The FirstSave STAR Biphasic AED is intended for use by or on the order of a Physician or persons licensed by State law. It is designed for emergency treatment of cardiac arrest patients by trained personnel. The user assesses the patient's condition and confirms that the patient is unconscious, pulseless and is not breathing prior to use of the device. The FirstSave STAR Biphasic AED is intended to be used on patients older than eight years'.
Device Story
Portable, battery-operated, semi-automatic external defibrillator (AED) for emergency cardiac arrest treatment. Device inputs EKG signals via electrodes; diagnostic algorithm analyzes rhythm to classify as shockable or non-shockable. Operator initiates shock via single button. Provides voice prompts and LED status indicators. Used by trained personnel in emergency settings. Output is biphasic defibrillation shock to terminate fatal arrhythmias. Benefits patient by restoring normal cardiac rhythm.
Clinical Evidence
Clinical studies were performed to evaluate the efficacy of the biphasic waveform. Results demonstrated that the biphasic waveform is equivalent to the monophasic truncated exponential waveform used in the predicate FirstSave device. Additional bench testing included rhythm detection, EMC, charge time, pulse shape, battery capacity, and defibrillation recovery, all complying with AAMI DF39-1993 and AAMI DF2-1996 standards.
Technological Characteristics
Portable, battery-powered, semi-automatic low-power DC defibrillator. Features biphasic defibrillation waveform, lithium battery with fuel gauge, LED diagnostic panel, and voice prompts. Dimensions: 3.3in x 10.6in x 12.4in; weight: 7.4 lbs. Complies with AAMI DF39-1993 and AAMI DF2-1996. Software and hardware developed under design controls.
Indications for Use
Indicated for emergency treatment of cardiac arrest in patients older than eight years who are unconscious, pulseless, and not breathing. For use by trained personnel or on the order of a physician.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
{0}------------------------------------------------
K982264
JAN 25 1999
# 510(k) Summary
#### ADMINSTRATIVE INFORMATION L-1.
## L-1.1 Name and address
SurVivaLink Corporation Submitted by: 5430 Feltl Road Minneapolis, MN 55343
| Contact Person: | Sew-Wah Tay, Ph.D. |
|-----------------|--------------------|
| Telephone No.: | 612-939-2942 |
| Facsimile No.: | 612-939-4191 |
June 24, 1998 Date Prepared:
## L-1.2 Device Name
Common or Usual Name: Semi-Automatic External Defibrillator (AED) Device Name: FirstSave™ STAR Biphasic™ AED
#### L-1.3 Classification
Classification Name: a) DC defibrillator
21CFR&870.5300; Class II
b) Cardiac Monitor (Cardiotachometer and Rate Alarm) 21CFR§870.2300; Class II
Note: FDA has determined that Semi-Automatic External Defibrillators are currently classified as class III devices
### L-1.4 Applicant
Applicant's Name: SurVivaLink, Corporation 5430 Feltl Road Minneapoils, MN 55343
#### L-2. PREDICATE DEVICE
FirstSave™ AED manufactured by SurVivaLink Corporation models 9100/9110 (K970481) and Forerunner AED manufactured by HeartStream models S, E and EM (K955628).
#### L-3. INDICATION FOR USE
The SurVivaLink's FirstSave STAR Biphasic AED is designed for emergency treatment of cardiac arrest by trained personnel to terminate potentially fatal arrhythmias for patients older than eight years old. The user assesses the patient's condition and confirms that the patient is unconscious, pulseless and is not breathing prior to use of the device.
{1}------------------------------------------------
### DEVICE DESCRIPTION L-4
The FirstSave STAR Biphasic AED is a portable battery operated semi-automatic low power DC defibrillator. The device's diagnostic algorithm analyzes the patient's cardiac rhythm to determine shockable versus non-shockable EKG rhythm. The operator then pushes the button to deliver the defibrillation shock. The FirstSave STAR Biphasic AED feature includes:
- . Biphasic defibrillation waveform
- . Lithium battery and battery fuel gauge
- Single user button .
- LED diagnostic panel ●
- Non-volatile status indicator ●
- Voice prompts .
The device weighs 7.4 lbs including battery, and electrodes with dimensions of 3.3in x 10.6in x 12.4in.
### PERFORMANCE DATA L-5.
The performance test data is provided in the 510(k) submission. The performance data demonstrate that the device complies with the applicable sections of AAMI DF39-1993 and AAMI DF2-1996. The device was developed under design control and the hardware and software tested in accordance with established industry standards.
Tests results include rhythm detection, EMC, charge time, pulse shape, battery capacity, defibrillation recovery, design verification and validation data for hardware and software incorporated into the FirstSave STAR Biphasic AED. Environmental tests were performed on the finished device.
Clinical studies was performed to test the efficacy of the biphasic waveform in the FirstSave STAR Biphasic AED device and was found to be equivalent to the monophasic truncated exponential waveform in the FirstSave.
Test data demonstrate that the safety and effectiveness of the FirstSave STAR Biphasic AED are substantially equivalent to the HeartStream's Forerunner and the SurVivaLink's FirstSave AEDs.
### L-6 SUBSTANTIAL EQUIVALENCE
The FirstSave STAR Biphasic AED is a portable, battery powered, semi-automatic, low energy DC external defibrillator. The materials, accessories and user interfaces are identical to the FirstSave AED (K970481).
The Company's FirstSave STAR Biphasic AED covered by this submission is substantially equivalent to other legally marketed semi-automatic low power DC
{2}------------------------------------------------
defibrillators. Specifically, the FirstSave STAR Biphasic AED is substantially equivalent to the SurVivaLink's FirstSave and the HeartStream Forerunner AED. The FirstSave STAR Biphasic AED has the same general indication for use, similar principles of operation, and similar technological characteristics as the previously cleared Forerunner and FirstSave devices. Although there are minor differences in the characteristics of the FirstSave STAR Biphasic AED and its predicate devices, those differences do not raise new questions of safety or efficacy.
的人物的人物的人物
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, featuring a staff with three intertwined snakes.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 25 1999
Sew-Wah Tay, Ph.D. SurVivaLink Corporation 5420 Feltl Road Minneapolis, MN 55343
K982264 Re: FirstSave STAR Biphasic Regulatory Class: III (three) Product Code: 74 MKJ November 13, 1998 Dated: November 16, 1998 Received:
Dear Dr. Tay:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{4}------------------------------------------------
Page 2 - Sew-Wah Tay, Ph.D.
On August 16, 1993 the Final Rule for Device Tracking was published in the Federal Register, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices throuqh the distribution chain and then identifies and follows the patients who receive them. The specific requirements of the requlation are found in 21 CFR 821 as described in the August 16, 1993 Federal Register beginning on page 43447.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please not the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Arle A. Cich.
Yo Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosures
{5}------------------------------------------------
# 510(k) Number : K982264
# Device Name: FirstSave STAR Biphasic AED
# Indication For Use
The FirstSave STAR Biphasic AED is intended for use by or on the order of a Physician or persons licensed by State law. It is designed for emergency treatment of cardiac arrest patients by trained personnel. The user assesses the patient's condition and confirms that the patient is unconscious, pulseless and is not breathing prior to use of the device. The FirstSave STAR Biphasic AED is intended to be used on patients older than eight years'.
Marle Kramer
PRESCRIPTION USE
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number _
<sup>1</sup> American Heart Association (1994): "Advanced Cardiac Life Support" Ed by R.O. Cummins. pp 4-11.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.