BAYER ADVIA IMS SYSTEM (IN VITRO DIAGNOSTIC SYSTEM)
Applicant
Bayer Corp.
Product Code
CJO · Clinical Chemistry
Decision Date
Jan 29, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1050
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Bayer ADVIA IMS alkaline phosphatase (ALP) assay is an in-vitro diagnostic device intended to measure ALP in human serum or plasma. Such measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. This diagnostic method is not intended for use on any other diagnostic system. This in vitro method is intended to quantitatively measure calcium in human serum and plasma on the Bayer ADVIA IMS. Measurements of calcium are used in the diagnosis, monitoring and treatment of a variety of diseases including parathyroid disease, a variety of bone diseases, chronic renal disease and tetany. The Bayer ADVIA IMS Glucose assay is an in vitro diagnostic device intended to measure glucose in human serum or plasma. Such measurements are used in the diagnosis, monitoring and treatment of carbohydrate metabolism disorders, including diabetes mellitus and neonatal hypoglycemia. This diagnostic method is not intended for use on any other diagnostic system. The Bayer ADVIA IMS Ferritin assay is an in vitro diagnostic device intended to quantitatively measure ferritin (an iron-storage protein) in human serum. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency. This diagnostic method is not intended for use on any other diagnostic system. The Bayer ADVIA IMS hCG assay is an in vitro diagnostic device intended to quantitatively measure total beta (8) in human chorionic gonadotropin (hcG) in human serum. Measurements of human chorionic gonadotropin are used in the detection of pregnancy. This diagnostic method is not intended for use on any other diagnostic system. The Bayer ADVIA IMS TSH assay is an in vitro diagnostic device intended to quantitatively measure thyroid stimulating hormone (TSH)in human serum. This assay allows the determination of TSH with 3d generation sensitivity of less than 20% total coefficient variation (CV) at 0.01 to 0.02 mIUL or uiU/mL, as defined by the American Thyroid Association. Measurements thyroid stimulating hormone are used in the diagnosis of thyroid or pituitary disorders. This diagnostic method is not intended for use on any other diagnostic system.
Device Story
ADVIA IMS is an integrated modular system for in vitro diagnostic analysis of human serum or plasma. System processes samples to measure ALP, Calcium, Glucose, Ferritin, hCG, and TSH. Operation involves automated reagent-sample mixing and photometric or immunoassay detection. Used in clinical laboratory settings by trained technicians. Output provides quantitative analyte concentrations to clinicians for diagnosis, monitoring, and treatment of metabolic, endocrine, and iron-related disorders. System performance validated via correlation with predicate methods and interference testing against hemoglobin, bilirubin, and lipemia.
Clinical Evidence
Bench testing only. Performance validated through comparative studies against predicate devices using clinical serum/plasma samples. Metrics include correlation coefficients (r=0.943 to 1.0), regression equations, total precision (CV%), and analytical range comparisons. Interference studies evaluated effects of hemoglobin, bilirubin, and lipemia on analyte recovery, showing minimal percentage changes.
Technological Characteristics
Automated modular diagnostic system. Reagents and calibrators specific to ADVIA IMS. Detection principles include photometric and immunoassay methods. Connectivity is integrated within the modular system. Software-controlled analysis. Sterilization not applicable (reagents).
Indications for Use
Indicated for patients requiring diagnostic testing for liver, bone, parathyroid, intestinal diseases (ALP); calcium metabolism disorders including parathyroid, bone, renal disease, tetany (Calcium); carbohydrate metabolism disorders including diabetes, neonatal hypoglycemia (Glucose); iron metabolism disorders including hemochromatosis, iron deficiency (Ferritin); pregnancy detection (hCG); and thyroid or pituitary disorders (TSH).
Regulatory Classification
Identification
An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
Predicate Devices
Bayer Chem 1 ALP(AMP) method
Technicon CHEM 1 Calcium method
Bayer Technicon Chem 1 Glucose method
Immuno 1 Ferritin assay
Bayer Immuno 1 hCG Assay
Bayer Immuno 1 TSH Assay
Submission Summary (Full Text)
{1}------------------------------------------------
# Alkaline Phosphatase (AMP) Method for the ADVIA IMS Systems
Listed below is a comparison of the performance of the Bayer ADVIA Alkaline Phosphatase (AMP) method and a similar device that was granted clearance of substantial equivalence (Bayer Chem 1 ALP(AMP) method). The information was extracted from the Bayer ADVIA ALP(AMP) method and Bayer Chem 1 method sheet.
