CHEM-CHEX STAT (NEW)

K982290 · Streck Laboratories, Inc. · JJY · Jul 22, 1998 · Clinical Chemistry

Device Facts

Record IDK982290
Device NameCHEM-CHEX STAT (NEW)
ApplicantStreck Laboratories, Inc.
Product CodeJJY · Clinical Chemistry
Decision DateJul 22, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

Chem-Chex STAT is intended to be used for verifying the accuracy of i-STAT chemistry analyzers used with i-STAT test cartridges. Chem-Chex STAT is assayed for pH, Na*, Ca**, CIT, K*, BUN, Hgb, Hct, and Glucose.

Device Story

Chem-Chex STAT is a multi-analyte, liquid quality control product; consists of stabilized mammalian red blood cells in an aqueous buffered medium. Used by laboratory personnel to verify performance of i-STAT chemistry analyzers and test cartridges. Provides positive procedural control for pH, Na+, Ca++, Cl-, K+, BUN, Hgb, Hct, and Glucose. Packaged in dropper-tip polyethylene vials. Users compare analyzer results against provided assay values to ensure system accuracy; facilitates clinical decision-making by confirming diagnostic equipment reliability.

Clinical Evidence

Bench testing only. Studies included lot-to-lot reproducibility, open and closed vial stability, and site-to-site comparisons at four locations. Results demonstrated consistent reproducibility and stability for the duration of product dating.

Technological Characteristics

Stabilized mammalian red blood cells in aqueous buffered medium. Form factor: 2.0 mL dropper-tip polyethylene vials with polypropylene screw-top caps. Analyte composition: Na+, Cl-, K+, BUN, Glucose, pH, Ca++, Hgb, Hct. Standalone control product; no connectivity or software.

Indications for Use

Indicated for use as an assayed quality control material to verify the accuracy of i-STAT chemistry analyzers and test cartridges. No specific patient population is targeted as this is an in vitro diagnostic control product.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness | 510(k) Submitter: | Streck Laboratories, Inc.<br>14124 Industrial Road<br>Omaha, Nebraska 68144 | |-------------------------|-----------------------------------------------------------------------------| | Official Correspondent: | Paul Kittelson<br>Quality Assurance Manager | K982290 (402) 691-7465 June 26. 1998 Date Prepared: ## Names of Device: Trade Name: Common Name: Classification Name: Classification Number: Chem-Chex STAT Assaved chemistry control Multi-analyte control (§862.1660) 75JJY Predicate Device: i-STAT control (K934641) manufactured by i-STAT Corp. Chem-Chex STAT control (K96115) manufactured by Streck Laboratory. #### Description: Chem-Chex STAT is a suspension of stabilized mammalian red blood cells in an aqueous buffered medium. Manufactured in three clinically significant levels, it is packaged in dropper tip polyethylene vials containing 2.0 mL volumes. Closures are injection molded polypropylene screw-top caps. Vials are packaged in polystyrene jars. #### Intended Use: Chem-Chex STAT is intended to be used for verifying the accuracy of i-STAT chemistry analyzers used with i-STAT test cartridges. Chem-Chex STAT is assayed for pH, Na*, Ca**, CIT, K*, BUN, Hgb, Hct, and Glucose. # Comparison with Predicate Devices: Like the i-STAT Control and former Chem-Chex STAT (96115), Chem-Chex STAT (New) is intended to enable the user to verify satisfactory performance of the i-STAT analyzer and cartridge system. All levels of the device are liquid containing sodium, chloride, potassium, blood urea nitrogen, and glucose. Like its predecessor control, Chem-Chex STAT (New) is an assayed control that provides the user a complete positive procedural control. It contains red blood cells and is assayed for hemoglobin and hematocrit, it is not assayed for PCO2, or PO2, like the i-STAT. ## Discussion of Tests and Test Results: Three studies of Chem-Chex STAT (New) were conducted: () lot to lot reproducibility; 11) open and closed vial stability; and III) site to site comparisons at 4 different locations. Study results showed Chem-Chex STAT (New) to be consistently reproducible, substantially equivalent to the predicate products, and stable for the entire product dating. ## Conclusions Drawn from Tests: Study results show Chem-Chex STAT (New) to be consistently reproducible, substantially equivalent to the predicate products, and stable for the entire product dating. Chem-Chex STAT (New) is a safe and effective product which fulfills its intended use when used as instructed in the product package insert. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 JUL 22 1998 Paul Kittelson Quality Assurance/Regulatory Affairs Manager Streck Laboratories, Inc. 14124 Industrial Road Omaha, Nebraska 68144 K982290 Re : Chem-Chex STAT (New) Regulatory Class: I Product Code: JJY Dated: June 26, 1998 Received: June 29, 1998 Dear Mr. Kittelson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Failure to Administration (FDA) will verify such assumptions. comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of ' Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven Litman Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ 510(k) Number (if known): Device Name: __________ Chem-Chex STAT Indications For Use: Chem-Chex STAT is intended to be used for verifying the accuracy of of i-STAT chemistry analyzers used with i-STAT test cartridges. Chem-Chex STAT is assayed for Nat CI · K* BUN, Hgb, Hct, Glucose, pH, and Ca**. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) **Prescription Use** (Per 21 CFR 801.109) Division " Divi. Jour 14:00 310/k; Aun. Over-The-Counter Use OR (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...