The AcroMed DOC™ Ventral Cervical Stabilization System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in whom stability is desired following anterior cervical fusion for the indications listed below. The intended levels for treatment range from C2 to T1. Indications include symptomatic cervical spondylosis, trauma (including fracture), posttraumatic kyphosis or lordosis, tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), re-operation for failed fusion, or instability following surgery for the above indications.
Device Story
DOC Ventral Cervical Stabilization System provides supplemental stability to cervical spine following anterior cervical fusion. System comprises titanium alloy rods, platforms, plates, screws, and cross-connectors; anchored to anterior vertebral surface via screws. Surgeon selects fixed (load-bearing) or unfixed (load-sharing) configuration based on platform locking mechanism. Rods are precontoured to cervical lordosity; may be cut or further bent intraoperatively. Cross-connectors provide torsional stability in long constructs. Device used in clinical setting by surgeons; output is mechanical stabilization of vertebral segments. Benefits include restored spinal alignment and support for fusion healing.
Clinical Evidence
Bench testing only. Mechanical characterization performed via static bending, compression, and torsion testing. Fatigue life testing also conducted to evaluate structural integrity.
Technological Characteristics
Implant grade Titanium alloy (ASTM F136). Components: rods (3mm OD), platforms (lockable/non-locking), plates, screws (solid/hollow), cross-connectors. Mechanical fixation via bicortical or unicortical screw insertion. Manual instrumentation set provided for contouring and assembly. No software or electronic components.
Indications for Use
Indicated for patients requiring stability following anterior cervical fusion at levels C2-T1. Conditions include symptomatic cervical spondylosis, trauma/fracture, posttraumatic kyphosis/lordosis, tumor, degenerative disc disease, failed fusion re-operation, or post-surgical instability. Contraindicated for pedicular screw fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
AcroMed Anterior Cervical Stabilization System (K970955)
AcroMed Anterior Cervical Stabilization System (K970462)
Submission Summary (Full Text)
{1}------------------------------------------------
K982443
# DOC VCSS™ 510(k) Summary
COMPANY:
DePuv Motech AcroMed 3303 Carnegie Avenue Cleveland, Ohio 44115
#### DOC Ventral Cervical Stabilization System TRADE NAME:
CLASSIFICATION:
Orthosis, fixation, spinal cervical intervertebral body. Class II
DESCRIPTION:
The DOC Ventral Cervical Stabilization System, in its fully assembled form, consists of two laterally placed rods joined by platforms, or a plate, which lie on the anterior vertebral surface. The system is anchored to the vertebrae with screws. The implant provides supplemental stability to the cervical spine following anterior cervical fusion. Depending upon the implant configuration, it may serve as either a fixed (load-bearing) or unfixed (load-sharing) construct.
The following types of components are available in the system: Rods, Platforms, Plates Screws and Cross Connectors. All components are manufactured from implant grade Titanium alloy which conforms to ASTM F136 specifications. Assemblies are also available for convenience of the surgeon. These are pre-assembled partial constructs consisting of one Platform, one Cross Connector, two Rods and two Locking Screws.
An instrument set is available specifically designed for use with the DOC Ventral Cervical Stabilization System.
Rod:
Precontoured Rods comprise the longitudinal structures of the system. All Rods have a 3mm outer diameter and an integral head. Eight lengths are available: 21mm, 36mm, 43mm, 51mm, 58mm, 65mm, 85mm and 110mm. Two rods are required for each construct. Rods may be cut to the appropriate length during the procedure if required. The Rods are precontoured to a curvature consistent with the cervical lordosity, but additional contouring using the Construct Bender contained in the instrument set may be performed if required.
{2}------------------------------------------------
## Platform:
Five Platform designs are available which are designed to connect the Rods and anchor to the vertebral bodies with bone screws. All Platforms may be used with either Solid Bone Screws (for bicortical fixation) or the combination of Outer and Inner Bone Screws (a hollow outer bone screw used with a solid inner locking screw designed for unicortical fixation). Screw selection is determined by physician preference. All Platforms are designed to be used with two Rods in forming the construct.