## INTENDED USE
The Bayer ADVIA IMS alkaline phosphatase (ALP) assay is an in-vitro diagnostic device intended to measure ALP in human serum or plasma. Such measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. This diagnostic method is not intended for use on any other diagnostic system.
| ALP(AMP) METHOD: | ADVIA IMS | CHEM 1 | | |
|---------------------------------------------|---------------------------------------------------------------------------------------------|---------------|------------|------|
| Part Number: | Reagents B41-3718-23 | T01-1459-53 | | |
| Analytical Range: | 0 to 2800 U/L | 0 to 2250 U/L | | |
| Precision (Total): | mean (U/L) | % CV | mean (U/L) | % CV |
| | 74 | 3.6 | 65 | 3.8 |
| | 160 | 2.9 | 263 | 3.3 |
| | 420 | 2.1 | 547 | 3.2 |
| Regression Equation: (serum) | $y = 0.96x + 2.0$ | | | |
| where: | y = ADVIA IMS<br>x = Chem 1<br>n = 74<br>r = 0.999<br>Sy.x = 11.2<br>range = 36 to 1599 U/L | | | |
| Regression Equation: (plasma qualification) | $y = 1.06x - 0.9$ | | | |
| where: | y = plasma<br>x = serum<br>n = 59<br>r = 0.999<br>Sy.x = 0.93<br>range = 33 to 111 U/L | | | |
{2}------------------------------------------------
# Interference
| | Interfering Substance<br>Concentration | ALP(AMP) | Effect<br>% Change |
|--------------------------|----------------------------------------|----------|--------------------|
| Hemoglobin | 500 mg/dL | 97 U/L | 9 |
| Bilirubin (conjugated) | 20 mg/dL | 100 U/L | 8 |
| Bilirubin (unconjugated) | 25 mg/dL | 107 U/L | 1 |
| Lipemia (Triglycerides) | 1000 mg/dL | 110 U/L | 3 |
Gabriel J. Muracy Jr.
Gabriel J. Muraco, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
6/30/98
Date
Date
00168
{3}------------------------------------------------
### Calcium Method for the Bayer ADVIA Integrated Modular System (IMS)
Listed below is a comparison of the performance between the Bayer ADVIA IMS Calcium method and a similar device that was granted clearance of substantial equivalence (Technicon CHEM 1 Calcium method). The information used in the Summary of Safety and Effectiveness was extracted from the Bayer ADVIA IMS Calcium method sheet and the CHEM 1 Calcium method sheet.
#### INTENDED USE
This in vitro method is intended to quantitatively measure calcium in human serum and plasma on the Bayer ADVIA IMS. Measurements of calcium are used in the diagnosis, monitoring and treatment of a variety of diseases including parathyroid disease, a variety of bone diseases, chronic renal disease and tetany.
| METHOD | | ADVIA IMS | CHEM 1 |
|-------------------|-------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------|
| Part No. | Reagents<br>Calibrators | B41-3725-25<br>T03-1291-62 | T01-1614-01<br>T03-1291-62 |
| Analytical Range | | 0 to 14.0 mg/dL | 1 to 15 mg/dL |
| Precision (Total) | | 1.1% @ 8.4 mg/dL<br>2.2% @ 10.4 mg/dL<br>0.9% @ 13.6 mg/dL | 2.3% @ 8.4 mg/dL<br>2.5% @10.6 mg/dL<br>2.2% @ 13.4 mg/dL |
| Correlation | | y=0.96-0.8 mg/dL<br>where<br>y=ADVIA IMS<br>x=CHEM 1<br>n=48<br>r=0.996<br>Sy.x=0.01 mg/dL | |
| Plasma/Serum Equivalence | Plasma | Serum | Difference |
|--------------------------|-----------|-----------|------------|
| | 8.4 mg/dL | 8.4 mg/dL | 0.00 |
| Within-run CV | 0.75% | 0.73% | 0.02% |
{4}------------------------------------------------
Interfering Substances
| Bilirubin (unconjugated) 25 mg/dL | 1.1% effect change @ 9.0 mg/dL Ca |
|-----------------------------------|-----------------------------------|
| Hemoglobin 1000 mg/dL | 3.0% effect change @ 8.9 mg/dL Ca |
| Triglycerides 500 mg/dL | 3.0% effect change @ 9.1 mg/dL Ca |
Gabriel J. Muraca Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
6/30/98
__
Date
Date
000170
ﻪ ﻣﺴ
{5}------------------------------------------------
## Glucose Method for the ADVIA IMS Systems
Listed below is a comparison of the performance of the Bayer ADVIA IMS Glucose method and a similar device that was granted clearance of substantial equivalence (Bayer Technicon Chem 1 Glucose method). The information was extracted from the Bayer ADVIA IMS Glucose method sheet.
