The Bartels CINAkit™ CMV Antigenemia Test is intended for use as an aid in the diagnosis of Cytomegalovirus (CMV) infection by the rapid direct qualitative detection of CMV pp65 antigen in human blood leukocytes by indirect immunofluorescence (IF). This product is not intended to be used for testing (i.e. screening) blood or plasma donors.
Device Story
Device is an indirect immunofluorescence assay (IFA) for qualitative detection of CMV pp65 antigen in human peripheral blood leukocytes. Input: anticoagulated venous blood (EDTA, heparin, or ACD). Process: leukocyte isolation via dextran sedimentation and centrifugation; hypotonic lysis of residual erythrocytes; slide preparation; fixation in formalin; permeabilization with NP40 detergent; staining with murine monoclonal antibody pool (1C3, AYM-1) targeting two epitopes on pp65; secondary staining with F(Ab)2 fluorescein-conjugated anti-mouse immunoglobulin. Output: fluorescently labeled cells visualized via fluorescence microscopy. Used in clinical laboratories by trained personnel. Healthcare providers interpret fluorescence patterns to aid diagnosis of acute or reactivated CMV infection. Benefits: rapid detection of CMV antigenemia compared to traditional culture methods.
Clinical Evidence
Clinical study evaluated 335 specimens comparing Bartels CINAkit to viral culture (gold standard). Results: Sensitivity 83% (95% CI: 70.4-95.3%), Specificity 89% (95% CI: 85.1-92.3%). Cross-reactivity testing performed against 14 common viruses (HIV, HSV-1/2, VZV, EBV, HHV-6, Influenza A/B, Parainfluenza, Adenovirus, RSV) showed no cross-reactivity.
Indicated for individuals suspected of acute or reactivated CMV infection. Not for screening blood or plasma donors.
Regulatory Classification
Identification
Cytomegalovirus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to cytomegalovirus in serum. The identification aids in the diagnosis of diseases caused by cytomegaloviruses (principally cytomegalic inclusion disease) and provides epidemiological information on these diseases. Cytomegalic inclusion disease is a generalized infection of infants and is caused by intrauterine or early postnatal infection with the virus. The disease may cause severe congenital abnormalities, such as microcephaly (abnormal smallness of the head), motor disability, and mental retardation. Cytomegalovirus infection has also been associated with acquired hemolytic anemia, acute and chronic hepatitis, and an infectious mononucleosis-like syndrome.
{1}------------------------------------------------
000017
Bartels CINAkit CMV Antigenemia DEC 1 4 1998 510(k) Premarket Notification
Confidential
#### 6.0 510(K) SUMMARY
Name of Device and Classification (A)
Tradename: Bartels CINAkit CMV Antigenemia
Classification: Cytomegalovirus indirect immunofluorescence assay that consists of antisera which has been classified as a Class II (Performance Standards) device, product code, 834N, (21 CFR 866.3175).
(B) Legally Marketed Device
BARTELS CINAKIT CMV Antigenemia claims substantial equivalance to the BIOTEST CMV BRITE TEST Kit (K951550), currently in commercial distribution by BIOTEST Diagnostics, Denville, New Jersey (USA).
Device Description (C)
BARTELS CINAKIT CMV Antigenemia is an indirect immunofluorescence test that allows detection of Human Cytomegalovirus antigen in leukocytes from peripheral blood. The test uses a monoclonal antibody pool (1C3, AYM-1) which recognizes the 65-68kDa lower matrix structural phosphoprotein (pp) (protein kinase, pp65, present in the nucleus of cells. The antibody pool (blended antibodies) recognizes two epitopes on the protein.
Leukocytes are prepared from whole blood by dextran sedimentation and centrifugation; slides are prepared, fixed in formalin and permeabilized on detergent (NP40). Staining is accomplished with primary murine monoclonal antibodies to the pp65 antigen and F(Ab-)2 fluorescein-conjugated anti-mouse immunoglobulin secondary antibody. The slides are read using a fluorescence microscope.
The substantial equivalence claim between Bartels CINAkit CMV Antigenernia and Biotest CMV Brite Test is based on the attached comparison table (Table 1) and clinical study data
BARTELS CINAKIT CMV Antigenemia Method :
- Isolation of peripheral blood leukocytes 1.