Platforms are available in Lockable and Non-Locking designs. Lockable Platforms are recommended for use at the caudal end of all constructs. If a fixed configuration is preferred to a non-fixed (i.e., load-sharing) configuration, a Lockable Platform must also be utilized at the rostral end of the construct. If a load-sharing configuration is desired, the rostral-most Platform is not locked to the Rods.
Three Platforms are designed for angled insertion of the screw into the vertebral body. The resulting screw angle is thirty degrees and may be directed either rostrally or caudally. These Platforms have a fin feature on one side which engages the vertebra from within the disk space for optimal positioning and fixation. Two of these three Platforms are lockable and thus may be utilized at either the rostral or caudal end of the construct. Lockable Platforms (Type C and Type H) have extensions which accept Platform / Cross Connector Locking Screws for securing the Platform to the Rods. Selection of Type C versus Type H Platforms is based on physician preference and patient anatomy. The third Platform for Angled Screws does not lock to the Rods, and thus is not designed for use at the caudal end of the construct.
Two Platforms are designed for straight (perpendicular) insertion of the screw into the They are available in both lockable and non-locking designs. vertebral body. As described above, the Lockable Platform for Straight Screws may be utilized at either the rostral or caudal end of the construct. The Non-Locking Platform for Straight Screws does not lock to the Rods, and thus is not designed for use at the caudal end of the construct.
One Platform design is available for fixation of the construct to an intervertebral graff. This Platform features a horizontal slot and spherical nest to accommodate variation in screw entry site and in trajectory into the graft.
{3}------------------------------------------------
#### Assembly:
As mentioned in the overall device description, Assemblies are pre-assembled, partial constructs consisting of one Platform, one Cross Connector, two Rods and two Platform / Cross Connector Locking Screws. They are provided for the convenience of the surgeon. Assemblies are provided in five lengths ranging from 36mm. All five available Assemblies utilize a Platform for Angled Screws, Lockable, Type H.
#### Plates
Plates are single piece structures comprised of laterally placed longitudinal sections The horizontal sections feature two screw holes for and horizontal sections. receiving an outer and inner bone screw. Different plate sizes feature different overall lengths as well as different spacing of the horizontal sections to accommodate various vertebral sizes. Plates substitute for rods and platforms in constructs, and constructs featuring plates serve only as fixed configurations.
#### Cross Connector:
Cross Connectors are used to connect and lock to the two Rods. They have a "band clamp" design in which the upper and lower halves of the Cross Connector are squeezed together by the Platform / Cross Connector Locking Screws, clamping the Cross Connector to the Rod. The Cross Connector does not provide for fixation to the vertebra. It is intended to provide torsional stability in longer constructs.
#### Outer Bone Screw:
Outer Bone Screws are designed to be used in conjunction with Inner Bone Screws. They are available in three diameters: 3.75mm, 4.0mm and 4.35mm. Each diameter is available in four lengths: 10mm, 12mm, 14mm and 16mm. The Outer Bone Screw has an expanding screw head which locks to the Platform. It also has a screw tip which, like the screw head, expands upon full insertion of the Inner Bone Screw. The combination of Outer and Inner Bone Screws is intended for unicortical fixation.
#### Inner Bone Screw:
Inner Bone Screws are available in four lengths: 10mm, 12mm, 14mm and 16mm, Insertion of the Inner Bone Screw into the Outer Bone Screw results in expansion of the Outer Bone Screw head (thus, locking the screw to the Platform) as well as expansion of the Outer Bone Screw tip.
{4}------------------------------------------------
#### Solid Bone Screw:
Solid Bone Screws are available in two diameters: 3.75mm and 4.0mm. Each diameter is available in eight lengths ranging from 16mm to 24mm. The Solid Bone Screw is designed for bicortical fixation.
### Platform / Cross Connector Locking Screw:
The Platform / Cross Connector Locking Screw is designed for two purposes. First, it is used to tighten together the two sides of the Cross Connector ("band clamp" design), thus locking it to the Rods. Second, it is used to lock a Platform to the Rods. It is available in only one size.
#### Graft Screw:
The Graft Screw is a solid screw design which is specifically designed to be used with the Platform for Graft Fixation. It is available in only one size, with a diameter of 3.5mm and a length of 10mm. The head of the Graft Screw is designed to match the Platform for Graft Fixation.