#### INTENDED USE
The Bayer ADVIA IMS Glucose assay is an in vitro diagnostic device intended to measure glucose in human serum or plasma. Such measurements are used in the diagnosis, monitoring and treatment of carbohydrate metabolism disorders, including diabetes mellitus and neonatal hypoglycemia. This diagnostic method is not intended for use on any other diagnostic system.
| GLUCOSE METHOD: | ADVIA IMS | | CHEM 1 | |
|------------------------------------------------|-------------------------|-------------------|-----------------|------|
| Part Number: | Reagents B41-3732-26 | | T01-1460-53 | |
| | Calibrators T03-1291-62 | | T03-1291-62 | |
| Analytical Range: | 0 to 800 mg/dL | | 0 to 675 mg/dL | |
| Precision (Total): | mean<br>(mg/dL) | % CV | mean<br>(mg/dL) | % CV |
| | 48 | 3.66 | 87 | 2.7 |
| | 97 | 3.03 | 262 | 2.9 |
| | 293 | 1.80 | 305 | 2.3 |
| | | | 562 | 1.3 |
| Regression Equation:<br>(serum) | y = $1.03x + 0.71$ | | | |
| where: | y | = ADVIA IMS | | |
| | x | = Chem 1 | | |
| | n | = 58 | | |
| | r | = 0.993 | | |
| | Sy.x | = 11.4 | | |
| | range | = 51 to 534 mg/dL | | |
| Regression Equation:<br>(plasma qualification) | y = $0.97x + 4.12$ | | | |
| where: | y | = plasma | | |
| | x | = serum | | |
| | n | = 60 | | |
| | r | = 0.943 | | |
| | Sy.x | = 4.1 | | |
range = 58 to 174 mg/dL
{6}------------------------------------------------
# Interference
| | Interfering Substance<br>Concentration | Glucose<br>Concentration | Effect<br>% Change |
|--------------------------|----------------------------------------|--------------------------|--------------------|
| Hemoglobin | 1000 mg/dL | 297 mg/dL | 0.5 |
| Bilirubin (conjugated) | 25 mg/dL | 288 mg/dL | 0.5 |
| Bilirubin (unconjugated) | 25 mg/dL | 271 mg/dL | 1.3 |
| Lipemia (Triglycerides) | 500 mg/dL | 274 mg/dL | 9.5 |
Gabriel J. Munoz, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
6/30/98
Date
00172
{7}------------------------------------------------
# Ferritin Method for Bayer ADVIA™ MODULAR SYSTEM
Listed below is a comparison of the performance between the ADVIA Ferritin method and a similar device that was granted clearance of substantial equivalence (Immuno 1 Ferritin assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA ferritin method sheet and the Immuno 1 ferritin method sheet.
#### INTENDED USED
This in vitro method is intended to quantitatively measure ferritin, an iron storage protein, in human serum using Bayer Immuno 1 ferritin reagents on a Bayer ADVIA Modular System. Measurements of ferritin are used in the diagnosis and treatment of iron deficiency or iron overload disease states.
| METHOD | Immuno 1 Ferritin<br>(predicate device) | | ADVIA Ferritin | |
|--------------------------|-----------------------------------------|----------------------------------------|----------------|--------------------------------------|
| Part No. | Reagents | T01-2863-51<br>Calibrators T03-3251-01 | Reagents | B42-3902-2<br>Calibrators B43-3933-0 |
| Minimum Detectable Conc. | 0.3 ng/mL | | 0.06 ng/mL | |
| Precision (Total CV%) | 21 ng/mL | 7.1% | 28 ng/mL | 4.1% |
| | 148 ng/mL | 5.0% | 108 ng/mL | 4.8% |
| | 344 ng/mL | 5.0% | 245 ng/mL | 5.0% |
Correlation
y = 1.040 x - 6.132
| where | |
|-------|----------------------------|
| y = | Bayer ADVIA Modular System |
| x = | Bayer Immuno 1 System |
| n = | 50 |
| r = | 0.993 |
| Syx = | 57.50 ng/mL |
{8}------------------------------------------------
### Interference
| | Interfering Substance<br>Concentration | Ferritin<br>concentration | Ferritin (Neat<br>concentration) | Effect<br>% Change |
|-------------------------|----------------------------------------|---------------------------|----------------------------------|--------------------|
| Hemoglobin | 1000 mg/dL | 179.8 ng/mL | 178.3 ng/mL | 0.8 |
| Bilirubin | 27.5 mg/dL | 34.74 ng/mL | 36.01ng/mL | 3.5 |
| Urea Nitrogen | 200 mg/dL | 175.1 ng/mL | 182.0 ng/ml | 3.8 |
| Lipemia (Triglycerides) | 1110 mg/dL | 212.6 ng/mL | 222.6 ng/ml | 4.5 |
Gabriel J. Muraca, Jr.
