CAREGRAPH

K982319 · Siemens Medical Solutions USA, Inc. · JAA · Sep 22, 1998 · Radiology

Device Facts

Record IDK982319
Device NameCAREGRAPH
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeJAA · Radiology
Decision DateSep 22, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesSoftware as a Medical Device

Intended Use

CAREGRAPH is an angiographic system accessory for recording, displaying and documenting the regional distribution of the skin entry dose of a patient, depending on the relevant angulations in X-ray examination. The dose is calculated depending on the format size, the angulation and the table position and displayed on a graphical patient body surface.

Device Story

CAREGRAPH is a PC-based software accessory for Siemens angiographic and fluoroscopic X-ray systems. It receives input from a commercially available dose measurement system sensor mounted in the imaging chain. The software calculates patient skin entry dose based on X-ray format size, gantry angulation, and table position. It produces a 2D graphical representation of radiation impact on a patient body surface model, updated every 10 seconds. The system provides numerical 'Hot Spot' indications and triggers an audible alarm if user-defined skin dose limits (up to 2000 mGy) are reached. Operated by clinicians via keyboard and mouse, the software allows for patient data management, report generation, and storage of at least one week of patient records. It assists healthcare providers in monitoring radiation exposure during procedures to manage patient safety.

Clinical Evidence

No clinical data. Bench testing only. The system performs dose calculations with an accuracy of ±5% and provides a 'Hot Spot' resolution of ≤ 5x5 mm².

Technological Characteristics

PC-based software running on Windows 95/NT 4.0. Requires Pentium 133 MHz processor and 16-32 MB RAM. Interfaces with Siemens Angio/Fluoro systems via ACS or XCS interfaces. Uses external dose measurement sensor. Configurable user interface for alarms and display parameters.

Indications for Use

Indicated for patients undergoing X-ray examinations using angiographic or fluoroscopic systems to record, display, and document regional skin entry dose distribution.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP 22 1998 Attachment 10 K982319 # 510(k) Summary ## CAREGRAPH Software Program for Image-Intensified Fluoroscopic X-ray System Submitted by: Siemens Medical Systems, Inc. 186 Wood Avenue South Iselin, NJ 08830 Establishment Registration Number: 2240869 June 30, 1998 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. #### Contact Person: 1. Kathleen Rutherford Phone: (732) 321-4779 Fax: (732) 321-4841 #### 2. Device Name and Classification: | Trade Name: | CAREGRAΡΗ | |-----------------------|----------------------------------------------------------------------------| | Classification Name: | Angiographic X-ray Systems and Image Intensified Fluoroscopic X-ray System | | Classification Panel: | Radiology | | CFR Section: | 21 CFR §892.1600 and 21 CFR §892.1650 | | Device Class: | Class II | | Device Code: | 90IZZ | ### 3. Intended Use: CAREGRAPH is an angiographic system accessory for recording, displaying and documenting the regional distribution of the skin entry dose of a patient, depending on the relevant angulations in X-ray examination. The dose is calculated depending on the format size, the angulation and the table position and displayed on a graphical patient body surface. #### 4. Substantial Equivalence: We believe within the meaning of the Safe Medical Devices Act of 1990, the CAREGRAPH Software Program for Angiographic X-ray Systems and Image Intensified Fluoroscopic X-ray Systems as addressed in this Premarket notification, is substantially equivalent to the following devices for Angiographic X-ray Systems and Image Intensified Fluoroscopic X-ray Systems currently in commercial distribution. | Device | Manufacturer | 510(k) Number | |-----------------------------------------------|--------------------------------------------------------------|---------------| | Diamentor patient<br>x-ray exposure indicator | Manufacturer: PTW - Freiburg<br>Importer: Nuclear Associates | K853753 | | Diamentor M1/M2/M3 | Manufacturer: PTW - Freiburg | exempt | {1}------------------------------------------------ This statement is based on the fact that the CAREGRAPH system is based a similar PTW detector system and complies with the same or equivalent standards and has the same intended uses as the aforementioned predicate devices. The software raises no new safety or effectiveness concerns. #### 5. Device Description: CAREGRAPH is a PC program which provides a graphical display of the dose exposure to the patient as measured by a commercially available dose measurement system whose sensor is mounted in the imaging chain. CAREGRAPH can be used with all Siemens Angio and Fluoro systems that are configured with the ACS-, or XCS-interface. The software, which is delivered on CD-ROM, runs on Windows 95/Windows NT 4.0, and is designed to operate with existing Siemens dose counter features in products such as the cardiac workstations (marketed as Siemens Quantcor and Siemens ACOM.PC). The program follows the Microsoft conventions and can be configured by the customer. Minimum PC requirements are: - Pentium 133 MHz (or better), 16 Mbyte RAM (or more) for Windows 95 . - Pentium 133 MHz (or better), 32 Mbyte RAM (or more) for Windows NT 4.0. ● The standard display is a two dimensional presentation of the radiation impact in a scaled window, with numerical indications at foci of skin doses for certain limits. The numerical value of the "Hot Spot" is focused in the graphical presentation, which is updated every 10 seconds. The radiation field must be presented within 2 seconds if relevant parameters have been changed. Calculation is done with an accuracy of ±5%. The "Hot Spot" resolution shall be ≤ 5x5 mm² and limits and colors for the indication of the spot is configurable by the user. The program is operated with keyboard and mouse. The customer can set the skin dose alarm limit, though not higher than 2000 mGy. When a critical dose is reached an alarm sounds. The display of the dose area is possible without radiation. Patient data can be set parallel to data survey. It is possible to process the patient data afterwards and also generate a report on the patient and examination data. As a minimum, one week's patient reports shall be storable. Data may then be transferred and stored on another media. ## 6. Summary of Technological Characteristics of the Device Compared to the Predicate Devices: The CAREGRAPH accessory has the same technological characteristics as the predicate PTW detector system and complies with the same or equivalent standards and has the same intended uses as the aforementioned predicate devices. Kathleen Rutherford Kathleen Rutherford Manager, Regulatory Submissions Siemens Medical Systems, Inc. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP 2 2 1998 Kathleen Rutherford Manage Regulatory Submissions Siemens Medical Systems, Inc. Imaging Systems Group 186 Wood Avenue South Iselin, NJ 08830 K982319 Re: Caregraph (Accessory to Angiographic X-ray Systems and Image Intensified Fluoroscopic Systems Dated: July 1, 1998 Received: July 2, 1998 Regulatory class: II 21 CFR 892.1650/Procode: 90 JAA Dear Ms. Rutherford: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into cither class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to preceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdch/dsmamain.html". Sincerely yours, Lillian Yin, Ph.D Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Hcalth Enclosure Image /page/2/Picture/17 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and head. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. {3}------------------------------------------------ ## Attachment 1 ### Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: Siemens CAREGRAPH Accessory for Angiographic X-ray Systems and Image Intensified Fluoroscopic X-ray Systems ### Indications for Use: CAREGRAPH is an angiographic system accessory for recording, displaying and documenting the regional distribution of the skin entry dose of a patient, depending on the relevant angulations in X-ray examination. The dose is calculated depending on the format size, the angulation and the table position and displayed on a graphical patient body surface. Concurrence of the CDRH, Office of Device Evaluation (ODE) Prescription Use C (per 21 CFR 801.109) OR Over-The-Counter Use __________ David le. Seymore (Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological Devices 510(k) Number K982319
Innolitics
510(k) Summary
Decision Summary
Classification Order