- 2. Preparation of cytospin slides
- ತೆ. Fixation and permeabilization
- 4. Indirect immunofluorescence staining using anti CMV pp65
- 5. Interpretation of results
Leukocytes are isolated from whole blood by dextran sedimentation of erythrocytes and centrifugation of leukocytes followed by hypotonic lysis of residual erythrocytes with
. : : }
{2}------------------------------------------------
@003/007
000018
Bartels CINAkit CMV Antigenernia 510(k) Premarket Notification
Confidential
ammonium chloride. Slides are prepared fixed in formalin and permeabilized with detergent. Staining is accomplished with primary murine monoclonal antibodies to the pp65 antigen and F(Ab)2 fluorescein-conjugated anti-mouse immunoglobulin secondary antibody. The slides are read using a fluorescence microscope.
Immunofluorescence is the preferred method of staining. It has been reported that techniques used to block endogenous enzymatic activity (peroxidase or alkaline phosphatase) can destroy the antigen recognized in antigenemia. Bartels' blended antibody preparation targets different epitopes on the pp65 antigen and thus ensures avoidance of false negatives that may be caused by mutation of one of the epitopes recognized. This anti-pp65 preparation stains cells in positive samples.
#### (D) Intended Use
Bartels CINAkit CMV Antigenemia is intended for use as an aid in the diagnosis of Cytomegalovirus (CMV) infection by the rapid direct qualitative detection of CMV pp65 antigen in human blood leukocytes by indirect immunofluorescence (IF). This product is not intended to be used for testing (i.e. screening) blood or plasma donors.
#### (E) Comparison with the Predicate Device
## TABLE 1
| Product Name | Bartels CINAkit CMV<br>Antigenemia | Biotest CMV Brite |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use Statement | "Bartels CINAkit CMV<br>Antigenemia is intended for use as<br>an aid in the diagnosis of<br>Cytomegalovirus (CMV) infection<br>by the rapid direct qualitative<br>detection of CMV pp65 antigen in<br>human blood leukocytes by indirect<br>immunofluorescence (IF). This<br>product is not intended to be used<br>for testing (i.e. screening) blood or<br>plasma donors." | "The Biotest CMV Brite Test Kit is<br>intended for the qualitative<br>detection of Cytomegalovirus<br>(CMV) lower matrix protein pp65<br>by indirect immunofluorescence<br>using microscopy in isolated<br>peripheral blood leukocytes<br>obtained from<br>ethylenediaminetetraacetic acid<br>(EDTA) and heparin anticoagulated<br>human peripheral blood. The<br>detection of CMV pp65 in human<br>peripheral blood cells aids in the<br>diagnosis of acute or reactivated<br>CMV infection. This product is not<br>FDA cleared (approved) for use in<br>testing (i.e. screening) of blood or<br>plasma donors." |
{3}------------------------------------------------
# 000019
Bartels CINAkit CMV Antigencmia 510(k) Premarket Notification
Confidential
| Target Population | Individuals suspected of acute or reactivated CMV infection. | Individuals suspected of acute or reactivated CMV infection. |
|-----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design/format | Indirect immunofluorescence assay (IFA) | Indirect immunofluorescence assay (IFA) |
| Materials | Cocktail (2) of murine monoclonal antibodies (CINA pool, 1C3/AYM-1) against lower matrix structural phospoprotein (pp65), FITC-conjugated secondary polyclonal antibody, Evans Blue counterstain, dextran solution, erythrocyte lysing reagents, fixative, sample diluents, and mounting fluid. | Cocktail (2) of murine monoclonal antibodies (C10/C11) against lower matrix structural phospoprotein (pp65), FITC-conjugated secondary polyclonal antibody, Evans Blue counterstain, dextran solution, erythrocyte lysing reagents, fixative, sample diluents, and mounting fluid. |
| Performance Characteristics | Sensitivity: 82.86%<br>Specificity: 88.67%<br>Accuracy: 88.06% | Sensitivity: 82.86%<br>Specificity: 87.33%<br>Accuracy: 86.87% |
| Risk to patient | No unique issues of safety or effectiveness. | |
| Specimen Type | Anticoagulated [EDTA, heparin, or Adenine Citrate Dextrose (ACD)] human venous blood. | Anticoagulated (EDTA, heparin) human venous blood. |
| Analyte | CMV lower matrix structural phospoprotein (pp65) | CMV lower matrix structural phospoprotein (pp65) |