Manager Regulatory Affairs
Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
6/30/98
---
Date
Date
{9}------------------------------------------------
#### hCG Assay for Bayer ADVIA® Integrated Modular System
Listed below is a comparison of the performance between the ADVIA hCG method and a similar device that was granted clearance of substantial equivalence (Bayer Immuno 19 hCG Assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA hCG insert and the Immuno 1° hCG Assay method sheet.
#### INTENDED USED
This in viro method is intended to quantitatively measure hCG, human chorionic gonadotropin, in human serum using ADVIA hCG Assay on a Bayer ADVIA® Integrated Modular System. Measurements of hCG are used in the detection of pregnancy.
| METHOD | ADVIA hCG Assay | Immuno 1 hCG Assay<br>(predicate Device) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Part No. | Reagents B42-3907-43<br>Calibrators B43-3941-01 | Reagents T01-2966-5<br>Calibrators T03-3148-0 |
| Minimum Detectable Conc. | 0.1 mIU/mL | 0.5 mIU/mL |
| Precision (Total CV) | 4.2% @ 12.1 mIU/mL<br>4.5% @ 23.9 mIU/mL<br>5.5% @ 198.1 mIU/mL | 4.0% @ 18.3 mIU/mL<br>3.7% @ 55.7 mIU/mL<br>3.7% @ 198.1 mIU/mL |
| Correlation | y = 0.99x + 1.96<br>where<br>y = ADVIA hCG Assay<br>x = Immuno 1 hCG Assay<br>n = 40<br>r = 1.0<br>Syx = 17.5 mIU/mL | |
#### Interfering Substances
| Interfering<br>Substance | Interfering Substance<br>Concentration | | Analyte<br>Concentration, mIU/mL | | Effect |
|--------------------------|----------------------------------------|---------|----------------------------------|----------|--------|
| | SI Units | (mg/dL) | Expected | Observed | (%) |
| Hemoglobin | 1.0 g/L | 1000 | 17.2 | 16.8 | -2.3 |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 17.2 | 15.7 | -8.7 |
| Bilirubin | 428 $\mu$ mol/L | 25 | 16.5 | 15.5 | -6.1 |
| Urea Nitrogen | 71.4 mmol/L | 200 | 17.2 | 15.9 | -7.6 |
{10}------------------------------------------------
Gabriel J. Menaca Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs Bayer Corporation 511 Benedict Avenue Tarrytown, New York 10591-5097
6/30/98
Date
{11}------------------------------------------------
# 3rd Generation TSH Assay for Bayer ADVIA® Integrated Modular System
Listed below is a comparison of the performance between the ADVIA 3d Generation TSH method and a similar device that was granted clearnice of substantial equivalence (Bayer Immuno 19 TSH Assay). The information used in the Summary of Safety and Effectiveness was extracted from the ADVIA 3ª Generation TSH insert and the Immuno 1 TSH Assay method sheet.
#### INTENDED USED
This in viro method is intended to quantitatively measure TSH, Thyroid Stimulating Hormone, in human serum using ADVIA 3d Generation TSH Assay on a Bayer ADVIA® Integrated Modular System. Measurements of TSH are used in the diagnosis of thyroid or pituitary disorders. This assay allows the determination of TSH with 3d generation sensitivity of less than 20% total coefficient of variation (CV) at 0.02 µIU/mL, as defined by the American Thyroid Association.
| METHOD | ADVIA TSH Assay | | Immuno 1 TSH Assay<br>(predicate Device) | |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|------------------------------------------|--------------------------------------------------------------|
| Part No. | Reagents | B42-3921-43<br>Calibrators B43-3948-01 | Reagents | T01-2942-51<br>Calibrators T03-3568-01 |
| Minimum Detectable Conc. | 0.005 µIU/mL | | 0.03 µIU/mL | |
| Precision (Total CV) | | 13.2% @ 0.02 µIU/mL<br>2.9% @ 0.52 µIU/mL<br>2.3% @ 4.95 µIU/mL<br>2.6% @ 31.10 µIU/mL | | 6.3% @ 1.3 µIU/mL<br>2.0% @ 9.0 µIU/mL<br>1.8% @ 22.5 µIU/mL |
| Correlation | y = 0.98x - 0.357<br>where<br>y = ADVIA TSH Assay<br>x = Immuno 1 TSH Assay<br>n = 50<br>r = 0.997<br>Syx = 1.48 µIU/mL | | | |
#### Interfering Substances
| Interfering | Interfering Substance | | Analyte | | Effect |
|------------------------|-----------------------|---------|-----------------------|----------|--------|
| Substance | Concentration | | Concentration, ulU/mL | | |
| | SI Units | (mg/dL) | Expected | Observed | (%) |
| Hemoglobin | 1.0 g/L | 1000 | 3.32 | 3.38 | 1.8 |
| Lipids (Triglycerides) | 11.3 mmol/L | 1000 | 3.49 | 3.45 | -0.9 |
| Bilirubin | 428 umol/L | ટર્ડ | 3.25 | 3.30 | ાં રે |
| Urea Nitrogen | 71.4 mmol/L | 200 | 3.30 | 3.31 | -1.5 |
{12}------------------------------------------------
Gabriel J. Muraca, Jr.