#### PERFORMANCE DATA:
#### Non-Clinical:
Static bending compression and torsion were performed on the system to characterize its mechanical properties. Additionally; testing was also performed to characterize fatigue life.
#### INTENDED USE:
The DOC Ventral Cervical Stabilization System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in whom stability is desired following anterior cervical fusion. The intended levels for treatment range from C2 to T1.
Indications include symptomatic cervical spondylosis, trauma (including fracture), posttraumatic kyphosis or lordosis, tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), reoperation for failed fusion, or instability following surgery for the above indications.
#### SUBSTANTIAL EQUIVALENCY
The AcroMed Anterior Cervical Stabilization System: K970955 and K970462
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Jul 2 9 1998
Mr. William Christianson Vice President, Regulatory Affairs DePuy Motech AcroMed 3303 Carnegie Avenue Cleveland, Ohio 44115
Re : K982443 DOC™ Ventral Cervical Stabilization System Requlatory Class: II Product Code: KWQ July 9, 1998 Dated: Received: July 14, 1998
Dear Mr. Christianson:
We have reviewed your Section 510(k) notification of intent to market the device system referenced above and we have determined the device system is substantially equivalent (for the indications for use stated in the enclosure) to device systems marketed in interstate commerce prior to May 28, 1976 or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act You may, therefore, market your device system subject (Act). to the general controls provisions of the Act and the limitations identified below.
The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Note that labeling or otherwise promoting this device system for pedicular screw fixation/attachment would cause the device system to be adulterated under 501(f) (1) of the Act.
FDA identifies that any device system, if intended for use in pedicular screw fixation/attachment, except for some limited indications, would be found not substantially equivalent and would be a class III device under Section 513(f) of the Act. Class III devices are required to have an approved premarket approval (PMA) application prior to marketing. Accordingly:
{6}------------------------------------------------
#### Page 2 - Mr. William Christianson
- You may not label or in any way promote this device 1. system for pedicular screw attachment to, or fixation of the cervical, thoracic or lumbar vertebral column. Therefore, in order to prevent off-label promotion, the package insert must include the following statement, This device system is not approved for screw "WARNING: attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.";
- All labeling for this device system, including the 2. package label, must state that there are labeling The package insert must prominently state limitations. that the device system is intended for the specific use(s) described in the enclosure only; and
- Pedicular screw fixation/attachment to the cervical, 3. thoracic or lumbar vertebral column, except for limited indications, of any device system is considered investigational and may only be investigated as a significant risk device in accordance with the investigational device exemption (IDE) regulations under All users of the device system for 21 CFR, Part 812. pedicular screw fixation/attachment must receive approval from their respective institutional review boards (IRBs) and the Food and Drug Administration (FDA) prior to conducting an investigation.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਚੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. - In addition; FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{7}------------------------------------------------
Page 3 - Mr. William Christianson
FDA advises that the use of the subject device system and/or device components with any other device components but those identified in this 510(k) would require submission of a new 510(k) providing documentation of design, material, and labeling compatibility between the device components. Mechanical testing of a spinal system consisting of the subject device components and other device components, whether yours or other manufacturers', may also be required.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification immediately. An FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
Page 1 of _1
510(k) Number (if known): K982443
Device Name: __ DOC VCSS System™
Indications for Use;
The AcroMed DOC™ Ventral Cervical Stabilization System is intended for anterior cervical intervertebral body fixation. This system is indicated for patients in whom stability is desired following anterior cervical fusion for the indications listed below. The intended levels for treatment range from C2 to T1.
Indications include symptomatic cervical spondylosis, trauma (including fracture), posttraumatic kyphosis or lordosis, tumor, degenerative disc disease (defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies), re-operation for failed fusion, or instability following surgery for the above indications.
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Sign-Off | |
|---------------|--------------------------|
| Device Name | Oral Restorative Devices |
| 510(k) Number | K98243 |
| Prescription Use (Per 21 CFR 801.109) | X |
|---------------------------------------|-----------------------------------------------|
| | OR |
| | Over-The-Counter Use (Optional Format 1-2-96) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.