hemostasis. The intended use is identical to that of the predicate devices, i.e. a standard bipolar forceps for use by a trained medical practitioner. There are no differences in the devices concerning intended use or the safety or effectiveness when used as labeled. {1}------------------------------------------------ ### Statement of technological characteristics: The ITI Medical Technologies, Inc. Midas Touch™ Bipolar Forceps incorporate no significant change in design, materials, energy source or other technological characteristics than those found in the predicate devices. Their form factors are similar to the predicate devices. They can be used with the same standard bipolar electrosurgical generator units and cables as the predicate devices. The only difference between the Midas Touch™ Bipolar Forceps and the predicate devices, other than minor shape differences, is the type of non-stick coating. The ITI forceps are made from a stainless steel or Titanium alloy having ITI's own non-stick coating as compared to the predicate devices: Kirwan, Nickel forceps and Valleylab, Titanium forceps. Special Controls: Although there are no performance standards established by the FDA for bipolar forceps devices, the ITI Medical Technologies. Inc. Bipolar Forceps have been designed to comply with, and are manufactured to pertinent parts of the following standards: #### ANSI/AAMI American National Standard HF18-1993 Electrosurgical devices In addition, the device and its development process also comply with pertinent parts of: - FDA, CDRH, ODE DGRD: October 19, 1993, 510(k) Guideline General Surgical . Electrosurgical Devices and; - FDA, CDRH, ODE DRAERD: August 16, 1995 510(k) Checklist for Endoscopic . Electrosurgical Unit (ESU) and Accessories used in Gastroenterology and Urology. - FDA CDRH ODE March 1995 DRAFT: Labeling Reusable Medical Devices for . Reprocessing in Health Care Facilities: FDA Reviewer Guidance Data regarding special controls are summarized in the submission, and support the safety and efficacy of the device. Comparison tests following written protocols were conducted using the ITI devices and the predicate devices. The performance evaluations indicate that the ITI Medical Technologies, Inc. Midas Touch™ Bipolar forceps met all of the performance requirements, consistently performed within their design parameters; and performed equivalently or superior to the predicate devices, thus demonstrating that they are safe and effective when used as intended. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 8 1998 SEP Dr. Roger W. Werne President and CEO ITI Medical Technologies, Inc. 2452 Armstrong Street Livermore, California 94550 Re: K982705 > Trade Name: Midas Touch Bipolar Forceps Regulatory Class: II Product Code: GEI Dated: July 31, 1998 Received: August 3, 1998 Dear Dr. Werne: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. . . . . . . . . . . . . . . . If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - Dr. Roger W. Werne This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # ITI Medical Technologies, Inc. Bipolar Forceps 510(k) Premarket Notification PREFACE - Page 3 rev2 510(k) Number (if known): _982705_ Device Name: Midas Touch™ Bipolar Forceps # Statement of intended use: The ITI Medical Technologies, Inc. Midas Touch™ Bipolar Forceps are intended for use attached by cable to a standard electrosurgical bipolar general surgery procedure that requires coagulation and or tissue cauterization to achieve hemostasis. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | | |--------------------------------------------------------|----------------| | (Division Sign-Off) | | | Division of General Restorative Devices | <i>K982705</i> | | 510(k) Number | | | Prescription Use <span style="text-decoration: overline;">X</span><br>(Per 21 CFR 801.109) | OR | Over - The - Counter - Use ______ | |--------------------------------------------------------------------------------------------|----|-----------------------------------| | | | (Optional Format 1-2-9) |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...