TENOR SPINAL SYSTEM
K982490 · Sofamor Danek USA,Inc. · KWP · Oct 5, 1998 · Orthopedic
Device Facts
| Record ID | K982490 |
| Device Name | TENOR SPINAL SYSTEM |
| Applicant | Sofamor Danek USA,Inc. |
| Product Code | KWP · Orthopedic |
| Decision Date | Oct 5, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3050 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The TENOR™ Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and (d) who are having the device removed after the development of a solid fusion mass. The TENOR™ Spinal System, when used as a non-pedicle screw fixation system, is intended for the following indications: 1.) Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2.)Pseudarthrosis, 3.) Stenosis, 4.) Spondylolisthesis, 5.) Spinal deformities: scoliosis, kyphosis, lordosis, 6.) Fracture, 7.) Unsuccessful previous attempts at spinal fusion, 8.) Tumor resection.
Device Story
Spinal stabilization system providing temporary, bilateral support to augment spinal fusion. Components include titanium alloy clamps, cross connectors, nuts, and screws. Used with GDLH™ 5.5mm titanium rods and compatible with TSRH® hooks and CROSSLINK® plates. Surgeon assembles components intraoperatively to fit patient-specific anatomy for posterior thoracic and lumbar spine fixation. System facilitates solid fusion mass development; pedicle screw components intended for removal post-fusion. Mechanical testing confirms substantial equivalence to existing spinal systems.
Clinical Evidence
Bench testing only.
Technological Characteristics
Medical grade titanium alloy components including clamps, cross connectors, nuts, and screws. Designed for use with 5.5mm titanium alloy rods. Modular assembly for posterior thoracic and lumbar spinal fixation.
Indications for Use
Indicated for patients requiring spinal stabilization and fusion. Pedicle screw fixation: severe spondylolisthesis (Grades 3-4) at L5-S1, using autogenous bone graft, levels L3-sacrum, removal after fusion. Non-pedicle fixation: degenerative disc disease, pseudarthrosis, stenosis, spondylolisthesis, spinal deformities (scoliosis, kyphosis, lordosis), fracture, failed fusion, or tumor resection.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
- CD™ Horizon Spinal System
- TSRH® Spinal System
- DYNA-LOK® Spinal System
Related Devices
- K991528 — MODIFICATION TO TENOR SPINAL SYSTEM-ROD/BOLT CONNECTOR · Danek Medical, Inc. · May 28, 1999
- K991364 — MODIFICATION TO TENOR SPINAL SYSTEM-PLATES · Sofamor Danek USA,Inc. · May 19, 1999
- K022191 — MODIFICATION TO TENOR SPINAL SYSTEM · Medtronic Sofamor Danek, Inc. · Aug 28, 2002
- K964254 — TITANIUM LIBERTY POSTERIOR SPINAL SYSTEM · Sofamor Danek USA,Inc. · Apr 21, 1997
- K022374 — CADENCE SPINAL FIXATION SYSTEM · Sulzer Spine-Tech · Sep 24, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 5 1998
# TENOR™ Spinal System 510(k) Summary K982490
- I. Company: Sofamor Danek USA 1800 Pyramid Place Memphis, TN 38132 (901) 396-3133
#### Proposed Proprietary Trade Name: II. TENOR™ Spinal System
#### III. Product Description
The TENOR™ Spinal System is a spinal device intended to provide temporary, bilateral stabilization and augment the development of a solid spinal fusion. The system comprises a variety of shapes and sizes of clamps, cross connectors, nuts, and screws made of medical grade titanium alloy. The TENOR™ Spinal System is used in conjunction with GDLH™ 5.5mm titanium alloy rods. Additionally, the TENOR™ Spinal System may be used with titanium alloy TSRH® hooks, TSRH® Low Profile CROSSLINK® plates, and MULTI AXIAL Low Profile MULTI-SPAN™ CROSSLINK® plates for attachment to the posterior thoracic and lumbar spine. These components are assembled to fit the patient's specific anatomic needs.
### IV. Indications
The TENOR™ Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (L5-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and (d) who are having the device removed after the development of a solid fusion mass.
The TENORTM Spinal System, when used as a non-pedicle screw fixation system, is intended for the following indications: 1.) Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), 2.)Pseudarthrosis, 3.) Stenosis, 4.) Spondylolisthesis, 5.) Spinal deformities: scoliosis, kyphosis, lordosis, 6.) Fracture, 7.) Unsuccessful previous attempts at spinal fusion, 8.) Tumor resection.
#### V. Substantial Equivalence
The TENOR™ Spinal System is substantially equivalent to the CD™ Horizon, TSRH®, and DYNA-LOK® Spinal Systems manufactured by Sofamor Danek. Mechanical testing and other information were provided to demonstrate substantial equivalence.
{1}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 5 1998
Richard W. Treharne, Ph.D. Vice President Research and Regulatory Affairs Sofamor Danek 1800 Pyramid Place Memphis, Tennessee 38132
Re: K982490 TENOR™ Spinal System Requlatory Class: II Product Codes: MNH and KWP Dated: July 13, 1998 Received: July 17, 1998
Dear Dr. Treharne:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will werify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{2}------------------------------------------------
Page 2 - Richard W. Treharne, Ph.D.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling. regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
M. Witten, Ph.D., M.D. Director
Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known): K982490
TENOR™ Spinal System Device Name:
## Indications for Use:
The TENORTM Spinal System, when used for pedicle screw fixation, is intended only for patients: (a) having severe spondylolisthesis (Grades 3 and 4) of the fifth lumbar - first sacral (LS-S1) vertebral joint; (b) who are receiving fusions using autogenous bone graft only; (c) who are having the device fixed or attached to the lumbar and sacral spine (levels of pedicle screw fixation may be from L3 to sacrum); and (d) who are having the device removed after the development of a solid fusion mass.
The TENOR™ Spinal System, when used as a posterior, non-pedicle screw fixation system, is intended for the following indications: 1. Degenerative disc disease (as defined by back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies); 2. Pseudarthrosis; 3. Stenosis; 4. Spondylolisthesis; 5. Spinal deformities: scoliosis, kyphosis; 6. Fracture; 7. Unsuccessful previous attempts at spinal fusion; 8. Tumor resection.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Evaluation (ODE
Evaluation (ODE)
Division Sion-Off ivision of General Restorative Devi 510(k) Number
Prescription Use (Per 21 CFR 801.109) (Optional 1-2-96)
**OR**
Over-the-counter Use