Gabriel J. Muraca, Jr. Manager Regulatory Affairs
Bayer Corporation 511 Benedict Avenue
Tarrytown, New York 10591-5097
6/30/98
Date
Date
{13}------------------------------------------------
JAN 2 9 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Gabriel J. Muraca, Jr. Manager Regulatory Affairs BAYER CORPORATION Business Group Diagnostics 511 Benedict Avenue Tarrytown, New York 10591-5097
K982328 Re: Trade Name: Bayer ADVIA® IMS™ System Regulatory Class: II Product Code: CJO, CFR, CIC, CJO, DHA, JLW, JMG Dated: November 12, 1998 Received: November 19, 1998
Dear Mr. Muraca:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{14}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{15}------------------------------------------------
510(k) Number (if known):___
Device Name: Bayer ADVIA™ Integrated Module System (IMS™)
Indications For Use:
## ALKALINE PHOSPHATASE (ALP)
Intended Use:
The Bayer ADVIA IMS alkaline phosphatase (ALP) assay is an in vitro diagnostic device intended to measure ALP in human serum or plasma. Such measurements are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases. This diagnostic method is not intended for use on any other diagnostic system.
## CALCIUM (CA)
Intended Use:
The Bayer ADVIA IMS Calcium assay is an in vitro diagnostic device intended to measure calcium in human serum or plasma. Measurements of calcium are used in the diagnosis, monitoring and treatment of a variety of diseases including parathyroid disease, a variety of bone diseases, chronic renal disease, and tetany. This diagnostic method is not intended for use on any other diagnostic system.
## GLUCOSE (GLU)
Intended Use:
The Bayer ADVIA IMS Glucose assay is an in vitro diagnostic device intended to measure glucose in human serum or plasma. Such measurements are used in the diagnosis, monitoring and treatment of carbohydrate metabolism disorders, including diabetes mellitus and neonatal hypoglycemia. This diagnostic method is not intended for use on any other diagnostic system.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
Optional Format 1-2-96)
000252
Page 1 of 2
{16}------------------------------------------------
510(k) Number (if known): A4893238
# Device Name: Bayer ADVIA™ Integrated Module System (IMS™)
Indications For Use:
## FERRITIN (FERR)
Intended Use:
The Bayer ADVIA IMS Ferritin assay is an in vitro diagnostic device intended to quantitatively measure ferritin (an iron-storage protein) in human serum. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency. This diagnostic method is not intended for use on any other diagnostic system.
# HUMAN CHORIONIC GONDADOTROPIN (hCG)
Intended Use:
The Bayer ADVIA IMS hCG assay is an in vitro diagnostic device intended to quantitatively measure total beta (8) in human chorionic gonadotropin (hcG) in human serum. Measurements of human chorionic gonadotropin are used in the detection of pregnancy. This diagnostic method is not intended for use on any other diagnostic system.
## THYROID STIMULATING HORMONE (TSH)
Intended Use:
"، "ﺍﻷﻣ
The Bayer ADVIA IMS TSH assay is an in vitro diagnostic device intended to quantitatively measure thyroid stimulating hormone (TSH)in human serum. This assay allows the determination of TSH with 3d generation sensitivity of less than 20% total coefficient variation (CV) at 0.01 to 0.02 mIUL or uiU/mL, as defined by the American Thyroid Association. Measurements thyroid stimulating hormone are used in the diagnosis of thyroid or pituitary disorders. This diagnostic method is not intended for use on any other diagnostic system.
(PLEASE DO NOT WRITE BELOW THIS LINE -
| (Division Sign-Off) |
|-----------------------------------------|
| Division of Clinical Laboratory Devices |
NEEDED)
ANOTHER PAGE IF
Division of Clinica 510(k) Number -
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
000253 Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.