| Controls | Optional Positive and Negative Intracel CMV Control Slides (B1029-81B) not included in kit. | Positive and Negative Control Slides (20 per 100 test lab)<br>Leukocytes and SF9 Insect cells |
| Kit Size | 100 Tests | 100 Tests |
| Cell Count Required | $2 x 10^5$ cells | $1.5 x 10^5$ cells |
#### (F) Performance Data
### 】-Cross Reactivity
Bartels CINAkit CMV Antigenemia has been tested against the following viruses and has shown no cross-reactivity :
> Human Immunodeficiency Virus (HIV) - Virion (Switzerland) Human Herpes Simplex Virus Type 1 (HSV-1) strain McIntyre (HHV-1) Human Herpes Simplex Virus Type 2 (HSV-2) strain M.S. (HHV-2) Varicella Zoster Virus (VZV) strain ATCC VR586 (HHV-3) Epstein-Barr Virus (EBV) strain B95.8 (HHV-4)
{4}------------------------------------------------
INTRACEL CORPORATION
の005/007
000020
| Bartels CINAkit CMV Antigenemia | |
|---------------------------------|--|
| 510(k) Premarket Notification | |
Confidential
Herpesvirus Type 6 (HHV-6) A variant strain TAN Herpesvirus Type 6 (HHV-6) B variant strain HST Influenza A - Puerto Rico/8/34 strain Influenza B - Russia/69 strain Parainfluenza Type I - C35 strain Parainfluenza Type II -Greer strain Parainfluenza Type III -Russia/69 strain or S.B. strain Adenovirus -Type 3 GB strain Respiratory Syncytial Virus - Long VR26 strain
Each of the aforementioned 14 viruses was found to be negative by immunofluorescence using the CINAkit reagents. However, positive control antibodies produced staining of +3 to +4 for its corresponding slide. It can be concluded that the Bartels CINAkit CMV monoclonal antibodies are highly specific for CMV.
Note: Absence of cross-reactivity against Echovirus 11 and 30 has not been established
#### Performance Characteristics 2.
Bartels CINAkit CMV Antigenemia was compared to CMV virus detection by the culture method using human peripheral blood leukocytes. Three hundred thirty-five (335) clinical specimens were evaluated in comparison to the culture method. In the same study, the performance of a comparable commercial CMV Antigementia test ("Other CMV") was also was compared to CMV virus detection by culture using human peripheral blood leukocytes from the same specimens. The performance data for the two Antigenemia tests, using the culture method as the "gold standard", are presented below.
| Antigenemia Method/Result | Culture (+)<br>n=35 | Culture (-)<br>n=300 |
|-------------------------------|---------------------|----------------------|
| Bartels CINAkit CMV (+) n=63 | 29 | 34 |
| Bartels CINAkit CMV (-) n=272 | 6 | 266 |
| Other CMV (+) n=67 | 29 | 38 |
| Other CMV (-) n=268 | 6 | 262 |
### TABLE 2
Total n = 335
The performance characteristics for Bartels CINAkit CMV Antigeneraia, using the culture method as the "gold standard", are as follows:
Sensitivity = 29/35 = 83% (95% Confidence interval = 70.4 - 95.3%) Specificity = 266/300 = 89% (95% Confidence interval = 85.1 - 92.3%)
{5}------------------------------------------------
Bartels CINAkit CMV Antigenemia 510(k) Premarket Notification
- Conclusion (G)
Bartels Cina kit CMV Antigenemia is substantially equivalent to the predicate device, Biotest CMV Brite.
{6}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
DEC 1 4 1998
Fedora Daye Contreras, MPH Regulatory Affairs Associate Intracel 1330 Piccard Drive Rockville, MD 20850
Re: K982311 Trade Name: Bartels CINAkit CMV Antigenemia Regulatory Class: II Product Code: GQH Dated: October 2, 1998 Received: October 2, 1998
Dear Ms. Knapp:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{7}------------------------------------------------
## Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
510(k) Number (if known): j
Device Name: Bartels CINAkit™ CMV Antigenemia Test
Indications For Use:
The Bartels CINAkit™ CMV Antigenemia Test is intended for use as an aid in the diagnosis of Cytomegalovirus (CMV) infection by the rapid direct qualitative detection of CMV pp65 antigen in human blood leukocytes by indirect immunofluorescence (IF). This product is not intended to be used for testing (i.e. screening) blood or plasma donors.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubois
pratory Devices 510(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.